Evaluating the effect of reduced entrance surface dose on neonatal chest imaging using subjective image quality evaluation

被引:12
|
作者
Conradie, Anita [1 ]
Herbst, Charles Petrus [1 ]
机构
[1] Univ Free State, Dept Med Phys, ZA-9300 Bloemfontein, South Africa
来源
关键词
Neonatal; Chest X-rays; Dose; Rabbit phantom; Image quality; INTENSIVE-CARE-UNIT; RADIATION-EXPOSURE; RADIOGRAPHY; OPTIMIZATION; REDUCTION;
D O I
10.1016/j.ejmp.2016.07.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: As hospitalized neonates are radiosensitive and often require numerous X-rays, institutions should investigate the optimal beam parameter combinations to deliver diagnostically acceptable quality images at the lowest possible entrance surface dose (ESD). Using a subjective approach, this study evaluated the effect of different beam parameter combinations on image quality. Methods and materials: Five rabbits simulated the neonatal chest. The ESD was reduced using a variation of voltage, tube current and filtration. Eight radiology registrars, blinded to the dose parameter information, ranked the digital X-ray images of three anatomical regions from best to worst using a variation of the multiple rank order method. T-tests compared the average values, obtained from the scores assigned by each observer to images acquired at different ESDs, for each region. The calculated intraclass correlation coefficient (ICC) assessed observer agreement. Results: Results showed that a 64% dose reduction was achievable by altering the beam parameters. Large variation among the observers was confirmed by an ICC value <0.5. A 95% confidence interval could not conclude that different ESD values, resulting in a 50-77% dose reduction compared to current practice, would result in different overall observed image quality. This was noted for all three regions indicating that no preference existed towards an image acquired with a specific beam parameter combination. Conclusions: The large variation in observers' opinion of acceptable image quality emphasizes the importance of subjective image quality evaluation in the clinical environment. The rabbit phantom and multiple rank order method are considered appropriate for these evaluations. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1368 / 1374
页数:7
相关论文
共 50 条
  • [1] Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems
    Sun, Zhonghua
    Lin, Chenghsun
    Tyan, YeuSheng
    Ng, Kwan-Hoong
    CLINICAL IMAGING, 2012, 36 (04) : 279 - 286
  • [2] Computed radiography for neonatal chest imaging: The effect of image processing on the perception of image quality
    Don, S
    Whiting, BR
    Ellinwood, JS
    Foos, DH
    Kronemer, KA
    Kraus, RA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (04) : 37 - 37
  • [3] Chest Computed Radiography Imaging Parameters for Nodule Detection: A Comparison of Image Quality and Entrance Skin Dose
    Sun, Zhonghua
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2014, 4 (04) : 587 - 592
  • [4] Standard-dose and 50%-reduced-dose chest CT: Comparing the effect on image quality
    Prasad, SR
    Wittram, C
    Shepard, JA
    McLoud, T
    Rhea, J
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (02) : 461 - 465
  • [5] Optimisation of radiation dose and image quality in mobile neonatal chest radiography
    Hinojos-Armendariz, V. I.
    Mejia-Rosales, S. J.
    Franco-Cabrera, M. C.
    RADIOGRAPHY, 2018, 24 (02) : 104 - 109
  • [6] Subjective image quality evaluation in security imaging systems
    Klíma, M
    Páta, P
    Fliegel, KF
    Hanzlík, P
    39TH ANNUAL 2005 INTERNATIONAL CARNAHAN CONFERENCE ON SECURITY TECHNOLOGY, PROCEEDINGS, 2005, : 19 - 22
  • [7] The practice of chest radiography using different digital imaging systems: Dose and image quality
    Moey S.-F.
    Shazli Z.A.
    Sayed I.S.
    Iranian Journal of Medical Physics, 2018, 15 (02) : 101 - 107
  • [8] Image quality and entrance surface dose evaluation of lateral cervical spine: A study using grid and non-grid techniques
    Moey S.F.
    Ramlee N.N.B.M.
    Iranian Journal of Medical Physics, 2019, 16 (02) : 166 - 170
  • [9] Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit
    Kim, Tae-Hoon
    Ryu, Jong-Hyun
    Jeong, Chang-Won
    Jun, Hong Young
    Heo, Dong-Woon
    Lee, Seung Hyun
    Oh, Yeon Kyun
    Lee, Mi-Jung
    Yoon, Kwon-Ha
    CLINICAL IMAGING, 2017, 42 : 165 - 171
  • [10] Optimising image quality and radiation dose for neonatal incubator imaging
    Tugwell-Allsup, J.
    Morris, R. W.
    Hibbs, R.
    England, A.
    RADIOGRAPHY, 2020, 26 (04) : E258 - E263