Characterisation of a CZT detector for dosimetry of molecular radiotherapy

被引:9
|
作者
McAreavey, L. H. [1 ]
Harkness-Brennan, L. J. [1 ]
Colosimo, S. J. [1 ]
Judson, D. S. [1 ]
Boston, A. J. [1 ]
Boston, H. C. [1 ]
Nolan, P. J. [1 ]
Flux, G. D. [2 ]
Denis-Bacelar, A. M. [3 ]
Harris, B. [4 ]
Radley, I. [5 ]
Carroll, M. [6 ]
机构
[1] Univ Liverpool, Oliver Lodge Lab, Liverpool L69 7ZE, Merseyside, England
[2] Royal Marsden Hosp, Sutton SM2 5PT, Surrey, England
[3] Inst Canc Res, Sutton SM2 5NG, Surrey, England
[4] eV Prod Inc, Kromek USA, Saxonburg, PA 16056 USA
[5] Kromek Grp Plc, Sedgefield TS21 3FD, England
[6] Royal Liverpool & Broadgreen Univ Hosp, Liverpool L7 8XP, Merseyside, England
来源
基金
英国科学技术设施理事会;
关键词
Gamma camera; SPECT; PET PET/CT; coronary CT angiography (CTA); Gamma detectors (scintillators; CZT; HPG; HgI etc); Radiotherapy concepts; Solid state detectors; THYROID-CANCER; RADIOIODINE; THERAPY; SPECT; PET;
D O I
10.1088/1748-0221/12/03/P03001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Apixelated cadmium zinc telluride (CZT) detector has been characterised for the purpose of developing a quantitative single photon emission computed tomography (SPECT) system for dosimetry of molecular radiotherapy (MRT). This is the aim of the Dosimetric Imaging with CZT (DEPICT) project, which is a collaboration between theUniversity of Liverpool, The Royal Marsden Hospital, The Royal Liverpool and Broadgreen University Hospital, and the commercial partner Kromek. CZT is a direct band gap semiconductor with superior energy resolution and stopping power compared to scintillator detectors used in current SPECT systems. The inherent detector properties have been investigated and operational parameters such as bias voltage and peaking time have been selected to optimise the performance of the system. Good energy resolution is required to discriminate gamma-rays that are scattered as they are emitted from the body and within the collimator, and high photon throughput is essential due to the high activities of isotopes administered in MRT. The system has an average measured electronic noise of 3.31 keV full width at half maximum (FWHM), determined through the use of an internal pulser. The energy response of the system was measured across the energy region of interest 59.5 keV to 364.5 keV and found to be linear. The reverse bias voltage and peaking time producing the optimum FWHM and maximum photon throughput were 600 V and 0.5 mu s respectively. The average dead time of the system was measured as 4.84 mu s and charge sharing was quantified to be 0.71% at 59.5 keV. A pixel sensitivity calibration map was created and planar images of the medical imaging isotopes Tc-99m and I-123 were acquired by coupling the device to a prototype collimator, thereby demonstrating the suitability of the detector for the DEPICT project.
引用
收藏
页数:17
相关论文
共 50 条
  • [41] Recommendations for Multicentre Clinical Trials Involving Dosimetry for Molecular Radiotherapy
    Taprogge, J.
    Wadsley, J.
    Miles, E.
    Flux, G. D.
    CLINICAL ONCOLOGY, 2021, 33 (02) : 131 - 136
  • [42] Development of a calibration protocol for quantitative imaging for molecular radiotherapy dosimetry
    Wevrett, J.
    Fenwick, A.
    Scuffham, J.
    Nisbet, A.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 355 - 360
  • [43] Implementation of dosimetry for molecular radiotherapy; results from a European survey
    Peters, Steffie
    Tran-Gia, Johannes
    Agius, Sam
    Ivashchenko, Oleksandra V.
    Badel, Jean Noel
    Cremonesi, Marta
    Kurth, Jens
    Gabina, Pablo Minguez
    Richetta, Elisa
    Gleisner, Katarina Sjogreen
    Tipping, Jill
    Bardies, Manuel
    Stokke, Caroline
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 117
  • [44] Time-Resolved Dosimetry of Pulsed Photon Beams for Radiotherapy Based on Diamond Detector
    Pettinato, Sara
    Girolami, Marco
    Olivieri, Riccardo
    Stravato, Antonella
    Caruso, Cristina
    Salvatori, Stefano
    IEEE SENSORS JOURNAL, 2022, 22 (12) : 12348 - 12356
  • [45] The use of a silicon strip detector dose magnifying glass in stereotactic radiotherapy QA and dosimetry
    Wong, J. H. D.
    Knittel, T.
    Downes, S.
    Carolan, M.
    Lerch, M. L. F.
    Petasecca, M.
    Perevertaylo, V. L.
    Metcalfe, P.
    Jackson, M.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2011, 38 (03) : 1226 - 1238
  • [46] CZT detector and HXI development at CASS/UCSD
    Rothschild, Richard E.
    Tomsick, John A.
    Matteson, James L.
    Pelling, Michael R.
    Suchy, Slawomir
    TRANSIENT MILKY WAY: A PERSPECTIVE FOR MIRAX, 2006, 840 : 107 - +
  • [47] Evaluation of a MR-compatible CZT detector
    Despres, Philippe
    Izaguirre, Enrique W.
    Liu, Siyuan
    Cirignano, Leonard J.
    Kim, Hadong
    Wendland, Michael F.
    Shah, Kanai S.
    Hasegawa, Bruce H.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4324 - +
  • [48] Study of Spectral Distortion in Pixel CZT Detector
    Fu, Jianqiang
    Li, Yulan
    Liu, Yilin
    Zhang, Lan
    Li, Yuanjing
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [49] Design and performance of the HEXIS CZT detector module
    Rothschild, RE
    Matteson, JL
    Heindl, WA
    Pelling, MR
    X-RAY AND GAMMA-RAY DETECTORS AND APPLICATIONS IV, 2002, 4784 : 45 - 53
  • [50] Polarimetry with a Laue Lens and CZT Pixel Detector
    Caroli, Ezio
    da Silva, Rui M. Curado
    Stephen, John B.
    Pisa, Alessandro
    Auricchio, Natalia
    del Sordo, Stefano
    Donati, Ariano
    Schiavone, Filomena
    Landini, Gianni
    Honkimaki, Veijo
    Frontera, Filippo
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4897 - +