Commissioning an image-guided localization system for radiotherapy

被引:32
|
作者
Phillips, MH [1 ]
Singer, K [1 ]
Miller, E [1 ]
Stelzer, K [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Radiat Oncol, Seattle, WA 98195 USA
关键词
image-guided radiation; radiation target localization; positioning; IMRT; conformal therapy; radiosurgery;
D O I
10.1016/S0360-3016(00)00581-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To describe the design and commissioning of a system for the treatment of classes of tumors that require highly accurate target localization during a course of fractionated external-beam therapy. This system uses image-guided localization techniques in the linac vault to position patients being treated for cranial tumors using stereotactic radiotherapy, conformal radiotherapy, and intensity-modulated radiation therapy techniques. Design constraints included flexibility in the use of treatment-planning software, accuracy and precision of repeat localization, limits on the time and human resources needed to use the system, and ease of use. Methods and Materials: A commercially marketed, stereotactic radiotherapy system, based on a system designed at the University of Florida, Gainesville, was adapted for use at the University of Washington Medical Center. A stereo pair of cameras in the linac vault were used to detect the position and orientation of an array of fiducial markers that are attached to a patient's biteblock. The system was modified to allow the use of either a treatment-planning system designed for stereotactic treatments, or a general, three-dimensional radiation therapy planning program. Measurements of the precision and accuracy of the target localization, dose delivery, and patient positioning were made using a number of different jigs and devices. Procedures were developed for the safe and accurate clinical use of the system. Results: The accuracy of the target localization is comparable to that of other treatment-planning systems. Gantry sag, which cannot be improved, was measured to be 1.7 mm, which had the effect of broadening the dose distribution, as confirmed by a comparison of measurement and calculation. The accuracy of positioning a target point in the radiation field was 1.0 +/- 0.2 mm. The calibration procedure using the room-based lasers had an accuracy of 0.76 mm, and using a floor-based radiosurgery system it was 0.73 mm. Target localization error in a phantom was 0.64 +/- 0.77 mm. Errors in positioning due to couch rotation error were reduced using the system. Conclusion: The system described has proven to have acceptable accuracy and precision for the clinical goals for which it was designed. It is robust in detecting errors, and it requires only a nominal increase in setup time and effort. Future work will focus on evaluating its suitability for use in the treatment of head-and-neck cancers not contained within the cranial vault. (C) 2000 Elsevier Science Inc.
引用
收藏
页码:267 / 276
页数:10
相关论文
共 50 条
  • [21] Evaluation of ultrasound-based prostate localization for image-guided radiotherapy
    Langen, KM
    Pouliot, J
    Anezinos, C
    Aubin, M
    Gottschalk, AR
    Hsu, IC
    Lowther, D
    Liu, YM
    Shinohara, K
    Verhey, LJ
    Weinberg, V
    Roach, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (03): : 635 - 644
  • [22] Fluoroscopic Image-Guided Intervention System for Transbronchial Localization
    Rai, Lav
    Keast, Thomas M.
    Wibowo, Henky
    Yu, Kun-Chang
    Draper, Jeffrey W.
    Gibbs, Jason D.
    MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2012, 8316
  • [23] From image-guided radiotherapy to dose-guided radiotherapy
    Cazoulat, G.
    Lesaunier, M.
    Simon, A.
    Haigron, P.
    Acosta, O.
    Louvel, G.
    Lafond, C.
    Chajon, E.
    Leseur, J.
    de Crevoisier, R.
    CANCER RADIOTHERAPIE, 2011, 15 (08): : 691 - 698
  • [24] The future of image-guided radiotherapy will be MR guided
    Pollard, Julianne M.
    Wen, Zhifei
    Sadagopan, Ramaswamy
    Wang, Jihong
    Ibbott, Geoffrey S.
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1073):
  • [25] Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator
    Muralidhar, K. R.
    Murthy, P. Narayana
    Kumar, Rajneesh
    JOURNAL OF MEDICAL PHYSICS, 2008, 33 (02) : 72 - 77
  • [26] Treatment time of image-guided radiotherapy with a Halcyon 2.0 system
    Hermida-Lopez, Marcelino
    Garcia-Relancio, David
    Comino-Muriano, Marina
    Perez-Esteve, Brenda
    Castillo-Elias, Esther
    Carrera-Garcia, Yolanda
    Giralt, Jordi
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2023, 54 (01) : 117 - 122
  • [27] Small animal image-guided radiotherapy
    Verhaegen, Frank
    Vaz, Pedro
    Prise, Kevin M.
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1095):
  • [28] Physics Support of Image-guided Radiotherapy
    Eaton, D. J.
    CLINICAL ONCOLOGY, 2013, 25 (03) : E31 - E31
  • [29] Image-guided radiotherapy for gynaecological cancer
    Lucia, F.
    Schick, U.
    Pradier, O.
    Espenel, S.
    de Crevoisier, R.
    Chargari, C.
    CANCER RADIOTHERAPIE, 2018, 22 (6-7): : 608 - 616
  • [30] Recent advances in image-guided radiotherapy
    Mohan, R
    Dong, L
    Zhang, XD
    RADIOTHERAPY AND ONCOLOGY, 2006, 78 : S27 - S28