Bedside Ultrasonography as an Alternative to Computed Tomography Scan for the Measurement of Optic Nerve Sheath Diameter

被引:6
|
作者
Yanamandra, Uday [1 ]
Gupta, Amul [3 ]
Yanamandra, Sushma [2 ]
Das, Subrat Kumar [2 ]
Patyal, Sagarika [4 ]
Nair, Velu [5 ]
机构
[1] Army Hosp Res & Referral, Dept Hematol, New Delhi, India
[2] Army Hosp Res & Referral, New Delhi, India
[3] Mil Hosp, Dept Radiol, Jaipur, Rajasthan, India
[4] Minist Def, New Delhi, India
[5] Gen Med Serv Army, New Delhi, India
关键词
Computed tomography scan; high altitude; intracranial pressure; optic nerve sheath diameter;
D O I
10.4103/jnrp.jnrp_537_17
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Optic nerve sheath diameter (ONSD) as measured by optic nerve sheath ultrasonography (ONSU) is used as a surrogate marker of intracranial pressure (ICP), especially in resource-limited settings. There is a growing interest in the use of ONSU in emergency and high-altitude setups. Notwithstanding multiple studies done on this subject, there is a paucity of data regarding standardization of techniques and comparison of ONSU with computed tomography (CT). Materials and Methods: Thirty-five patients with a diagnosis of high-altitude cerebral edema were enrolled in the study. ONSD was measured in all patients using ONSU, along visual and coronal axis, and CT scan. We repeated ONSU in these patients on days 3, 7, 10, and 15 (day of discharge). Correlation between visual and coronal axis as well as CT scan was analyzed. Results: The correlation of visual to coronal and coronal to visual was equally significant (both correlation coefficients being R-2 = 0.983). Correlation of ONSD by visual axis to CT scan was better than coronal axis (correlation coefficient R-2 = 0.986 vs. 0.96, respectively). Conclusion: In our study, we found a strong correlation between the visual and coronal axes. Thus, either of the two axes can be used for monitoring ICP. However, it has been found that measurements along the coronal axis are challenging, especially in the emergency setup. ONSD measured along visual axis correlated better with CT scan as compared to the coronal axis.
引用
收藏
页码:252 / 255
页数:4
相关论文
共 50 条
  • [31] Accuracy of bedside sonographic measurement of optic nerve sheath diameter for intracranial hypertension diagnosis in the emergency department
    Busti, Chiara
    Marcosignori, Matteo
    Marchetti, Francesco
    Batori, Giuseppe
    Giovenali, Laura
    Corea, Francesco
    Calabro, Giuseppe
    Monti, Manuel
    Germini, Federico
    EMERGENCY CARE JOURNAL, 2023, 19 (02)
  • [32] ULTRASONOGRAPHY OF OPTIC NERVE SHEATH DIAMETER (ONSD) FOR RAISED INTRACRANIAL PRESSURE
    Siddiqui, Naveed Ur Rehman
    CRITICAL CARE MEDICINE, 2018, 46 (01) : 397 - 397
  • [33] Optic nerve sheath diameter ultrasonography in differentiation of ischemic and hemorrhagic strokes
    De Bernardo, Maddalena
    Vitiello, Livio
    Rosa, Nicola
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2019, 37 (07): : 1384 - 1385
  • [34] Optic nerve sheath diameter ultrasonography for elevated intracranial pressure detection
    Wang Li-Juan
    Chen Hong-Xiu
    Chen Ying
    Yu Ze-Yang
    Xing Ying-Qi
    ANNALS OF CLINICAL AND TRANSLATIONAL NEUROLOGY, 2020, 7 (05): : 865 - 868
  • [35] The relationship between computed tomography measurement of the optic nerve sheath diameter and elevated intracranial pressure in non-trauma patients
    Luyt, Daniel
    Hurter, Delme
    Joubert, Gina
    SA JOURNAL OF RADIOLOGY, 2016, 20 (01):
  • [36] Ultrasonographic measurement of the optic nerve sheath diameter in dysthyroid optic neuropathy
    Xian Ji
    Wei Xiao
    Huijing Ye
    Rongxin Chen
    Jinghong Wu
    Yuxiang Mao
    Huasheng Yang
    Eye, 2021, 35 : 568 - 574
  • [37] Ultrasonographic measurement of the optic nerve sheath diameter in dysthyroid optic neuropathy
    Ji, Xian
    Xiao, Wei
    Ye, Huijing
    Chen, Rongxin
    Wu, Jinghong
    Mao, Yuxiang
    Yang, Huasheng
    EYE, 2021, 35 (02) : 568 - 574
  • [38] Elevated intracranial pressure detected by bedside emergency ultrasonography of the optic nerve sheath
    Blaivas, M
    Theodoro, D
    Sierzenski, PR
    ACADEMIC EMERGENCY MEDICINE, 2003, 10 (04) : 376 - 381
  • [39] Optic nerve sheath diameter
    Lovrencic-Huzjan, A.
    CEREBROVASCULAR DISEASES, 2015, 39 : 4 - 4
  • [40] Comparative and retrospective evaluation of the predictive performance of optic nerve sheath thickness and optic nerve sheath diameter for traumatic brain injury using facial computed tomography
    Sim, Woo Sung
    Lee, Sun Hwa
    Yun, Seong Jong
    Ryu, Seokyong
    Choi, Seung Woon
    Kim, Hye Jin
    Kang, Tae Kyung
    Oh, Sung Chan
    Cho, Suk Jin
    CLINICAL AND EXPERIMENTAL EMERGENCY MEDICINE, 2020, 7 (02): : 122 - 130