Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system

被引:0
|
作者
Theresa V. Feddersen
Pejman Rowshanfarzad
Tamara N. Abel
Martin A. Ebert
机构
[1] University of Western Australia,Department of Physics
[2] University of Western Australia,Telethon Kids Institute
[3] Sir Charles Gairdner Hospital,Department of Radiation Oncology
[4] Erasmus University Medical Center,Department of Radiotherapy
[5] Erasmus University Medical Center,Department of Radiology & Nuclear Medicine
关键词
Small animal radiotherapy; Pre-clinical; Commissioning; Dose delivery; Image guidance;
D O I
暂无
中图分类号
学科分类号
摘要
Characteristics of a small-animal radiotherapy device, the X-RAD SmART, are described following commissioning of the device for pre-clinical radiotherapy research. Performance characteristics were assessed using published standards and compared with previous results published for similar systems. Operational radiation safety was established. Device X-ray beam quality and output dose-rate were found to be consistent with those reported for similar devices. Output steadily declined over 18 months though remained within tolerance levels. There is considerable variation in output factor across the international installations for the smallest field size (varying by more than 30% for 2.5 mm diameter fields). Measured depth dose and profile data was mostly consistent with that published, with some differences in penumbrae and generally reduced flatness. Target localisation is achieved with an imaging panel and with automatic corrections for panel flex and device mechanical instability, targeting within 0.2 mm is achievable. The small-animal image-guided radiotherapy platform has been implemented and assessed and found to perform as specified. The combination of kV energy and high spatial precision make it suitable for replicating clinical dose distributions at the small-animal scale, though dosimetric uncertainties for the narrowest fields need to be acknowledged.
引用
收藏
页码:541 / 551
页数:10
相关论文
共 50 条
  • [21] Image-guided and adaptive radiotherapy
    Louvel, G.
    Cazoulat, G.
    Chajon, E.
    Le Maitre, A.
    Simon, A.
    Henry, O.
    Bensadoun, R. J.
    de Crevoisier, R.
    CANCER RADIOTHERAPIE, 2012, 16 (5-6): : 423 - 429
  • [22] The Future of Image-guided Radiotherapy
    Choudhury, A.
    Budgell, G.
    MacKay, R.
    Falk, S.
    Faivre-Finn, C.
    Dubec, M.
    van Herk, M.
    McWilliam, A.
    CLINICAL ONCOLOGY, 2017, 29 (10) : 662 - 666
  • [23] Commissioning experience of the X-ray volume imaging system of an image-guided radiotherapy capable linear accelerator
    Raina, Payal
    Sahoo, Susanta Kumar
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (01) : 108 - 113
  • [24] Innovations in image-guided radiotherapy
    Verellen, Dirk
    De Ridder, Mark
    Linthout, Nadine
    Tournel, Koen
    Soete, Guy
    Storme, Guy
    NATURE REVIEWS CANCER, 2007, 7 (12) : 949 - 960
  • [25] A review of image-guided radiotherapy
    Chen G.T.Y.
    Sharp G.C.
    Mori S.
    Radiological Physics and Technology, 2009, 2 (1) : 1 - 12
  • [26] Implementation of Image-guided Radiotherapy
    Franks, K. N.
    McNair, H. A.
    CLINICAL ONCOLOGY, 2012, 24 (09) : 625 - 626
  • [27] Image-guided Gated Radiotherapy
    Heaton, A.
    Shakeshaft, J.
    CLINICAL ONCOLOGY, 2009, 21 (03) : 268 - 268
  • [28] Innovations in image-guided radiotherapy
    Dirk Verellen
    Mark De Ridder
    Nadine Linthout
    Koen Tournel
    Guy Soete
    Guy Storme
    Nature Reviews Cancer, 2007, 7 : 949 - 960
  • [29] A review of treatment planning for precision image-guided photon beam pre-clinical animal radiation studies
    Verhaegen, Frank
    van Hoof, Stefan
    Granton, Patrick V.
    Trani, Daniela
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2014, 24 (04): : 323 - 334
  • [30] From image-guided radiotherapy to dose-guided radiotherapy
    Cazoulat, G.
    Lesaunier, M.
    Simon, A.
    Haigron, P.
    Acosta, O.
    Louvel, G.
    Lafond, C.
    Chajon, E.
    Leseur, J.
    de Crevoisier, R.
    CANCER RADIOTHERAPIE, 2011, 15 (08): : 691 - 698