Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study

被引:0
|
作者
Dan Wang
Xiu-hua Hu
Shi-zheng Zhang
Run-ze Wu
Su-su Xie
Bin Chen
Qiao-wei Zhang
机构
[1] Zhejiang University,Department of Radiology, Sir Run Run Shaw Hospital, School of Medicine
[2] Sir Run Run Shaw Institute of Clinical Medicine of Zhejiang University,undefined
[3] Siemens Healthcare,undefined
关键词
High-pitch; Coronary CT angiography; Computed tomography; Dose; Image quality;
D O I
暂无
中图分类号
学科分类号
摘要
To investigate the image quality and dose performance of 80 kV high-pitch spiral (HPS) coronary CT angiography (CCTA). 106 patients consecutively enrolled into prospectively ECG-triggering HPS CCTA (pitch = 3.4) exam using kV/ref. mAs = 80/400, 100/370, and 120/370 when patient BMI was ≤22.5 (n = 40), between 22.5 and 27.5 (n = 53) and >27.5 kg/m² (n = 13). Image quality was assessed per-segment by two observers independently using a 4-point scale (1—excellent, 4—non-diagnosable). Image noise and signal-to-noise ratio (SNR), contrast-to-noise ratio were measured. Diagnostic image quality was obtained in 503 of 507, 687 of 693, 164 of 167 coronary segments in 80, 100, 120 kV groups without significant difference (P = 0.482). The proportions of segments with score 1–4 were not significantly different among three kV groups (all P > 0.05). Image noise were significantly higher in 80 kV group than 100 and 120 groups (P < 0.001), while SNR was not (P = 0.097). The effective dose of 80 kV group (0.36 ± 0.03 mSv) was significantly lower than that of 100 kV group (0.86 ± 0.08 mSv) and 120 kV group (1.77 ± 0.18 mSv). The mean ± SD of HR in all patients was 54.8 ± 5.1 bpm. 80 kV HPS CCTA is feasible for patient with BMI ≤ 22.5 kg/m² which can save 58% dose than 100 kV group, while maintain diagnosable image quality.
引用
收藏
页码:415 / 423
页数:8
相关论文
共 50 条
  • [41] Coronary CT Angiography at 140 kV vs 120 kV: Assessment of Image Quality and Effective Dose
    Lee, A.
    Engel, L.
    Hui, G.
    Ghoshhajra, B.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (05)
  • [42] High-pitch prospective ECG-triggered helical coronary computed tomography angiography in clinical practice: image quality and radiation dose
    Kjell Bogaard
    Friso M. van der Zant
    Remco J. J. Knol
    Sjoerd Reinders
    Marije M. G. Krul
    Albert C. van Rossum
    Paul Knaapen
    Jan H. Cornel
    The International Journal of Cardiovascular Imaging, 2015, 31 : 125 - 133
  • [43] High-Pitch Spiral Computed Tomography Effect on Image Quality and Radiation Dose in Pediatric Chest Computed Tomography
    Lell, Michael M.
    May, Matthias
    Deak, Paul
    Alibek, Sedat
    Kuefner, Michael
    Kuettner, Axel
    Koehler, Henrik
    Achenbach, Stephan
    Uder, Michael
    Radkow, Tanja
    INVESTIGATIVE RADIOLOGY, 2011, 46 (02) : 116 - 123
  • [44] High-pitch prospective ECG-triggered helical coronary computed tomography angiography in clinical practice: image quality and radiation dose
    Bogaard, Kjell
    van der Zant, Friso M.
    Knol, Remco J. J.
    Reinders, Sjoerd
    Krul, Marije M. G.
    van Rossum, Albert C.
    Knaapen, Paul
    Cornel, Jan H.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2015, 31 (01): : 125 - 133
  • [45] High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose
    Rajiah, Prabhakar Shantha
    Dunning, Chelsea A. S.
    Rajendran, Kishore
    Tandon, Yasmeen K.
    Ahmed, Zaki
    Larson, Nicholas B.
    Collins, Jeremy D.
    Thorne, Jamison
    Williamson, Eric
    Fletcher, Joel G.
    McCollough, Cynthia
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2023, 58 (09) : 681 - 690
  • [46] Low-Dose CT of the Patients with 80kV Coronary Angiography Bypass Grafts Using Iterative Model Reconstruction: Evaluation of Image Quality and Radiation Dose Reduction
    Luo, C. C.
    Li, T.
    Yang, L.
    Huang, Z. L.
    Li, X. P.
    Liu, B.
    Yang, T.
    Qi, Y. Q.
    JOURNAL OF THE AMERICAN GERIATRICS SOCIETY, 2019, 67 : S682 - S682
  • [47] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    HEART, 2010, 96 (12) : 933 - 938
  • [48] Image quality and diagnostic accuracy of coronary CT angiography derived from low-dose dynamic CT myocardial perfusion: a feasibility study with comparison to invasive coronary angiography
    Xu Dai
    Mengmeng Yu
    Jingwei Pan
    Zhigang Lu
    Chengxing Shen
    Yining Wang
    Bin Lu
    Jiayin Zhang
    European Radiology, 2019, 29 : 4349 - 4356
  • [49] Image quality and diagnostic accuracy of coronary CT angiography derived from low-dose dynamic CT myocardial perfusion: a feasibility study with comparison to invasive coronary angiography
    Dai, Xu
    Yu, Mengmeng
    Pan, Jingwei
    Lu, Zhigang
    Shen, Chengxing
    Wang, Yining
    Lu, Bin
    Zhang, Jiayin
    EUROPEAN RADIOLOGY, 2019, 29 (08) : 4349 - 4356
  • [50] Successful low-dose CT assessment of coronary stents from a large consecutive cohort utilizing high-pitch spiral acquisition mode
    Ho, K. T.
    Yong, Q. W.
    Chua, K. C.
    Panknin, C.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2011, 13 (0A) : A23 - A23