Commissioning an image-guided localization system for radiotherapy

被引:32
|
作者
Phillips, MH [1 ]
Singer, K [1 ]
Miller, E [1 ]
Stelzer, K [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Radiat Oncol, Seattle, WA 98195 USA
关键词
image-guided radiation; radiation target localization; positioning; IMRT; conformal therapy; radiosurgery;
D O I
10.1016/S0360-3016(00)00581-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To describe the design and commissioning of a system for the treatment of classes of tumors that require highly accurate target localization during a course of fractionated external-beam therapy. This system uses image-guided localization techniques in the linac vault to position patients being treated for cranial tumors using stereotactic radiotherapy, conformal radiotherapy, and intensity-modulated radiation therapy techniques. Design constraints included flexibility in the use of treatment-planning software, accuracy and precision of repeat localization, limits on the time and human resources needed to use the system, and ease of use. Methods and Materials: A commercially marketed, stereotactic radiotherapy system, based on a system designed at the University of Florida, Gainesville, was adapted for use at the University of Washington Medical Center. A stereo pair of cameras in the linac vault were used to detect the position and orientation of an array of fiducial markers that are attached to a patient's biteblock. The system was modified to allow the use of either a treatment-planning system designed for stereotactic treatments, or a general, three-dimensional radiation therapy planning program. Measurements of the precision and accuracy of the target localization, dose delivery, and patient positioning were made using a number of different jigs and devices. Procedures were developed for the safe and accurate clinical use of the system. Results: The accuracy of the target localization is comparable to that of other treatment-planning systems. Gantry sag, which cannot be improved, was measured to be 1.7 mm, which had the effect of broadening the dose distribution, as confirmed by a comparison of measurement and calculation. The accuracy of positioning a target point in the radiation field was 1.0 +/- 0.2 mm. The calibration procedure using the room-based lasers had an accuracy of 0.76 mm, and using a floor-based radiosurgery system it was 0.73 mm. Target localization error in a phantom was 0.64 +/- 0.77 mm. Errors in positioning due to couch rotation error were reduced using the system. Conclusion: The system described has proven to have acceptable accuracy and precision for the clinical goals for which it was designed. It is robust in detecting errors, and it requires only a nominal increase in setup time and effort. Future work will focus on evaluating its suitability for use in the treatment of head-and-neck cancers not contained within the cranial vault. (C) 2000 Elsevier Science Inc.
引用
下载
收藏
页码:267 / 276
页数:10
相关论文
共 50 条
  • [1] Commissioning of a KV image-guided system
    Moseley, D.
    White, E.
    Purdie, T.
    Bissonnette, J.
    Sharpe, M.
    Brock, K.
    Siewerdsen, J.
    Jaffray, D.
    MEDICAL PHYSICS, 2006, 33 (06) : 2281 - 2282
  • [2] Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system
    Feddersen, Theresa V.
    Rowshanfarzad, Pejman
    Abel, Tamara N.
    Ebert, Martin A.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (02) : 541 - 551
  • [3] Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system
    Theresa V. Feddersen
    Pejman Rowshanfarzad
    Tamara N. Abel
    Martin A. Ebert
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 541 - 551
  • [4] Image-guided radiotherapy
    de Crevoisier, R.
    Isambert, A.
    Lisbona, A.
    Bodez, V.
    Marguet, M.
    Lafay, F.
    Remonnay, R.
    Lagrange, J. -L.
    CANCER RADIOTHERAPIE, 2007, 11 (6-7): : 296 - 304
  • [5] Image-guided radiotherapy
    de Crevoisier, R.
    Lafond, C.
    Mervoyer, A.
    Hulot, C.
    Jaksic, N.
    Bessieres, I
    Delpon, G.
    CANCER RADIOTHERAPIE, 2022, 26 (1-2): : 34 - 49
  • [6] CALCIFICATIONS ARE POTENTIAL SURROGATES FOR PROSTATE LOCALIZATION IN IMAGE-GUIDED RADIOTHERAPY
    Zeng, Grace G.
    McGowan, Tom S.
    Larsen, Tessa M.
    Brue, Lisa M.
    Moran, Natasha K.
    Tsao, Jonathan R.
    MacPherson, Miller S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (04): : 963 - 966
  • [7] Automatic localization of the prostate for on-line image-guided radiotherapy
    Smitsmans, MHP
    Wolthaus, JWH
    Artignan, X
    De Bois, J
    Jaffray, DA
    Lebesque, JV
    Van Herk, M
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S23 - S23
  • [8] MULTIPLE 3D MARKER LOCALIZATION AND TRACKING SYSTEM IN IMAGE-GUIDED RADIOTHERAPY
    Mao, Songan
    Wu, Huanmei
    Lu, Minghui
    Cheng, Chee-Wai
    INTERNATIONAL JOURNAL OF ROBOTICS & AUTOMATION, 2017, 32 (05): : 517 - 523
  • [9] The IMPORT HIGH Image-guided Radiotherapy Study: A Model for Assessing Image-guided Radiotherapy
    Donovan, E. M.
    Harris, E. J.
    Mukesh, M. B.
    Haviland, J. S.
    Titley, J.
    Griffin, C.
    Coles, C. E.
    Evans, P. M.
    CLINICAL ONCOLOGY, 2015, 27 (01) : 3 - 5
  • [10] Positioning accuracy of a new image-guided radiotherapy system
    Miyabe, Yuki
    Sawada, Akira
    Takayama, Kenji
    Kaneko, Shuji
    Mizowaki, Takashi
    Kokubo, Masaki
    Hiraoka, Masahiro
    MEDICAL PHYSICS, 2011, 38 (05) : 2535 - 2541