Output factor measurements for a kilovoltage x-ray therapy unit

被引:25
|
作者
Healy B.J. [1 ]
Gibbs A. [1 ]
Murry R.L. [1 ]
Prunster J.E. [1 ]
Nitschke K.N. [1 ]
机构
[1] Southern Zone Radiation Oncology Service - Mater Centre, Princess Alexandra Hospital, South Brisbane, QLD 4101, Raymond Tce.
关键词
Ionisation chambers; Kilovoltage x-rays; Output factors; Water substitutes;
D O I
10.1007/BF03178702
中图分类号
学科分类号
摘要
Output factors at the surface for treatment cones and lead cut-outs have been measured for a Pantak Therapax SXT 150 superficial therapy unit with x-ray beam qualities from 1 to 13 min A1 HVL. A variety of phantom materials and two ionisation chambers were tested for their suitability in output factor and percentage depth dose measurement. Solid water proved a useful water-equivalent phantom material with discrepancies between measurements in water and solid water less than 2.3% for percentage depth dose and less than 0.6% for output factors. Larger measurement discrepancies were found for Plastic Water and Perspex. A PTW Markus chamber was found to compare well with a NE 2532/3 low energy chamber in percentage depth dose measurement, but discrepancies arose between the chambers in output factor measurements, up to 5% for small field sizes. Measurements indicated that the Markus chamber had an energy dependent response in the kilovoltage range, which could account for the discrepancy in output factor measurement. Copyright © 2005 ACPSEM/EA.
引用
收藏
页码:115 / 121
页数:6
相关论文
共 50 条
  • [1] Angular dependence of the output of a kilovoltage X-ray therapy unit
    Goharian, Mehran
    Nygren, Ian
    Tambasco, Mauro
    Barajas, Jose Eduardo Villarreal
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (03): : 276 - 284
  • [2] A Study On Output Factor Determination for Kilovoltage X-Ray Beams
    Kim, J.
    Mackonis, E. Claridge
    Hill, R.
    Kuncic, Z.
    MEDICAL PHYSICS, 2009, 36 (06) : 2582 - +
  • [3] Acceptance and commissioning of a modern kilovoltage X-ray therapy unit
    Aspradakis, M. M.
    Zucchetti, P.
    Seiler, R.
    Goetzfried, Th
    von Deschwanden, A.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (08) : 751 - 751
  • [4] Use of calculations to validate beam quality and relative dose measurements for a kilovoltage X-ray therapy unit
    B. J. Healy
    R. F. Hill
    Physical and Engineering Sciences in Medicine, 2022, 45 : 537 - 546
  • [5] Use of calculations to validate beam quality and relative dose measurements for a kilovoltage X-ray therapy unit
    Healy, B. J.
    Hill, R. F.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (02) : 537 - 546
  • [6] A study on the use of Gafchromic™ EBT3 film for output factor measurements in kilovoltage X-ray beams
    Gill, Simran
    Hill, Robin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2013, 36 (04) : 465 - 471
  • [7] A study on the use of Gafchromic™ EBT3 film for output factor measurements in kilovoltage X-ray beams
    Simran Gill
    Robin Hill
    Australasian Physical & Engineering Sciences in Medicine, 2013, 36 : 465 - 471
  • [8] METHOD OF SELECTION OF KILOVOLTAGE ON X-RAY SETS, AND ITS EFFECT ON X-RAY OUTPUT
    STERN, BE
    LINDSAY, DD
    BRITISH JOURNAL OF APPLIED PHYSICS, 1950, 1 (MAY): : 119 - 124
  • [9] Performance assessment of the Gulmay D3300 kilovoltage X-ray therapy unit
    Evans, PA
    Moloney, AJ
    Mountford, PJ
    BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (882): : 537 - 547
  • [10] Backscatter factor measurements for kilovoltage X-ray beams using thermoluminescent dosimeters (TLDs)
    Nelson, Vinod K.
    Hill, Robin F.
    RADIATION MEASUREMENTS, 2011, 46 (12) : 2097 - 2099