Quality assurance of stereotactic treatment planning system: implementation of Monte Carlo technique on Leksell Gamma Knife

被引:0
|
作者
Cheung, YC [1 ]
Yu, KN [1 ]
Yu, CP [1 ]
Ho, TK [1 ]
机构
[1] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon, Hong Kong, Peoples R China
来源
关键词
D O I
暂无
中图分类号
O59 [应用物理学];
学科分类号
摘要
The purpose of this report is to describe the quality assurance (QA) program on the treatment planning system, Leksell GammaPlan, using the Monte Carlo (MC) technique. Routine dose distribution measurements using a spherical polystyrene phantom and GafChromic films may not give accurate results since the measurements are not performed in a water medium and the small energy dependence of GafChromic films may affect the measurement results. Comparison of the dose distributions solely from measurement results and the GammaPlan output is not sufficient, hn alternative QA program on GammaPlan using the MC technique is introduced. The results from GammaPlan showed good consistency with the MC results. Small discrepancies of the relative dose along the z axis were observed on the 8 mm collimator helmet. Such discrepancies may however be too small to give clinical significance.
引用
收藏
页码:95 / 99
页数:5
相关论文
共 50 条
  • [31] GammaPlan®-Leksell Gamma Knife® radiosurgery treatment planning verification method
    Marcu, SM
    Wu, QJ
    Pillai, K
    Weinhous, MS
    MEDICAL PHYSICS, 2000, 27 (09) : 2146 - 2149
  • [32] GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion™
    Thomas Schaarschmidt
    Tae Hoon Kim
    Yong Kyun Kim
    Hye Jeong Yang
    Hyun-Tai Chung
    Journal of the Korean Physical Society, 2018, 73 : 1814 - 1820
  • [33] GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion
    Schaarschmidt, Thomas
    Kim, Tae Hoon
    Kim, Yong Kyun
    Yang, Hye Jeong
    Chung, Hyun-Tai
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 73 (12) : 1814 - 1820
  • [34] Monte Carlo treatment planning for stereotactic radiosurgery
    Solberg, TD
    DeMarco, JJ
    Holly, FE
    Smathers, JB
    DeSalles, AAF
    RADIOTHERAPY AND ONCOLOGY, 1998, 49 (01) : 73 - 84
  • [35] Monte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system
    Dept. of Phys. and Materials Science, City University of Hong Kong, Tat Chee Avenue, Kowloon Tong, Hong Kong
    不详
    Med. Phys., 9 (1673-1675):
  • [36] Monte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system
    Cheung, JYC
    Yu, KN
    Yu, CP
    Ho, RTK
    MEDICAL PHYSICS, 1998, 25 (09) : 1673 - 1675
  • [37] A quality assurance program in stereotactic radiosurgery using the gamma knife unit
    Stuecklschweiger, GE
    Feichtinger, K
    STRAHLENTHERAPIE UND ONKOLOGIE, 1998, 174 : 43 - 46
  • [38] Correction factors for commissioning and patient specific quality assurance of stereotactic fields in a Monte Carlo based treatment planning system TPS correction factors
    Goodall, Simon K.
    Rowshanfarzad, Pejman
    Ebert, Martin A.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 735 - 745
  • [39] Characterization and commissioning of a Leksell Gamma Knife ICON system for framed and frameless stereotactic radiosurgery
    Hu, Yue-Houng
    Hickling, Susannah, V
    Qian, Jing
    Blackwell, C. Robert
    McLemore, Luke B.
    Tryggestad, Erik J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (03):
  • [40] Forward Treatment Planning Techniques On Leksell Gamma Knife for the Normalization Effect Reduction
    Cheng, H.
    Pan, D.
    Tsai, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3603 - 3603