GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion™

被引:0
|
作者
Thomas Schaarschmidt
Tae Hoon Kim
Yong Kyun Kim
Hye Jeong Yang
Hyun-Tai Chung
机构
[1] Hanyang University,Department of Nuclear Engineering
[2] Catholic University College of Medicine,Department of Biomedicine & Health Sciences and Biomedical Engineering
[3] Seoul National University College of Medicine,Department of Neurosurgery
来源
关键词
GEANT4; Gamma Knife; Perfexion; Beam quality correction factor; Calibration; Small field dosimetry; TRS-483;
D O I
暂无
中图分类号
学科分类号
摘要
With the publication of TRS-483 in late 2017 the IAEA has established an international code of practice for reference dosimetry in small and non-standard fields based on a formalism first suggested by Alfonso et al. in 2008. However, data on beam quality correction factors (kQmsr,Q0fmsr,fref\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$k_{{Q_{msr,}}{Q_0}}^{{f_{msr,}}{f_{ref}}}$$\end{document}) for the Leksell Gamma Knife® Perfexion™ is scarce and what little data is available was obtained under conditions not necessarily in accordance with the IAEA’s recommendations. This study constitutes the first systematic attempt to calculate those correction factors by applying the new code of practice to Monte Carlo simulation using the GEANT4 toolkit. kQmsr,Q0fmsr,fref\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$k_{{Q_{msr,}}{Q_0}}^{{f_{msr,}}{f_{ref}}}$$\end{document} values were determined for three common ionization chamber detectors and five different phantom materials, with results indicating that in most phantom materials, all chambers were well suited for reference dosimetry with the Gamma Knife®. Similarities and differences between the results of this study and previous ones were also analyzed and it was found that the results obtained herein were generally in good agreement with earlier PENELOPE and EGSnrc studies.
引用
收藏
页码:1814 / 1820
页数:6
相关论文
共 50 条
  • [1] GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion
    Schaarschmidt, Thomas
    Kim, Tae Hoon
    Kim, Yong Kyun
    Yang, Hye Jeong
    Chung, Hyun-Tai
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 73 (12) : 1814 - 1820
  • [2] Geant4-based Monte Carlo simulation of the Leksell Gamma Knife®
    Romano, F.
    Sabini, M. G.
    Cuttone, G.
    Russo, G.
    Mongelli, V.
    Foroni, R.
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2581 - +
  • [3] FLUKA MONTE CARLO SIMULATION FOR THE LEKSELL GAMMA KNIFE PERFEXION: HOMOGENEOUS MEDIA.
    Battistoni, G.
    Bertolino, N.
    Brambilla, M. G.
    Cappucci, F.
    Mainardi, H. S.
    Torresin, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S176 - S176
  • [4] A modular Geant4 model of Leksell Gamma Knife Perfexion™
    Pipek, J.
    Novotny, J.
    Novotny, J., Jr.
    Kozubikova, P.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (24): : 7609 - 7623
  • [5] FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion radiosurgery system: Homogeneous media
    Battistoni, Giuseppe
    Cappucci, Fabrizio
    Bertolino, Nicola
    Brambilla, Maria Grazia
    Mainardi, Hae Song
    Torresin, Alberto
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (06): : 656 - 661
  • [6] Characterization of Gamma Knife Perfexion™ source based on Monte Carlo simulation
    Junios, J.
    Irhas, I.
    Novitrian, N.
    Soediatmoko, E.
    Haryanto, F.
    Su'ud, Z.
    Fielding, A. L.
    [J]. RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2020, 13 (04) : 398 - 404
  • [7] Characterization of Gamma Knife Perfexion™ source based on Monte Carlo simulation
    J. Junios
    I. Irhas
    N. Novitrian
    E. Soediatmoko
    F. Haryanto
    Z. Su’ud
    A. L. Fielding
    [J]. Radiological Physics and Technology, 2020, 13 : 398 - 404
  • [8] Monte Carlo calculated and experimentally determined output correction factors for small field detectors in Leksell Gamma Knife Perfexion beams
    Benmakhlouf, H.
    Johansson, J.
    Paddick, I.
    Andreo, P.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (10): : 3959 - 3973
  • [9] A Monte Carlo model and its commissioning for the Leksell Gamma Knife Perfexion radiosurgery system
    Yuan, Jiankui
    Machtay, Mitchell
    [J]. MEDICAL PHYSICS, 2017, 44 (09) : 4910 - 4918
  • [10] Inhomogeneity Correction using Monte Carlo Simulation for Leksell Gamma Knife Stereotactic Radiosurgery
    Moskvin, V. P.
    Henderson, M. A.
    Barriger, B.
    Can, M.
    DesRosiers, C.
    Johnstone, P. A. S.
    Das, I. J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S680 - S680