Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system

被引:0
|
作者
Theresa V. Feddersen
Pejman Rowshanfarzad
Tamara N. Abel
Martin A. Ebert
机构
[1] University of Western Australia,Department of Physics
[2] University of Western Australia,Telethon Kids Institute
[3] Sir Charles Gairdner Hospital,Department of Radiation Oncology
[4] Erasmus University Medical Center,Department of Radiotherapy
[5] Erasmus University Medical Center,Department of Radiology & Nuclear Medicine
关键词
Small animal radiotherapy; Pre-clinical; Commissioning; Dose delivery; Image guidance;
D O I
暂无
中图分类号
学科分类号
摘要
Characteristics of a small-animal radiotherapy device, the X-RAD SmART, are described following commissioning of the device for pre-clinical radiotherapy research. Performance characteristics were assessed using published standards and compared with previous results published for similar systems. Operational radiation safety was established. Device X-ray beam quality and output dose-rate were found to be consistent with those reported for similar devices. Output steadily declined over 18 months though remained within tolerance levels. There is considerable variation in output factor across the international installations for the smallest field size (varying by more than 30% for 2.5 mm diameter fields). Measured depth dose and profile data was mostly consistent with that published, with some differences in penumbrae and generally reduced flatness. Target localisation is achieved with an imaging panel and with automatic corrections for panel flex and device mechanical instability, targeting within 0.2 mm is achievable. The small-animal image-guided radiotherapy platform has been implemented and assessed and found to perform as specified. The combination of kV energy and high spatial precision make it suitable for replicating clinical dose distributions at the small-animal scale, though dosimetric uncertainties for the narrowest fields need to be acknowledged.
引用
收藏
页码:541 / 551
页数:10
相关论文
共 50 条
  • [1] Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system
    Feddersen, Theresa V.
    Rowshanfarzad, Pejman
    Abel, Tamara N.
    Ebert, Martin A.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (02) : 541 - 551
  • [2] Commissioning an image-guided localization system for radiotherapy
    Phillips, MH
    Singer, K
    Miller, E
    Stelzer, K
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (01): : 267 - 276
  • [3] SyncMRT: a solution to image-guided synchrotron radiotherapy for quality assurance and pre-clinical trials
    Barnes, M.J.
    Paino, J.
    Day, L.R.
    Butler, D.
    Häusermann, D.
    Pelliccia, D.
    Crosbie, J.C.
    Journal of Synchrotron Radiation, 2022, 29 : 1074 - 1084
  • [4] SyncMRT: a solution to image-guided synchrotron radiotherapy for quality assurance and pre-clinical trials
    Barnes, M. J.
    Paino, J.
    Day, L. R.
    Butler, D.
    Hausermann, D.
    Pelliccia, D.
    Crosbie, J. C.
    JOURNAL OF SYNCHROTRON RADIATION, 2022, 29 : 1074 - 1084
  • [5] An image-guided irradiator for pre-clinical radiation therapy studies
    Jaffray, D.
    Moseley, D.
    Chow, J.
    Kim, S.
    Ansell, S.
    Wilson, G.
    Chiarot, C.
    MEDICAL PHYSICS, 2006, 33 (06) : 2241 - 2241
  • [6] Commissioning of a KV image-guided system
    Moseley, D.
    White, E.
    Purdie, T.
    Bissonnette, J.
    Sharpe, M.
    Brock, K.
    Siewerdsen, J.
    Jaffray, D.
    MEDICAL PHYSICS, 2006, 33 (06) : 2281 - 2282
  • [7] Intraoperative image-guided transoral robotic surgery: pre-clinical studies
    Liu, Wen P.
    Reaugamornrat, Sureerat
    Sorger, Jonathan M.
    Siewerdsen, Jeffrey H.
    Taylor, Russell H.
    Richmon, Jeremy D.
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2015, 11 (02): : 256 - 267
  • [8] Fluorescence Tomography-Guided System for Pre-Clinical Radiotherapy Research
    Tseng, Y.
    Hardy, L.
    Xu, X.
    Iordachita, I.
    Wong, J.
    Wang, K.
    MEDICAL PHYSICS, 2022, 49 (06) : E563 - E563
  • [9] Image-guided radiotherapy
    de Crevoisier, R.
    Isambert, A.
    Lisbona, A.
    Bodez, V.
    Marguet, M.
    Lafay, F.
    Remonnay, R.
    Lagrange, J. -L.
    CANCER RADIOTHERAPIE, 2007, 11 (6-7): : 296 - 304
  • [10] Image-guided radiotherapy
    de Crevoisier, R.
    Lafond, C.
    Mervoyer, A.
    Hulot, C.
    Jaksic, N.
    Bessieres, I
    Delpon, G.
    CANCER RADIOTHERAPIE, 2022, 26 (1-2): : 34 - 49