Output factor measurements for a kilovoltage x-ray therapy unit

被引:25
|
作者
Healy B.J. [1 ]
Gibbs A. [1 ]
Murry R.L. [1 ]
Prunster J.E. [1 ]
Nitschke K.N. [1 ]
机构
[1] Southern Zone Radiation Oncology Service - Mater Centre, Princess Alexandra Hospital, South Brisbane, QLD 4101, Raymond Tce.
关键词
Ionisation chambers; Kilovoltage x-rays; Output factors; Water substitutes;
D O I
10.1007/BF03178702
中图分类号
学科分类号
摘要
Output factors at the surface for treatment cones and lead cut-outs have been measured for a Pantak Therapax SXT 150 superficial therapy unit with x-ray beam qualities from 1 to 13 min A1 HVL. A variety of phantom materials and two ionisation chambers were tested for their suitability in output factor and percentage depth dose measurement. Solid water proved a useful water-equivalent phantom material with discrepancies between measurements in water and solid water less than 2.3% for percentage depth dose and less than 0.6% for output factors. Larger measurement discrepancies were found for Plastic Water and Perspex. A PTW Markus chamber was found to compare well with a NE 2532/3 low energy chamber in percentage depth dose measurement, but discrepancies arose between the chambers in output factor measurements, up to 5% for small field sizes. Measurements indicated that the Markus chamber had an energy dependent response in the kilovoltage range, which could account for the discrepancy in output factor measurement. Copyright © 2005 ACPSEM/EA.
引用
收藏
页码:115 / 121
页数:6
相关论文
共 50 条
  • [31] A BRIEF NOTE ON A METHOD OF SWITCHING ON A REQUIRED KILOVOLTAGE AND ITS EFFECT ON X-RAY SET OUTPUT
    CONDON, JR
    JOURNAL OF THE FACULTY OF RADIOLOGISTS-LONDON, 1951, 3 (02): : 147 - 147
  • [32] Is a decrease in dose rate in a kilovoltage X-ray radiotherapy unit directly linked to tube metallization?
    Lizondo, M.
    Latorre-Mussol, A.
    Ruiz, A.
    Jornet, N.
    Eudaldo, T.
    Carrasco, P.
    Jordi, O.
    Delgado, P.
    Ribas, M.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S765 - S766
  • [33] Gold nanoparticles as dose-enhancement agent for kilovoltage X-ray therapy of melanoma
    Kim, So-Ra
    Kim, Eun-Hee
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2017, 93 (05) : 517 - 526
  • [34] Inverse optimization of low-cost kilovoltage x-ray arc therapy plans
    Breitkreutz, Dylan Y.
    Renaud, Marc-Andre
    Seuntjens, Jan
    Weil, Michael D.
    Zavgorodni, Sergei
    Bazalova-Carter, Magdalena
    MEDICAL PHYSICS, 2018, 45 (11) : 5161 - 5171
  • [35] The dosimetry of eye shields for kilovoltage X-ray beams
    Wang, D.
    Sobolewski, M.
    Hill, R.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (04) : 491 - 495
  • [36] The dosimetry of eye shields for kilovoltage X-ray beams
    D. Wang
    M. Sobolewski
    R. Hill
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 491 - 495
  • [37] Stereotactic breast irradiation with kilovoltage x-ray beams
    Garnica-Garza, H. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 983 - 995
  • [38] Determination of kilovoltage x-ray therapy depth doses with open-ended applicators
    Perkins, Anne
    Healy, Brendan
    Coldrey, Ben
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (03) : 1191 - 1201
  • [39] Study of dosimetry consistency for kilovoltage x-ray beams
    Ma, CM
    Li, XA
    Seuntjens, JP
    MEDICAL PHYSICS, 1998, 25 (12) : 2376 - 2384
  • [40] COMPARISON OF F FACTORS DERIVED FROM HVL AND X-RAY-SPECTRUM MEASUREMENTS FOR A SUPERFICIAL THERAPY X-RAY UNIT
    LEVY, L
    WAGGENER, R
    ZANCA, P
    MCDAVID, D
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (06): : 883 - &