Electromagnetic interference of implantable cardiac devices from a shoulder massage machine

被引:0
|
作者
Saeko Yoshida
Kousaku Fujiwara
Satoshi Kohira
Minoru Hirose
机构
[1] Kitasato University,Department of Medical Safety Engineering, Graduate School of Medical Science
[2] Kitasato University,Clinical Engineering Course, School of Allied Health Science
[3] Kitasato University Hospital,Department of Medical Engineering
来源
关键词
Pacemaker; EMI; Massage machine;
D O I
暂无
中图分类号
学科分类号
摘要
Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter–defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28 cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a “don’t use” warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.
引用
收藏
页码:243 / 249
页数:6
相关论文
共 50 条
  • [41] Electromagnetic interference with cardiac pacemakers and implantable cardioverter-defibrillators from low-frequency electromagnetic fields in vivo
    Tiikkaja, Maria
    Aro, Aapo L.
    Alanko, Tommi
    Lindholm, Harri
    Sistonen, Heli
    Hartikainen, Juha E. K.
    Toivonen, Lauri
    Juutilainen, Jukka
    Hietanen, Maila
    EUROPACE, 2013, 15 (03): : 388 - 394
  • [42] Prevalence of electromagnetic interference from left ventricular assist devices in patients with implantable cardioverter defibrillator/pacemakers
    Kooij, C.
    Yalcin, Y. C.
    Theuns, D. A. M. J.
    Constantinescu, A. A.
    Brugts, J. J.
    Manintveld, O. C.
    Yap, S. C.
    Szili-Torok, T.
    Bogers, A. J. J. C.
    Caliskan, K.
    EUROPEAN HEART JOURNAL, 2019, 40 : 3306 - 3306
  • [43] Electromagnetic interference from left ventricular assist devices detected in patients with implantable cardioverter-defibrillators
    Sheldon, Seth H.
    Jazayeri, Mohammad-Ali
    Pierpoline, Michael
    Mohammed, Moghniuddin
    Parikh, Valay
    Robinson, Alexander
    Noheria, Amit
    Haglund, Nicholas
    Sauer, Andrew J.
    Reddy, Y. Madhu
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (01) : 93 - 101
  • [44] Electromagnetic interference from left ventricular assist devices in patients with subcutaneous implantable cardioverter-defibrillators
    Ishida, Yuji
    Payne, Joshua E.
    Field, Michael E.
    Gold, Michael R.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (05) : 1195 - 1201
  • [45] Implantable devices in the electromagnetic environment
    Santini, Luca
    Forleo, Giovanni B.
    Santini, Massimo
    JOURNAL OF ARRHYTHMIA, 2013, 29 (06) : 325 - 333
  • [46] A setup for checking electromagnetic interference between implantable cardiac pacemaker and RFID UHF
    Tranfaglia, R.
    Sansone, M.
    Santise, A.
    Mirarchi, L.
    Bracale, M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 12, 2009, 25 (12): : 243 - +
  • [47] X-ray Radiation Causes Electromagnetic Interference in Implantable Cardiac Pacemakers
    Hirose, Minoru
    Tachikawa, Keiichi
    Ozaki, Masanori
    Umezawa, Naoki
    Shinbo, Toshihiro
    Kokubo, Kenichi
    Kobayashi, Hirosuke
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (10): : 1174 - 1181
  • [48] Effect of ElectroMagnetic interference from SmartPHone on cardiac ImplaNtable electronic device (EMI-PHONE study)
    Apakuppakul, Sanatcha
    Methachittiphan, Nilubon
    Apiyasawat, Sirin
    JOURNAL OF ARRHYTHMIA, 2022, 38 (05) : 778 - 782
  • [49] Video Capsule Endoscopy in Patients With Implantable Cardiac Devices: Safety, Interference and Beyond
    Nassani, Najib
    Tabet, Rabih
    Karam, Boutros
    Akhrass, Philippe
    Deeb, Liliane
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2017, 112 : S260 - S260
  • [50] Interference and implantable cardiac stimulators
    Corbucci, G
    Marton, F
    Riva, U
    Sciotto, F
    Venturini, D
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2004, 20 (01): : 3 - 12