Electromagnetic interference of implantable cardiac devices from a shoulder massage machine

被引:0
|
作者
Saeko Yoshida
Kousaku Fujiwara
Satoshi Kohira
Minoru Hirose
机构
[1] Kitasato University,Department of Medical Safety Engineering, Graduate School of Medical Science
[2] Kitasato University,Clinical Engineering Course, School of Allied Health Science
[3] Kitasato University Hospital,Department of Medical Engineering
来源
关键词
Pacemaker; EMI; Massage machine;
D O I
暂无
中图分类号
学科分类号
摘要
Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter–defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28 cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a “don’t use” warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.
引用
收藏
页码:243 / 249
页数:6
相关论文
共 50 条
  • [21] Electric Cars and Electromagnetic Interference With Cardiac Implantable Electronic Devices: A Cross-sectional Evaluation
    Lennerz, Carsten
    O'Connor, Matthew
    Horlbeck, Lorenz
    Michel, Jonathan
    Weigand, Severin
    Grebmer, Christian
    Blazek, Patrick
    Brkic, Amir
    Semmler, Verena
    Haller, Bernhard
    Reents, Tilko
    Hessling, Gabriele
    Deisenhofer, Isabel
    Whittaker, Peter
    Lienkamp, Markus
    Kolb, Christof
    ANNALS OF INTERNAL MEDICINE, 2018, 169 (05) : 350 - +
  • [22] New test system to assess the risk of electromagnetic interference in cardiac active implantable medical devices
    Stunder, D.
    Kaiser, C.
    Silny, J.
    Joosten, S.
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2012, 57 : 857 - 857
  • [23] Electromagnetic interference with implantable cardiac pacemakers by video capsule
    Dubner, S
    Dubner, Y
    Gallino, S
    Spallone, L
    Zagalsky, D
    Rubio, H
    Zimmerman, J
    Goldin, E
    GASTROINTESTINAL ENDOSCOPY, 2005, 61 (02) : 250 - 254
  • [24] Electromagnetic interference from lasers and intense light sources in the treatment of patients with artificial pacemakers and other implantable cardiac devices
    Lister, Tom
    Grant, Lindsay
    Lee, Siu-Man
    Cole, Richard P.
    Jones, Anthony
    Taylor, Timothy
    Mayo, Angela
    Wright, Philip A.
    LASERS IN MEDICAL SCIENCE, 2015, 30 (05) : 1619 - 1622
  • [25] Electromagnetic interference from lasers and intense light sources in the treatment of patients with artificial pacemakers and other implantable cardiac devices
    Tom Lister
    Lindsay Grant
    Siu-Man Lee
    Richard P. Cole
    Anthony Jones
    Timothy Taylor
    Angela Mayo
    Philip A. Wright
    Lasers in Medical Science, 2015, 30 : 1619 - 1622
  • [26] Investigation of Hyfrecators and Their In Vitro Interference with Implantable Cardiac Devices
    Weyer, Christopher
    Siegle, Ronald J.
    Girard, Guillaume
    DERMATOLOGIC SURGERY, 2012, 38 (11) : 1843 - 1848
  • [27] Electromagnetic Interference in Cardiac Contractility Modulation Devices
    Napp, Andreas
    Zink, Matthias
    Kraus, Thomas
    Marx, Nikolaus
    Stunder, Dominik
    CIRCULATION, 2017, 136
  • [28] Electromagnetic Interference in Cardiac Implantable Electronic Devices Due to Dynamic Wireless Power Systems for Electric Vehicles
    Campi, Tommaso
    Cruciani, Silvano
    Maradei, Francesca
    Feliziani, Mauro
    ENERGIES, 2023, 16 (09)
  • [29] Effects of electromagnetic interference on implanted cardiac devices
    Blitzer, ML
    Schoenfeld, MH
    CARDIAC ARRHYTHMIAS 2003, 2004, : 911 - 918
  • [30] Electromagnetic interference from automobile passive keyless entry in cardiovascular implantable electronic devices
    Prasertwitayakij, Narawudt
    Komolmis, Taradol
    Gunaparn, Siriluck
    Pisespongsa, Cheeranun
    Phrommintikul, Arintaya
    Wongcharoen, Wanwarang
    Nantsupawat, Teerapat
    EXPERT REVIEW OF MEDICAL DEVICES, 2023, 20 (11) : 973 - 977