Backscatter factor measurements for kilovoltage X-ray beams using thermoluminescent dosimeters (TLDs)

被引:10
|
作者
Nelson, Vinod K. [1 ,2 ]
Hill, Robin F. [3 ,4 ]
机构
[1] Liverpool Canc Therapy Ctr, Dept Med Phys, Campbelltown, NSW 2560, Australia
[2] Macarthur Canc Therapy Ctr, Dept Med Phys, Campbelltown, NSW 2560, Australia
[3] Royal Prince Alfred Hosp, Dept Radiat Oncol, Camperdown, NSW 2050, Australia
[4] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
关键词
Thermoluminescent dosimeters; Backscatter factors; Monte Carlo; Kilovoltage x-ray beams; MONTE-CARLO; KV; RADIOTHERAPY; DEPENDENCE; VOLTAGES; PHOTON; WATER; FILM;
D O I
10.1016/j.radmeas.2011.08.019
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Backscatter factors (BSF) are used for calculation of absorbed dose in water from kilovoltage x-ray beams through the measurements using air-kerma based dosimetry protocols. The determination of backscatter factors has usually been performed using Monte Carlo calculations. The experimental measurement of BSFs requires that a dosimeter causes minimal perturbation to the x-ray fluence. In this work, we present the results of measuring BSFs using two different thermoluminescent dosimeters (TLDs). These were LiF:Mg,Ti (TLD100) and LiF:Mg,Cu,P (TLD100H) for x-ray beams with energies ranging from 75 to 300 kVp. We also calculated the BSFs for these x-ray beams using the EGSnrc Monte Carlo code and compared our results with published BSFs. With the exception of the 180 kVp x-ray beam (HVL 0.85 mm Cu), the BSFs measured with both materials agree to within 5% of the BSFs published in IPEMB and AAPM protocols and calculated by Monte Carlo simulations. From these results, we propose that BSFs can be readily determined using measurements with TLD100 and TLD100H chips for comparison with published BSFs. (C) 2011 Published by Elsevier Ltd.
引用
收藏
页码:2097 / 2099
页数:3
相关论文
共 50 条
  • [1] Energy Response of Thermoluminescent Microcube Dosimeters in Water for Kilovoltage X-Ray Beams
    Di Maso, L.
    Lawless, M.
    Culberson, W.
    DeWerd, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3624 - 3624
  • [2] COMPARISON OF X-RAY BEAMS USING THERMOLUMINESCENT DOSIMETERS
    SPURNY, Z
    MILU, C
    RACOVEANU, N
    PHYSICS IN MEDICINE AND BIOLOGY, 1973, 18 (02): : 276 - 278
  • [3] Energy and Depth Response of Thermoluminescent Dosimeters and Ionization Chambers in Water for Kilovoltage X-Ray Beams
    Lawless, M.
    Palmer, B.
    DeWerd, L.
    MEDICAL PHYSICS, 2015, 42 (06) : 3683 - 3683
  • [4] Thermoluminescent dosimeters for low dose X-ray measurements
    Del Sol Fernandez, S.
    Garcia-Salcedo, R.
    Sanchez-Guzman, D.
    Ramirez-Rodriguez, G.
    Gaona, E.
    de Leon-Alfaro, M. A.
    Rivera-Montalvo, T.
    APPLIED RADIATION AND ISOTOPES, 2016, 107 : 340 - 345
  • [5] Energy Response of Thermoluminescent Dosimeters in 3D-Printed Mouse Phantoms for Kilovoltage X-Ray Beams
    Viscariello, N.
    Hammer, C.
    Kunugi, K.
    DeWerd, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E596 - E596
  • [6] Determination of Backscatter Factors in Diagnostic Kilovoltage X-Ray Beams
    Tanabe, R.
    Araki, F.
    Ohno, T.
    MEDICAL PHYSICS, 2020, 47 (06) : E525 - E525
  • [7] Laser plasma X-ray measurements using thermoluminescent dosimeters and Si photodiodes
    Farnikova, M
    Krasa, J
    Kralikova, B
    Skala, J
    Parys, P
    Ryc, L
    Wolowski, J
    Woryna, E
    CZECHOSLOVAK JOURNAL OF PHYSICS, 1998, 48 (03) : 355 - 363
  • [8] Mass-energy absorption coefficient and backscatter factor ratios for kilovoltage x-ray beams
    Ma, CM
    Seuntjens, JP
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (01): : 131 - 143
  • [9] A Study On Output Factor Determination for Kilovoltage X-Ray Beams
    Kim, J.
    Mackonis, E. Claridge
    Hill, R.
    Kuncic, Z.
    MEDICAL PHYSICS, 2009, 36 (06) : 2582 - +
  • [10] The measurement of backscatter factors of kilovoltage X-ray beams using Gafchromic™ EBT2 film
    L. Smith
    R. Hill
    M. Nakano
    J. Kim
    Z. Kuncic
    Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 261 - 266