Comparison of size modulation and conventional standard automated perimetry with the 24-2 test protocol in glaucoma patients

被引:0
|
作者
Kazunori Hirasawa
Nobuyuki Shoji
Masayuki Kasahara
Kazuhiro Matsumura
Kimiya Shimizu
机构
[1] Orthoptics and Visual Science,Department of Rehabilitation
[2] School of Allied Health Sciences,Department of Ophthalmology
[3] Kitasato University,undefined
[4] School of Medicine,undefined
[5] Kitasato University,undefined
来源
关键词
D O I
暂无
中图分类号
学科分类号
摘要
This prospective randomized study compared test results of size modulation standard automated perimetry (SM-SAP) performed with the Octopus 600 and conventional SAP (C-SAP) performed with the Humphrey Field Analyzer (HFA) in glaucoma patients. Eighty-eight eyes of 88 glaucoma patients underwent SM-SAP and C-SAP tests with the Octopus 600 24-2 Dynamic and HFA 24-2 SITA-Standard, respectively. Fovea threshold, mean defect and square loss variance of SM-SAP were significantly correlated with the corresponding C-SAP indices (P < 0.001). The false-positive rate was slightly lower and false-negative rate slightly higher, with SM-SAP than C-SAP (P = 0.002). Point-wise threshold values obtained with SM-SAP were moderately to strongly correlated with those obtained with C-SAP (P < 0.001). The correlation coefficients of the central zone were significantly lower than those of the middle to peripheral zone (P = 0.031). The size and depth of the visual field (VF) defect were smaller (P = 0.039) and greater (P = 0.043), respectively, on SM-SAP than on C-SAP. Although small differences were observed in VF sensitivity in the central zone, the defect size and depth and the reliability indices between SM-SAP and C-SAP, global indices of the two testing modalities were well correlated.
引用
收藏
相关论文
共 50 条
  • [21] Comparison of the Variability of Standard Automated Perimetry between Preperimetric Glaucoma Patients and Normal Controls
    Kim, Sung In
    Park, Hae-Young Lopilly
    Park, Chan Kee
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2018, 59 (01): : 44 - 49
  • [22] A COMPARISON OF GOLDMANN AND HUMPHREY AUTOMATED PERIMETRY IN PATIENTS WITH GLAUCOMA
    TROPE, GE
    BRITTON, R
    BRITISH JOURNAL OF OPHTHALMOLOGY, 1987, 71 (07) : 489 - 493
  • [23] Comparison of Standard Automated Perimetry, Frequency-Doubling Technology Perimetry, and Short-Wavelength Automated Perimetry for Detection of Glaucoma
    Liu, Shu
    Lam, Shi
    Weinreb, Robert N.
    Ye, Cong
    Cheung, Carol Y.
    Lai, Gilda
    Lam, Dennis Shun-Chiu
    Leung, Christopher Kai-Shun
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (10) : 7325 - 7331
  • [24] Effect of eye testing order on automated perimetry results using the Swedish interactive threshold algorithm standard 24-2
    Barkana, Y
    Gerber, Y
    Mora, R
    Liebmann, JM
    Ritch, R
    ARCHIVES OF OPHTHALMOLOGY, 2006, 124 (06) : 781 - 784
  • [25] Comparison between frequency-doubling technology perimetry and standard automated perimetry in early glaucoma
    Seong Ah Kim
    Chan Kee Park
    Hae-Young Lopilly Park
    Scientific Reports, 12
  • [26] Reproducibility of frequency doubling perimetry using a 24-2 test pattern.
    Spry, PGD
    Johnson, CA
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2001, 42 (04) : S150 - S150
  • [27] Comparison between frequency-doubling technology perimetry and standard automated perimetry in early glaucoma
    Kim, Seong Ah
    Park, Chan Kee
    Park, Hae-Young Lopilly
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [28] Repeatability of Automated Perimetry: A Comparison between Standard Automated Perimetry with Stimulus Size III and V, Matrix, and Motion Perimetry
    Wall, Michael
    Woodward, Kimberly R.
    Doyle, Carrie K.
    Artes, Paul H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (02) : 974 - 979
  • [29] Comparison of frequency doubling perimetry and standard achromatic computerized perimetry in patients with glaucoma
    Serguhn, S
    Spiegel, D
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2001, 239 (05) : 351 - 355
  • [30] Comparison of the Moorfields Motion Displacement Test to Standard Automated Perimetry
    Sharpe, G. P.
    Artes, P. H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)