To determine the source dwell positions of HDR brachytherapy using 2D 729 ion chamber array

被引:1
|
作者
Kumar, S. [1 ]
Gangadharan, Sitha [2 ]
Cheruparambil, Aswathi [2 ]
Parakat, Anjana [2 ]
Perumangat, Aparna [2 ]
机构
[1] Malabar Canc Ctr, Div Radiat Phys, Dept Radiat Oncol, Thalassery 670103, Kerala, India
[2] Univ Calicut, Dept Phys, Malappuram, Kerala, India
关键词
afterloading; applicators; HDR brachytherapy; 2D array;
D O I
10.1017/S1460396915000382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to determine the dwell position of a high-dose- rate (HDR) brachytherapy Ir-192 source using a PTW Seven29 2D detector array. A Nucletron Microselectron HDR device and 2D array ionisation chamber, equipped with 729 ionisation chambers uniformly arranged in a 27 x 27 matrix with an active array area of 27 x 27 cm(2), were used for this study. Different dwell positions were assigned in the HDR machine. Rigid interstitial needles and a vaginal applicator were positioned on the 2D array, which was then exposed according to the programmed dwell positions. Subsequently, the positional accuracy of the source position was analysed. This process was repeated for different dwell positions. The results were analysed using an in-house- developed Excel programme. Different random dwell position checks as well as dwell position measurements were performed using a radiochromic film. The dwell positions measured by the 2D array were found to be in good agreement with those measured by the film. The standard deviations between the doses obtained from the different dwell positions were 0.191828, 0.329973, 0.370632 and 0.779939, whereas the corresponding standard deviations of the doses at the vaginal cylinder were 0.60303, 0.242808, 0.242808 and 0.065309. When the planned and measured dwell positions were plotted, a linear relationship was obtained.
引用
收藏
页码:403 / 409
页数:7
相关论文
共 50 条
  • [21] A study of enhanced dynamic wedge dosimetry using a 2D ion chamber array detector
    Kumar, S.
    Aparna, P.
    Anjana, P.
    Aswathi, C.
    Sitha, G.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (04) : 394 - 402
  • [22] A New Method for Quality Control of Multileaf Collimators Using a 2D Ion Chamber Array
    Moreno, R. Diaz
    Venencia, C.
    Garrigo, E.
    Pipman, Y.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [23] Patient-specific QA for CyberKnife plans using 2D ion chamber array
    Adamczyk, S.
    Skrobala, A.
    Adamczyk, M.
    Pawalowski, B.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S171 - S171
  • [24] Measurement of Penumbra Using a 2D Ionization Chamber Array
    Roberts, D.
    Litzenberg, D.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [25] Single vs double resolution performance of a 2D ion chamber array
    Sevillano, D.
    Garcia-Vicente, F.
    Castro, P.
    Ferrer, C.
    Perez, L.
    Torres, J.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S195 - S195
  • [26] Measuring field parameters of a linear accelerator with a 2D ion chamber array
    Henkner, K.
    Rhein, B.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S192 - S193
  • [27] A multiple acquisition sequence for IMRT verification with a 2D ion chamber array
    Spezi, Emiliano
    Angelini, Anna Lisa
    Ferri, Andrea
    MEDICAL DOSIMETRY, 2006, 31 (04) : 269 - 272
  • [28] Dosimetric verification of enhanced dynamic wedges by a 2D ion chamber array
    Oh, Se An
    Kim, Sung Kyu
    Kang, Min Kyu
    Yea, Ji Woon
    Kim, Eng Chan
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 63 (11) : 2215 - 2219
  • [29] Dosimetric verification of enhanced dynamic wedges by a 2D ion chamber array
    Se An Oh
    Sung Kyu Kim
    Min Kyu Kang
    Ji Woon Yea
    Eng Chan Kim
    Journal of the Korean Physical Society, 2013, 63 : 2215 - 2219
  • [30] Measurement of Proton Beam Spot Size Using a 2D Ion Chamber Array for Daily QA
    Thorne, N.
    Kassaee, A.
    MEDICAL PHYSICS, 2016, 43 (06) : 3500 - 3500