The purpose of this study was to determine the dwell position of a high-dose- rate (HDR) brachytherapy Ir-192 source using a PTW Seven29 2D detector array. A Nucletron Microselectron HDR device and 2D array ionisation chamber, equipped with 729 ionisation chambers uniformly arranged in a 27 x 27 matrix with an active array area of 27 x 27 cm(2), were used for this study. Different dwell positions were assigned in the HDR machine. Rigid interstitial needles and a vaginal applicator were positioned on the 2D array, which was then exposed according to the programmed dwell positions. Subsequently, the positional accuracy of the source position was analysed. This process was repeated for different dwell positions. The results were analysed using an in-house- developed Excel programme. Different random dwell position checks as well as dwell position measurements were performed using a radiochromic film. The dwell positions measured by the 2D array were found to be in good agreement with those measured by the film. The standard deviations between the doses obtained from the different dwell positions were 0.191828, 0.329973, 0.370632 and 0.779939, whereas the corresponding standard deviations of the doses at the vaginal cylinder were 0.60303, 0.242808, 0.242808 and 0.065309. When the planned and measured dwell positions were plotted, a linear relationship was obtained.
引用
收藏
页码:403 / 409
页数:7
相关论文
共 50 条
  • [1] To Determine the Source Dwell Positions of HDR Brachytherapy Using 2D 729 Ion Chamber Array
    SyamKumar
    Sitha
    MEDICAL PHYSICS, 2015, 42 (06) : 3511 - 3511
  • [2] Autoactivation of source dwell positions for HDR brachytherapy treatment planning
    Giannouli, S
    Baltas, D
    Milickovic, N
    Lahanas, M
    Kolotas, C
    Zamboglou, N
    Uzunoglu, N
    MEDICAL PHYSICS, 2000, 27 (11) : 2517 - 2520
  • [3] Determination of the effective chamber position of 2D ion chamber array using an HDR afterloading device
    Biggs, P.
    MEDICAL PHYSICS, 2006, 33 (06) : 2085 - 2085
  • [4] A novel approach to locating source dwell positions in HDR brachytherapy gynaecological applicators
    Franich, R.
    Hanlon, M.
    Smith, R. L.
    Demsey, C.
    Millar, J. L.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S473 - S474
  • [5] HDR brachytherapy invivo source position verification using a 2D diode array: A Monte Carlo study
    Poder, Joel
    Cutajar, Dean
    Guatelli, Susanna
    Petasecca, Marco
    Howie, Andrew
    Bucci, Joseph
    Rosenfeld, Anatoly
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 163 - 172
  • [6] Dwell positions of the source in Ring Applicator ("z-coordinate") in HDR brachytherapy commissioning
    Stserbakov, V.
    Tsizik, A.
    Saveljev, N.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1609 - S1610
  • [7] Using PTW Seven29 Ion Chamber Array for HDR Source Acceptance and for HDR Brachytherapy Treatment Planning System QA
    Bassalow, R.
    Pavord, D.
    Gong, J.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [8] Examining Acquisition Rate for Using MatriXX Ion Chamber Array to Measure HDR Brachytherapy Treatments
    Wagar, M.
    Bhagwat, M.
    O'Farrell, D.
    Friesen, S.
    Buzurovic, I.
    Damato, A.
    Devlin, P.
    Cormack, R.
    MEDICAL PHYSICS, 2015, 42 (06) : 3429 - 3429
  • [9] IMRT plan verification using a 2D ion chamber array
    Spezi, E
    Angelini, AL
    Romani, F
    Ferri, A
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S176 - S177
  • [10] A Digital QA Solution Using 2D Ion Chamber Array
    Dou, Y.
    Yan, Y.
    Velasco, C.
    Zhang, X.
    Chao, M.
    Han, E.
    Moros, E.
    MEDICAL PHYSICS, 2010, 37 (06)