Evaluation of the AAA Treatment Planning Algorithm for SBRT Lung Treatment: Comparison with Monte Carlo and Homogeneous Pencil Beam Dose Calculations

被引:9
|
作者
Gete, Ermias [1 ]
Teke, Tony [2 ]
Kwa, William [1 ]
机构
[1] British Columbia Canc Agcy, Vancouver Ctr, 600 W 10th Ave, Vancouver, BC V5Z 4E6, Canada
[2] Univ British Columbia, Dept Phys & Astron, Vancouver, BC, Canada
关键词
D O I
10.1016/j.jmir.2011.09.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the dose calculation accuracy of the Varian Eclipse anisotropic analytical algorithm ( AAA) for stereotactic body radiation therapy (SBRT), and to investigate the dosimetric consequences of not applying tissue heterogeneity correction on complex SBRT lung plans. Materials and Methods: Nine cases of non-small-cell lung cancer (NSCLC) that were previously treated with SBRT at our center were selected for this study. Following Radiation Therapy Oncology Group 0236, the original plans were calculated using pencil beam without heterogeneity correction (PBNC). For this study, these plans were recalculated by applying tissue heterogeneity correction with the AAA algorithm and with the Monte Carlo (MC) method, keeping the number of monitor units the same as the original plans. Two kinds of plan comparison were made. First, the AAA calculations were compared with MC. Second, the treatment plans that were calculated with AAA were compared with the original PBNC calculations. The following dose-volume parameters were used for the comparison: V-100%; V-90%; the maximum, the minimum, and the mean planning target volume (PTV) doses (D-max, D-min, and D-mean, respectively); V-20Gy, V-15Gy, V-10Gy, V-5Gy; Dmean for the lung; and Dmax for the critical organs. Results: Comparable results were obtained for AAA and MC calculations: except for Dmax, Dmin, and Dmean, the differences in the patient-average values of all of the PTV dose parameters were less than 2%. The largest average difference was observed for Dmin (3.8 +/- 5.4%). Average differences in all the lung dose parameters were under 0.2%, and average differences in normal tissue Dmax were under 0.3 Gy, except for the skin dose. There were appreciable differences in the PTV and normal tissue dose-volume parameters when comparing AAA and PBNC calculations. Except for V-100% and V-90%, PBNC calculations on average underestimated the dose to the PTV. The largest discrepancy was in the PTV maximum dose, with a patient-averaged difference of 11.1 +/- 4.6%. Conclusions: Based on our MC investigation, we conclude that the Eclipse AAA algorithm is sufficiently accurate for dose calculations of lung SBRT plans involving small 6-MV photon fields. Our results also demonstrate that, although dose calculations at the periphery of the PTV showed good agreement when comparing PBNC with both AAA and MC calculations, there is a potential to significantly underestimate the dose inside the PTV and doses to critical structures if tissue heterogeneity correction is not applied to lung SBRT plans.
引用
收藏
页码:26 / 33
页数:8
相关论文
共 50 条
  • [31] A virtual-accelerator-based verification of a Monte Carlo dose calculation algorithm for electron beam treatment planning in homogeneous phantoms
    Wieslander, E
    Knöös, T
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06): : 1533 - 1544
  • [32] Accuracy of electron beam Monte Carlo calculations in radiotherapy treatment planning
    Arends, M
    Ausma, A
    van 't Veld, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S43 - S43
  • [33] Monte Carlo evaluation of a photon pencil kernel algorithm applied to fast neutron therapy treatment planning
    Söderberg, J
    Carlsson, GA
    Ahnesjö, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (20): : 3327 - 3344
  • [34] Pseudo Proton Radiography Beam Validation of Monte Carlo Dose Calculation in Two Pencil Beam Scanning Treatment Planning Systems
    Chang, C.
    Zhou, J.
    Yang, X.
    Dhabaan, A.
    Zhang, R.
    Liu, T.
    McDonald, M.
    Langen, K.
    Lin, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E592 - E592
  • [35] Comparison of Dose Calculation for Brain, Lung, Pelvis and Abdomen Targets with Monte Carlo Simulation and Eclipse Pencil Beam Algorithm
    Prusator, M.
    Muller, L.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E261 - E262
  • [36] SBRT of lung tumours: Monte Carlo simulation with PENELOPE of dose distributions including respiratory motion and comparison with different treatment planning systems
    Panettieri, Vanessa
    Wennberg, Berit
    Gagliardi, Giovanna
    Duch, Maria Amor
    Ginjaume, Merce
    Lax, Ingmar
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14): : 4265 - 4281
  • [37] USE OF MONTE-CARLO GENERATED PENCIL BEAM DATA FOR ELECTRON TREATMENT PLANNING
    RILEY, R
    MOHAN, R
    CHUI, C
    [J]. MEDICAL PHYSICS, 1985, 12 (04) : 516 - 516
  • [38] Comparison of XVMC Monte Carlo Dose Calculations with Eclipse AAA Calculations for RapidArc Plans
    Stelljes, T.
    Alber, M.
    Poppe, B.
    Laub, W.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3818 - 3818
  • [39] AAA vs Monte Carlo Dose Calculation Algorithm for Lung SABR
    Badragan, I.
    Huang, V.
    Shellenberg, D.
    Krauze, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S982 - S983
  • [40] A novel lateral disequilibrium inclusive, (LDI) pencil-beam based dose calculation algorithm: Evaluation in inhomogeneous phantoms and comparison with Monte Carlo calculations
    Wertz, Hansjoerg
    Jahnke, Lennart
    Schneider, Frank
    Polednik, Martin
    Fleckenstein, Jens
    Lohr, Frank
    Wenz, Frederik
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1627 - 1634