Output factor measurements for a kilovoltage x-ray therapy unit

被引:25
|
作者
Healy B.J. [1 ]
Gibbs A. [1 ]
Murry R.L. [1 ]
Prunster J.E. [1 ]
Nitschke K.N. [1 ]
机构
[1] Southern Zone Radiation Oncology Service - Mater Centre, Princess Alexandra Hospital, South Brisbane, QLD 4101, Raymond Tce.
关键词
Ionisation chambers; Kilovoltage x-rays; Output factors; Water substitutes;
D O I
10.1007/BF03178702
中图分类号
学科分类号
摘要
Output factors at the surface for treatment cones and lead cut-outs have been measured for a Pantak Therapax SXT 150 superficial therapy unit with x-ray beam qualities from 1 to 13 min A1 HVL. A variety of phantom materials and two ionisation chambers were tested for their suitability in output factor and percentage depth dose measurement. Solid water proved a useful water-equivalent phantom material with discrepancies between measurements in water and solid water less than 2.3% for percentage depth dose and less than 0.6% for output factors. Larger measurement discrepancies were found for Plastic Water and Perspex. A PTW Markus chamber was found to compare well with a NE 2532/3 low energy chamber in percentage depth dose measurement, but discrepancies arose between the chambers in output factor measurements, up to 5% for small field sizes. Measurements indicated that the Markus chamber had an energy dependent response in the kilovoltage range, which could account for the discrepancy in output factor measurement. Copyright © 2005 ACPSEM/EA.
引用
收藏
页码:115 / 121
页数:6
相关论文
共 50 条
  • [21] Quantifying the effect of off-focal radiation on the output of kilovoltage x-ray systems
    Ali, E. S. M.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2008, 35 (09) : 4149 - 4160
  • [22] Characterization of a prototype rapid kilovoltage x-ray image guidance system designed for a ring shape radiation therapy unit
    Cai, Bin
    Laugeman, Eric
    Mazur, Thomas R.
    Park, Justin C.
    Henke, Lauren E.
    Kim, Hyun
    Hugo, Geoffrey D.
    Mutic, Sasa
    Li, Hua
    MEDICAL PHYSICS, 2019, 46 (03) : 1355 - 1370
  • [23] Characterization of a prototype rapid kilovoltage x-ray image guidance system designed for a linear accelerator radiation therapy unit
    Arsenault, Theodore
    Arrue, Jonathan
    Gregg, Kenneth
    Baydoun, Atallah
    Rezai, Atefeh
    Langmack, Christian
    Henke, Lauren
    Spratt, Daniel E.
    Kashani, Rojano
    Price, Alex
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [24] FILM X-ray Revelation at Low Kilovoltage
    Wysnewski, Roy E.
    MATERIALS EVALUATION, 2012, 70 (12) : 1373 - 1380
  • [25] Image-guided radiation therapy based on kilovoltage X-ray imaging
    Jaffray, D
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S30 - S30
  • [26] Advances in kilovoltage x-ray beam dosimetry
    Hill, Robin
    Healy, Brendan
    Holloway, Lois
    Kuncic, Zdenka
    Thwaites, David
    Baldock, Clive
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (06): : R183 - R231
  • [27] Prediction of collimator scatter factor and phantom scatter factor for kilovoltage x-ray radiation fields
    Karabrahimi, V.
    Blais, N.
    MEDICAL PHYSICS, 2006, 33 (06) : 2130 - 2130
  • [28] COMPLETE CALIBRATION OF AN X-RAY THERAPY UNIT
    WILLIAMS, MD
    FELDMAN, A
    ANDERSON, JA
    AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1967, 99 (04): : 946 - &
  • [29] NEW SUPERFICIAL X-RAY THERAPY UNIT
    LILLICRAP, SC
    WILSON, BC
    EDSER, E
    STANLEY, D
    BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (555): : 193 - 195
  • [30] Ion recombination and polarity effect of ionization chambers in kilovoltage x-ray exposure measurements
    Das, IJ
    Akber, SF
    MEDICAL PHYSICS, 1998, 25 (09) : 1751 - 1757