Evaluation of surface dose and image quality using the half-scan mode in chest computed tomography-guided interventional radiology: a phantom study

被引:1
|
作者
Hasegawa H. [1 ]
Sato J. [1 ]
Kobayashi I. [2 ]
机构
[1] Department of Radiology, The University of Tokyo Hospital, 7-3-1, Hongo, Bunkyo-ku, Tokyo
[2] Nagase-Landauer, Ltd., Block C22-1, Suwa, Tsukuba, 300-2686, Ibaraki
关键词
Computed tomography; Half-scan; Interventional radiology; Optically stimulated luminescent dosimeter; Physicians’ radiation exposure; Radiation protection;
D O I
10.1007/s12194-018-0445-6
中图分类号
学科分类号
摘要
The authors aimed to evaluate the effects of the half-scan mode on image quality and physician exposure to radiation in computed tomography (CT)-guided interventional radiology (IVR) to the right lung using an intermittent CT fluoroscopy technique for measuring phantom surface dose distribution and image noise. For the half-scan mode, settings at 0°, 90°, 180°, and 270° were used as the central axis of the X-ray exposure range on the chest phantom. With the center of the ventral side in the chest phantom defined as 0°, optically stimulated luminescent dosimeters were attached at five positions at 30° intervals on the right side of the phantom surface. Securing a space for device operation during the procedure is necessary. The couch was shifted downward by 50 mm to reproduce the conditions used for measurement in clinical settings. Image noise and contrast-to-noise ratio were measured to assess image quality; subjective evaluation was performed using simulated lung nodules placed in the phantom. The phantom surface dose distribution in the measured half-scan mode depended on the angle setting. Additionally, the phantom surface dose in the half-scan mode at the 90° setting was reduced by approximately 50%; however, image quality was clearly decreased. In CT-guided IVR to the right lung, using a lead drape and half-scan mode according to the procedural situation is important. © 2018, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:138 / 148
页数:10
相关论文
共 50 条
  • [31] Evaluation of image quality and radiation dose using gold nanoparticles and other clinical contrast agents in dual-energy Computed Tomography (CT): CT abdomen phantom
    Zukhi, J.
    Yusob, D.
    Tajuddin, A. A.
    Vuanghao, L.
    Zainon, R.
    [J]. 10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [32] Comparison of Calcium Scoring Between 64-Multidetector Computed Tomography and 320-Multidetector Computed Tomography Using a Cardiac Phantom: Achieving Consistent Image Quality With Dose Optimization
    Tsai, Sung-Yi
    Ding, Xiao-Hua
    Jin, Shih-Chun
    Yi, Dong-Bo
    Chen, Jyh-Cheng
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2021, 45 (01) : 73 - 77
  • [33] The investigation of dose and image quality of chest computed tomography using different combinations of noise index and adaptive statistic iterative reconstruction level
    Sookpeng, Supawitoo
    Martin, Colin J.
    Butdee, Chitsanupong
    [J]. INDIAN JOURNAL OF RADIOLOGY AND IMAGING, 2019, 29 (01): : 53 - 60
  • [34] Evaluating the effect of reduced entrance surface dose on neonatal chest imaging using subjective image quality evaluation
    Conradie, Anita
    Herbst, Charles Petrus
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (10): : 1368 - 1374
  • [35] A cross-sectional study on the correlation of image quality, effective dose, and body composition with thyroid, chest, and abdominal computed tomography scans
    Lin, Chengxiang
    Shen, Xigang
    Gu, Yajia
    Qiao, Zhongwei
    Peng, Weijun
    [J]. QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (06) : 4031 - 4040
  • [36] Evaluation of Organ Dose and Image Quality Metrics of Pediatric CT Chest-Abdomen-Pelvis (CAP) Examination: An Anthropomorphic Phantom Study
    Muhammad, Nor Azura
    Kayun, Zunaide
    Abu Hassan, Hasyma
    Wong, Jeannie Hsiu Ding
    Ng, Kwan Hoong
    Karim, Muhammad Khalis Abdul
    [J]. APPLIED SCIENCES-BASEL, 2021, 11 (05): : 1 - 13
  • [37] Dose-optimised chest computed tomography for diagnosis of Coronavirus Disease 2019 (COVID-19) - Evaluation of image quality and diagnostic impact
    Steuwe, Andrea
    Rademacher, Christin
    Valentin, Birte
    Koehler, Marie-Helen
    Appel, Elisabeth
    Keitel, Verena
    Timm, Joerg
    Antoch, Gerald
    Aissa, Joel
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (03) : 877 - 891
  • [38] Evaluation of Ultra-low-dose (ULD) Lung Computed Tomography (CT) Using Deep-learning: A Phantom Study
    Kim, Daehong
    Son, Kihong
    Baek, Cheol-Ha
    Jeon, Pil-Hyun
    Lee, Sooyeul
    [J]. JOURNAL OF MAGNETICS, 2021, 26 (04) : 429 - 436
  • [39] Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique
    Gandhi, Diksha
    Crotty, Dominic J.
    Stevens, Grant M.
    Schmidt, Taly Gilat
    [J]. MEDICAL PHYSICS, 2015, 42 (11) : 6572 - 6578
  • [40] Assessment of the image quality of virtual monochromatic spectral computed tomography images: a phantom study considering object contrast, radiation dose, and frequency characteristics
    Masuda, Shota
    Sugisawa, Koichi
    Minamishima, Kazuya
    Yamazaki, Akihisa
    Jinzaki, Masahiro
    [J]. RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (01) : 41 - 49