Commissioning an image-guided localization system for radiotherapy

被引:32
|
作者
Phillips, MH [1 ]
Singer, K [1 ]
Miller, E [1 ]
Stelzer, K [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Radiat Oncol, Seattle, WA 98195 USA
关键词
image-guided radiation; radiation target localization; positioning; IMRT; conformal therapy; radiosurgery;
D O I
10.1016/S0360-3016(00)00581-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To describe the design and commissioning of a system for the treatment of classes of tumors that require highly accurate target localization during a course of fractionated external-beam therapy. This system uses image-guided localization techniques in the linac vault to position patients being treated for cranial tumors using stereotactic radiotherapy, conformal radiotherapy, and intensity-modulated radiation therapy techniques. Design constraints included flexibility in the use of treatment-planning software, accuracy and precision of repeat localization, limits on the time and human resources needed to use the system, and ease of use. Methods and Materials: A commercially marketed, stereotactic radiotherapy system, based on a system designed at the University of Florida, Gainesville, was adapted for use at the University of Washington Medical Center. A stereo pair of cameras in the linac vault were used to detect the position and orientation of an array of fiducial markers that are attached to a patient's biteblock. The system was modified to allow the use of either a treatment-planning system designed for stereotactic treatments, or a general, three-dimensional radiation therapy planning program. Measurements of the precision and accuracy of the target localization, dose delivery, and patient positioning were made using a number of different jigs and devices. Procedures were developed for the safe and accurate clinical use of the system. Results: The accuracy of the target localization is comparable to that of other treatment-planning systems. Gantry sag, which cannot be improved, was measured to be 1.7 mm, which had the effect of broadening the dose distribution, as confirmed by a comparison of measurement and calculation. The accuracy of positioning a target point in the radiation field was 1.0 +/- 0.2 mm. The calibration procedure using the room-based lasers had an accuracy of 0.76 mm, and using a floor-based radiosurgery system it was 0.73 mm. Target localization error in a phantom was 0.64 +/- 0.77 mm. Errors in positioning due to couch rotation error were reduced using the system. Conclusion: The system described has proven to have acceptable accuracy and precision for the clinical goals for which it was designed. It is robust in detecting errors, and it requires only a nominal increase in setup time and effort. Future work will focus on evaluating its suitability for use in the treatment of head-and-neck cancers not contained within the cranial vault. (C) 2000 Elsevier Science Inc.
引用
收藏
页码:267 / 276
页数:10
相关论文
共 50 条
  • [31] Erratum: Innovations in image-guided radiotherapy
    Dirk Verellen
    Mark De Ridder
    Nadine Linthout
    Koen Tournel
    Guy Soete
    Guy Storme
    Nature Reviews Cancer, 2008, 8 : 71 - 71
  • [32] Dose verification for image-guided radiotherapy
    Kapatoes, J
    Ruchala, K
    Olivera, G
    Forrest, L
    Turek, M
    Mackie, T
    Welsh, J
    Jaradat, H
    Balog, J
    Lu, W
    Jeraj, R
    Reckwerdt, P
    MEDICAL PHYSICS, 2003, 30 (06) : 1418 - 1419
  • [33] Image-guided radiotherapy in lung cancer
    Aboudaram, A.
    Khalifa, J.
    Massabeau, C.
    Sirrion, L.
    Henni, A. Hadj
    Thureau, S.
    CANCER RADIOTHERAPIE, 2018, 22 (6-7): : 602 - 607
  • [34] Image-guided motion adaptation in radiotherapy
    Murphy, Martin J.
    2007 4TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING : MACRO TO NANO, VOLS 1-3, 2007, : 996 - 999
  • [35] Good practice of image-guided radiotherapy
    de Crevoisier, R.
    Crehange, G.
    Castelli, J.
    Lafond, C.
    Delpon, G.
    CANCER RADIOTHERAPIE, 2015, 19 (6-7): : 489 - 500
  • [36] Dynamic targeting image-guided radiotherapy
    Huntzinger, C
    Munro, P
    Johnson, S
    Miettinen, M
    Zankowski, C
    Ahlstrom, G
    Glettig, R
    Filliberti, R
    Kaissl, W
    Kamber, M
    Amstutz, M
    Bouchet, L
    Klebanov, D
    Mostafavi, H
    Stark, R
    MEDICAL DOSIMETRY, 2006, 31 (02) : 113 - 125
  • [37] Image-guided radiotherapy for prostate cancer
    Lohr, F.
    Boda-Heggemann, J.
    Wenz, F.
    Wiegel, T.
    AKTUELLE UROLOGIE, 2007, 38 (05) : 386 - 391
  • [38] A (short) history of image-guided radiotherapy
    Verellen, Dirk
    De Ridder, Mark
    Storme, Guy
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 4 - 13
  • [39] Small animal image-guided radiotherapy
    Prise, Kevin M.
    Verhaegen, Frank
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1069):
  • [40] Image-guided radiotherapy for prostate cancer
    Dang, Audrey
    Kupelian, Patrick A.
    Cao, Minsong
    Agazaryan, Nzhde
    Kishan, Amar U.
    TRANSLATIONAL ANDROLOGY AND UROLOGY, 2018, 7 (03) : 308 - 320