Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system

被引:0
|
作者
Theresa V. Feddersen
Pejman Rowshanfarzad
Tamara N. Abel
Martin A. Ebert
机构
[1] University of Western Australia,Department of Physics
[2] University of Western Australia,Telethon Kids Institute
[3] Sir Charles Gairdner Hospital,Department of Radiation Oncology
[4] Erasmus University Medical Center,Department of Radiotherapy
[5] Erasmus University Medical Center,Department of Radiology & Nuclear Medicine
关键词
Small animal radiotherapy; Pre-clinical; Commissioning; Dose delivery; Image guidance;
D O I
暂无
中图分类号
学科分类号
摘要
Characteristics of a small-animal radiotherapy device, the X-RAD SmART, are described following commissioning of the device for pre-clinical radiotherapy research. Performance characteristics were assessed using published standards and compared with previous results published for similar systems. Operational radiation safety was established. Device X-ray beam quality and output dose-rate were found to be consistent with those reported for similar devices. Output steadily declined over 18 months though remained within tolerance levels. There is considerable variation in output factor across the international installations for the smallest field size (varying by more than 30% for 2.5 mm diameter fields). Measured depth dose and profile data was mostly consistent with that published, with some differences in penumbrae and generally reduced flatness. Target localisation is achieved with an imaging panel and with automatic corrections for panel flex and device mechanical instability, targeting within 0.2 mm is achievable. The small-animal image-guided radiotherapy platform has been implemented and assessed and found to perform as specified. The combination of kV energy and high spatial precision make it suitable for replicating clinical dose distributions at the small-animal scale, though dosimetric uncertainties for the narrowest fields need to be acknowledged.
引用
收藏
页码:541 / 551
页数:10
相关论文
共 50 条
  • [31] The future of image-guided radiotherapy will be MR guided
    Pollard, Julianne M.
    Wen, Zhifei
    Sadagopan, Ramaswamy
    Wang, Jihong
    Ibbott, Geoffrey S.
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1073):
  • [32] Clinical Application of Image-guided Radiotherapy in Bladder and Prostate Cancer
    Button, M. R.
    Staffurth, J. N.
    CLINICAL ONCOLOGY, 2010, 22 (08) : 698 - 706
  • [33] Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator
    Muralidhar, K. R.
    Murthy, P. Narayana
    Kumar, Rajneesh
    JOURNAL OF MEDICAL PHYSICS, 2008, 33 (02) : 72 - 77
  • [34] Treatment time of image-guided radiotherapy with a Halcyon 2.0 system
    Hermida-Lopez, Marcelino
    Garcia-Relancio, David
    Comino-Muriano, Marina
    Perez-Esteve, Brenda
    Castillo-Elias, Esther
    Carrera-Garcia, Yolanda
    Giralt, Jordi
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2023, 54 (01) : 117 - 122
  • [35] Physics Support of Image-guided Radiotherapy
    Eaton, D. J.
    CLINICAL ONCOLOGY, 2013, 25 (03) : E31 - E31
  • [36] Small animal image-guided radiotherapy
    Verhaegen, Frank
    Vaz, Pedro
    Prise, Kevin M.
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1095):
  • [37] Image-guided radiotherapy for gynaecological cancer
    Lucia, F.
    Schick, U.
    Pradier, O.
    Espenel, S.
    de Crevoisier, R.
    Chargari, C.
    CANCER RADIOTHERAPIE, 2018, 22 (6-7): : 608 - 616
  • [38] Recent advances in image-guided radiotherapy
    Mohan, R
    Dong, L
    Zhang, XD
    RADIOTHERAPY AND ONCOLOGY, 2006, 78 : S27 - S28
  • [39] Erratum: Innovations in image-guided radiotherapy
    Dirk Verellen
    Mark De Ridder
    Nadine Linthout
    Koen Tournel
    Guy Soete
    Guy Storme
    Nature Reviews Cancer, 2008, 8 : 71 - 71
  • [40] Image-guided motion adaptation in radiotherapy
    Murphy, Martin J.
    2007 4TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING : MACRO TO NANO, VOLS 1-3, 2007, : 996 - 999