To determine the source dwell positions of HDR brachytherapy using 2D 729 ion chamber array

被引:1
|
作者
Kumar, S. [1 ]
Gangadharan, Sitha [2 ]
Cheruparambil, Aswathi [2 ]
Parakat, Anjana [2 ]
Perumangat, Aparna [2 ]
机构
[1] Malabar Canc Ctr, Div Radiat Phys, Dept Radiat Oncol, Thalassery 670103, Kerala, India
[2] Univ Calicut, Dept Phys, Malappuram, Kerala, India
关键词
afterloading; applicators; HDR brachytherapy; 2D array;
D O I
10.1017/S1460396915000382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to determine the dwell position of a high-dose- rate (HDR) brachytherapy Ir-192 source using a PTW Seven29 2D detector array. A Nucletron Microselectron HDR device and 2D array ionisation chamber, equipped with 729 ionisation chambers uniformly arranged in a 27 x 27 matrix with an active array area of 27 x 27 cm(2), were used for this study. Different dwell positions were assigned in the HDR machine. Rigid interstitial needles and a vaginal applicator were positioned on the 2D array, which was then exposed according to the programmed dwell positions. Subsequently, the positional accuracy of the source position was analysed. This process was repeated for different dwell positions. The results were analysed using an in-house- developed Excel programme. Different random dwell position checks as well as dwell position measurements were performed using a radiochromic film. The dwell positions measured by the 2D array were found to be in good agreement with those measured by the film. The standard deviations between the doses obtained from the different dwell positions were 0.191828, 0.329973, 0.370632 and 0.779939, whereas the corresponding standard deviations of the doses at the vaginal cylinder were 0.60303, 0.242808, 0.242808 and 0.065309. When the planned and measured dwell positions were plotted, a linear relationship was obtained.
引用
收藏
页码:403 / 409
页数:7
相关论文
共 50 条
  • [31] Acceptance Modualted Radiation Intensity and Enhanced Dynamic Wedge using 2D Ion Chamber Array
    Contreras, O. J. Garcia
    Casas, L. D.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 357 - 357
  • [32] Direct reconstruction and associated uncertainties of 192Ir source dwell positions in ring applicators using gafchromic film in the treatment planning of HDR brachytherapy cervix patients
    Awunor, O. A.
    Dixon, B.
    Walker, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (10): : 3207 - 3225
  • [33] QA of a tomotherapy unit by means of a 2D ion chamber array and the octavius phantom
    Hambach, L.
    Clermont, C.
    Devillers, M.
    Van Esch, A.
    Huyskens, D.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S82 - S82
  • [34] Gantry-Angle Dependence of a 2D Ion Chamber Array for IMAT QA
    O'Daniel, J.
    Kishore, M.
    Das, S.
    Wu, Qj
    Yoo, S.
    Yin, F.
    MEDICAL PHYSICS, 2010, 37 (06) : 3237 - +
  • [35] Dosimetric Characteristics of 2D Ion Chamber Array Matrix for Verification Radiotherapy Treatments
    Sathiyan, S.
    Ravikumar, M.
    Supe, S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [36] A FLUKA STUDY OF CROSSTALK BETWEEN MEASURING VOLUMES IN A 2D ION CHAMBER ARRAY
    Popovici, Maria Ana
    Vasilache, Radu Alin
    Buzatu, Daniela
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN-SERIES A-APPLIED MATHEMATICS AND PHYSICS, 2018, 80 (04): : 301 - 308
  • [37] VMAT Monthly Quality Assurance Using a 2D Ionization Chamber Array
    Myers, P.
    Buckey, C.
    Mihailidis, D.
    Mavroidis, P.
    Esquivel, C.
    Gutierrez, A.
    Papanikolaou, N.
    Stathakis, S.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [38] The impact of activating source dwell positions outside the CTV on the dose to treated normal tissue volumes in TRUS guided 3D conformal interstitial HDR brachytherapy of prostate cancer
    Karlsson, Leif
    Thunberg, Per
    Jahansson, Bengt
    Persliden, Jan
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2014, 6 (03) : 282 - 288
  • [39] Diagnostic X-Ray Beam Profiles Measured with a 2D Ion Chamber Array
    Li, Texin
    Gong, Jason
    Tong, Shidong
    Fountain, Betty
    Duffy, Edward
    MEDICAL PHYSICS, 2010, 37 (06)
  • [40] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Rogelio Manuel Diaz Moreno
    Daniel Venencia
    Edgardo Garrigo
    Yakov Pipman
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 145 - 151