A Fluorescence Immunochromatographic Assay Using Europium (III) Chelate Microparticles for Rapid, Quantitative and Sensitive Detection of Creatine Kinase MB

被引:0
|
作者
Xiao-Hong Lai
Rong-Liang Liang
Tian-Cai Liu
Zhi-Ning Dong
Ying-Song Wu
Lin-Hai Li
机构
[1] Guangzhou General Hospital of Guangzhou Military Command of PLA,Department of Laboratory Medicine
[2] Guangzhou University of Chinese Medicine,State Key Laboratory of Organ Failure, Guangdong Provincial Key Laboratory of Tropical Disease Research, School of Biotechnology
[3] Southern Medical University,undefined
来源
Journal of Fluorescence | 2016年 / 26卷
关键词
Acute myocardial infarction; Creatine kinase MB; Eu (III) chelate microparticles; Fluorescence immunochromatographic assay;
D O I
暂无
中图分类号
学科分类号
摘要
The isoenzyme creatine kinase MB is very important for diagnosis of acute myocardial infarction (AMI). Some CK-MB immunoassays are sensitive, accurate and available for clinical application, but they are expensive and time-consuming procedures. Furthermore, conventional fluorescence immunochromatographic assays (FL-ICAs) have suffered from background fluorescence interference and low analytical sensitivity. A rapid and simple FL-ICA with Eu (III) chelate polystyrene microparticles was developed to determine CK-MB in 50uL serum samples using a portable test strip reader by measuring the fluorescence peak heights of the test line (HT) and the control line (HC) in 12 min. The assay was reliable with a good correlation coefficient between HT/HC ratio and CK-MB concentration in samples. A linear range was 0.85–100.29 ng/mL for CK-MB, and the LOD was 0.029 ng/mL. The intra- and inter-assay coefficients of variation (CV) were both <10 % and the average recoveries were from 90.17 % -112.63 % for CK-MB. The system performed well in interference experiments. Furthermore, a highly significant correlation (r = 0.9794, P < 0.001) between this method and the commercially available bioMérieux mini VIDAS system were attained for measuring 120 CK-MB samples. These results indicated that the Eu (III) chelate microparticles-based FL-ICA is simple, fast, highly sensitive, reliable, and reproducible for point-of-care testing of CK-MB concentrations in serum.
引用
收藏
页码:987 / 996
页数:9
相关论文
共 47 条
  • [41] Immunochromatographic assay for quantitative and sensitive detection of hepatitis B virus surface antigen using highly luminescent quantum dot-beads
    Shen, Jun
    Zhou, Yaofeng
    Fu, Fen
    Xu, Hengyi
    Lv, Jiaofeng
    Xiong, Yonghua
    Wang, Andrew
    TALANTA, 2015, 142 : 145 - 149
  • [42] Development of ZnCdSe/ZnS quantum dot-based fluorescence immunochromatographic assay for the rapid visual and quantitative detection 25-hydroxyvitamins D in human serum
    Wang, Jianfa
    Sun, Guoshao
    Li, Fang
    Zhu, Zhi
    Sun, Lei
    Lv, Pengju
    Yue, Han
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [43] Rapid and quantitative detection of neutrophil gelatinase-associated lipocalin in synovial fluid using fluorescence immunochromatographic test strips for diagnosing prosthetic joint infections
    Zhang, Zijie
    Huang, Ying
    Li, Mengqing
    Li, Wenbo
    Fang, Xinyu
    Huang, Zida
    Zhang, Wenming
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2023, 106 (04)
  • [44] A fluorescence-displacement assay using molecularly imprinted polymers for the visual, rapid, and sensitive detection of the algal metabolites, geosmin and 2-methylisoborneol
    Li, Cheng
    Mun Hong Ngai
    Reddy, Karnati Konda
    Leong, Sandric Chee Yew
    Tong, Yen Wah
    Chai, Christina Li Lin
    ANALYTICA CHIMICA ACTA, 2019, 1066 (121-130) : 121 - 130
  • [45] Development of a Rapid, Sensitive, and Field-Deployable Razor Ex BioDetection System and Quantitative PCR Assay for Detection of Phymatotrichopsis omnivora Using Multiple Gene Targets
    Arif, M.
    Fletcher, J.
    Marek, S. M.
    Melcher, U.
    Ochoa-Corona, F. M.
    APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2013, 79 (07) : 2312 - 2320
  • [46] Rapid, quantitative, and high-sensitivity detection of anti-phospholipase A2 receptor antibodies using a novel CdSe/ZnS-based fluorescence immunosorbent assay
    Chenxi Li
    Manyun Qian
    Qiaozhen Hong
    Xiaohong Xin
    Zichun Sun
    Yafeng Li
    Bo Tang
    Bing Gu
    Scientific Reports, 11
  • [47] Rapid, quantitative, and high-sensitivity detection of anti-phospholipase A2 receptor antibodies using a novel CdSe/ZnS-based fluorescence immunosorbent assay
    Li, Chenxi
    Qian, Manyun
    Hong, Qiaozhen
    Xin, Xiaohong
    Sun, Zichun
    Li, Yafeng
    Tang, Bo
    Gu, Bing
    SCIENTIFIC REPORTS, 2021, 11 (01)