Targeting accuracy at couch kick for a frameless image guided radiosurgery system

被引:1
|
作者
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Kim, Joshua [1 ]
Wen, Ning [1 ]
Chetty, Indrin J. [1 ]
Siddiqui, Salim [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 W Grand Blvd, Detroit, MI 48202 USA
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2018年 / 5卷 / 02期
关键词
frameless radiosurgery; trigeminal neuralgia; image-guided radiosurgery;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose/Objectives: Targeting accuracy at all possible couch angles needs to be carefully evaluated prior to initiating a frameless image-guided stereotactic radiosurgery program on a Linac for treating functional disorders such as trigeminal neuralgia. In this study, we report positioning accuracy with stereoscopic x-ray imaging over the complete range of couch rotation using anthropomorphic head phantoms. Materials/Methods: An anthropomorphic head phantom with three 5 mm tungsten BBs as hidden targets was CT simulated. A group of 7 arcs was planned with couch angles from 0 to 90 degrees in 15 degrees increments. A pair of stereoscopic x-ray images that auto-matches to planning CT according to bony anatomy was utilized to position each BB to machine isocenter at all planned couch angles. Targeting accuracy was measured by stereoscopic x-ray imaging of the BB itself, which provides the distance from the centroid of BB to the x-ray imaging isocenter. For each BB, the hidden target test was repeated 5 times at couch 0 degrees and 3 times at other couch angles, resulting in a total of 69 measurements, each with random initial setup deviation. Following the same workflow, a second anthropomorphic head phantom with two 5 mm BBs was utilized to evaluate localization accuracy at couch angles of 0 through 270 degrees in 15 degrees increments, resulting in another 18 measurements. Results: Residual setup deviation following image guidance in the first head phantom was 0.6 +/- 0.1, 0.4 +/- 0.1, and 0.4 +/- 0.1 mm, respectively, at the three BBs, and 0.5 +/- 0.1 mm overall (N=69). Comparable results of 0.4 +/- 0.1 mm (N=18) were achieved with the second head phantom. Combining results from both phantoms, the targeting accuracy was 0.5 +/- 0.1 mm (range 0.2-0.8 mm). No apparent correlation was observed between targeting accuracy and couch rotation. Conclusions: Accurate positioning within 1 mm can be achieved with stereoscopic x-ray imaging at any couch angle. Frameless image-guided stereotactic radiosurgery could achieve targeting accuracy similar to that of frame-based systems for high dose treatment of trigeminal neuralgia.
引用
收藏
页码:123 / 129
页数:7
相关论文
共 50 条
  • [1] Assessment of targeting accuracy at couch kick in frameless LINAC-based stereotactic radiosurgery of cranial metastases
    Grohmann, M.
    Todorovic, M.
    Petersen, C.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2020, 196 (SUPPL 1) : S47 - S48
  • [2] The accuracy of dose localization for an image-guided frameless radiosurgery system
    Murphy, MJ
    Cox, RS
    [J]. MEDICAL PHYSICS, 1996, 23 (12) : 2043 - 2049
  • [3] Novalis frameless image guided radiosurgery
    Wurm, R. E.
    Erbel, S.
    Schwenkert, I.
    Gum, F.
    Scheffler, D.
    Schlenger, L.
    Agaoglu, D.
    Schild, R.
    Budach, V.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S263 - S263
  • [4] Alignment Accuracy and Intrafraction Motion Analysis in Frameless Functional Image Guided Radiosurgery
    Mathew, P. K.
    Shankar, V.
    Haritha, C.
    Prajapathi, J.
    Tinimol, P.
    Rajesh, R.
    Joseph, J.
    Shinde, P. K.
    Chaudhari, L. N.
    Kumar, A.
    Kumar, R.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S307 - S307
  • [5] Alignment accuracy & intrafraction motion analysis in frameless functional image guided radiosurgery
    Shankar, V.
    Gupta, A. R.
    Vinoth, K.
    Sathish, P.
    Rajesh, R.
    Ranjan, K.
    Jijo, J.
    Haritha, C.
    Sachin, J.
    Das, S.
    Jain, D.
    [J]. EUROPEAN JOURNAL OF CANCER, 2016, 54 : S59 - S59
  • [6] The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch system
    Takakura, T.
    Mizowaki, T.
    Nakata, M.
    Yano, S.
    Fujimoto, T.
    Miyabe, Y.
    Nakamura, M.
    Hiraoka, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01): : 1 - 10
  • [7] Novalis frameless image guided radiosurgery and radiotherapy
    Wurm, R.
    Erbel, S.
    Schwenkert, I.
    Agaoglu, D.
    Schild, R.
    Scheffler, D.
    Schlenger, L.
    Burlach, V.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 : 25 - 25
  • [8] TECHNIQUE FOR TARGETING ARTERIOVENOUS MALFORMATIONS USING FRAMELESS IMAGE-GUIDED ROBOTIC RADIOSURGERY
    Hristov, Dimitre
    Liu, Lina
    Adler, John R.
    Gibbs, Iris C.
    Moore, Teri
    Sarmiento, Marily
    Chang, Steve D.
    Dodd, Robert
    Marks, Michael
    Do, Huy M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 79 (04): : 1232 - 1240
  • [9] Targeting Accuracy of Image-Guided Radiosurgery for Intracranial Lesions
    Huang, Y.
    Wen, N.
    Zhao, B.
    Kim, J.
    Gordon, J.
    Chetty, I.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 184 - 184
  • [10] Novalis frameless image-guided radiosurgery and radiotherapy
    Wurm, R.
    Erbel, S.
    Gum, F.
    Schlenger, L.
    Agaoglu, D.
    Schild, R.
    Budach, V
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2008, 184 : 59 - 59