Evaluation of surface dose and image quality using the half-scan mode in chest computed tomography-guided interventional radiology: a phantom study

被引:1
|
作者
Hasegawa H. [1 ]
Sato J. [1 ]
Kobayashi I. [2 ]
机构
[1] Department of Radiology, The University of Tokyo Hospital, 7-3-1, Hongo, Bunkyo-ku, Tokyo
[2] Nagase-Landauer, Ltd., Block C22-1, Suwa, Tsukuba, 300-2686, Ibaraki
关键词
Computed tomography; Half-scan; Interventional radiology; Optically stimulated luminescent dosimeter; Physicians’ radiation exposure; Radiation protection;
D O I
10.1007/s12194-018-0445-6
中图分类号
学科分类号
摘要
The authors aimed to evaluate the effects of the half-scan mode on image quality and physician exposure to radiation in computed tomography (CT)-guided interventional radiology (IVR) to the right lung using an intermittent CT fluoroscopy technique for measuring phantom surface dose distribution and image noise. For the half-scan mode, settings at 0°, 90°, 180°, and 270° were used as the central axis of the X-ray exposure range on the chest phantom. With the center of the ventral side in the chest phantom defined as 0°, optically stimulated luminescent dosimeters were attached at five positions at 30° intervals on the right side of the phantom surface. Securing a space for device operation during the procedure is necessary. The couch was shifted downward by 50 mm to reproduce the conditions used for measurement in clinical settings. Image noise and contrast-to-noise ratio were measured to assess image quality; subjective evaluation was performed using simulated lung nodules placed in the phantom. The phantom surface dose distribution in the measured half-scan mode depended on the angle setting. Additionally, the phantom surface dose in the half-scan mode at the 90° setting was reduced by approximately 50%; however, image quality was clearly decreased. In CT-guided IVR to the right lung, using a lead drape and half-scan mode according to the procedural situation is important. © 2018, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:138 / 148
页数:10
相关论文
共 50 条
  • [41] Assessment of the image quality of virtual monochromatic spectral computed tomography images: a phantom study considering object contrast, radiation dose, and frequency characteristics
    Shota Masuda
    Koichi Sugisawa
    Kazuya Minamishima
    Akihisa Yamazaki
    Masahiro Jinzaki
    [J]. Radiological Physics and Technology, 2021, 14 : 41 - 49
  • [42] Predictability and image quality of low-dose cone-beam computed tomography in computer-guided implantology: An experimental study
    Horsch, Laura
    Labis, Christos
    Trebing, Charlotte Theresa
    Kronsteiner, Dorothea
    Hilgenfeld, Tim
    Rammelsberg, Peter
    Schwindling, Franz Sebastian
    [J]. JOURNAL OF DENTISTRY, 2021, 112
  • [43] Dose reduction and image quality improvement of chest radiography by using bone-suppression technique and low tube voltage: a phantom study
    Takagi, Satoshi
    Yaegashi, Tatsuya
    Ishikawa, Masayori
    [J]. EUROPEAN RADIOLOGY, 2020, 30 (01) : 571 - 580
  • [44] Dose reduction and image quality improvement of chest radiography by using bone-suppression technique and low tube voltage: a phantom study
    Satoshi Takagi
    Tatsuya Yaegashi
    Masayori Ishikawa
    [J]. European Radiology, 2020, 30 : 571 - 580
  • [45] Effect of x-ray tube parameters, iodine concentration, and patient size on image quality in pulmonary computed tomography angiography: A chest-phantom-study
    Szucs-Farkas, Zsolt
    Verdun, Francis R.
    von Allmen, Gabriel
    Mini, Roberto L.
    Vock, Peter
    [J]. INVESTIGATIVE RADIOLOGY, 2008, 43 (06) : 374 - 381
  • [46] A qualitative study of improving megavoltage computed tomography image quality and maintaining dose accuracy using cycleGAN-based image synthesis
    Lv, Tie
    Xie, Chuanbin
    Zhang, Yihang
    Liu, Yaoying
    Zhang, Gaolong
    Qu, Baolin
    Zhao, Wei
    Xu, Shouping
    [J]. MEDICAL PHYSICS, 2024, 51 (01) : 394 - 406
  • [47] Trade-off between the radiation parameters and image quality using iterative reconstruction techniques in head computed tomography: a phantom study
    Lace, Eliza
    Mohammadian, Reza
    Aboltins, Ainars
    Sosars, Davis
    Apine, Ilze
    [J]. ACTA RADIOLOGICA, 2023, 64 (09) : 2618 - 2626
  • [48] Evaluation of efficacy of transcatheter arterial chemoembolization combined with computed tomography-guided radiofrequency ablation for hepatocellular carcinoma using magnetic resonance diffusion weighted imaging and computed tomography perfusion imaging A prospective study
    Shao, Guo-Liang
    Zheng, Jia-Ping
    Guo, Li-Wen
    Chen, Yu-Tang
    Zeng, Hui
    Yao, Zheng
    [J]. MEDICINE, 2017, 96 (03)
  • [49] Dose evaluation for foetal computed tomography with a 320-row unit in wide-volume mode and an 80-row unit in helical scanning mode: a phantom study
    Matsunaga, Yuta
    Kawaguchi, Ai
    Kobayashi, Masanao
    Suzuki, Shigetaka
    Suzuki, Shoichi
    Chida, Koichi
    [J]. RADIATION PROTECTION DOSIMETRY, 2016, 168 (04) : 523 - 530
  • [50] Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: a phantom study
    Greffier, Joel
    Dabli, Djamel
    Hamard, Aymeric
    Belaouni, Asmaa
    Akessoul, Philippe
    Frandon, Julien
    Beregi, Jean-Paul
    [J]. QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (01) : 229 - +