Trace Elemental Determination Using a Portable Total Reflection X-Ray Fluorescence Spectrometer with a Collodion Film Sample Holder

被引:0
|
作者
Shinsuke Kunimura
Tomoki Shinkai
机构
[1] Tokyo University of Science,Department of Industrial Chemistry, Faculty of Engineering
来源
Analytical Sciences | 2017年 / 33卷
关键词
Aluminum; chromium; collodion; iron; manganese; portable spectrometer; X-ray fluorescence;
D O I
暂无
中图分类号
学科分类号
摘要
Using a portable total-reflection X-ray fluorescence (TXRF) spectrometer with a collodion film sample holder, a spectrum of an analyte containing 50 ng of aluminum was measured. The Al Kα line (1.49 keV) that partially overlaps with the Si Kα line (1.74 keV) from a quartz glass substrate usually used as a sample holder for TXRF analysis, was clearly detected when using the collodion film sample holder. To investigate the quantitative performance of the portable spectrometer with a collodion film sample holder, the concentrations of Cr, Mn, and Fe in a certified reference material of river water (JSAC 0302-3b), whose certified values are 10.0, 5.1, and 59.6 μg/L, respectively, were determined by the internalstandard method. We showed that approximate concentrations of these elements were determined.