Quality assurance of stereotactic treatment planning system: implementation of Monte Carlo technique on Leksell Gamma Knife

被引:0
|
作者
Cheung, YC [1 ]
Yu, KN [1 ]
Yu, CP [1 ]
Ho, TK [1 ]
机构
[1] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon, Hong Kong, Peoples R China
来源
关键词
D O I
暂无
中图分类号
O59 [应用物理学];
学科分类号
摘要
The purpose of this report is to describe the quality assurance (QA) program on the treatment planning system, Leksell GammaPlan, using the Monte Carlo (MC) technique. Routine dose distribution measurements using a spherical polystyrene phantom and GafChromic films may not give accurate results since the measurements are not performed in a water medium and the small energy dependence of GafChromic films may affect the measurement results. Comparison of the dose distributions solely from measurement results and the GammaPlan output is not sufficient, hn alternative QA program on GammaPlan using the MC technique is introduced. The results from GammaPlan showed good consistency with the MC results. Small discrepancies of the relative dose along the z axis were observed on the 8 mm collimator helmet. Such discrepancies may however be too small to give clinical significance.
引用
收藏
页码:95 / 99
页数:5
相关论文
共 50 条
  • [11] System accuracy of Leksell Gamma Knife stereotactic radiosurgery
    Yu, C.
    Pagnini, P.
    Giannotta, S.
    Apuzzo, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2492 - 2492
  • [12] Motion Tracking Quality Assurance Platform for Leksell Gamma Knife HDMM System
    Chiu, T.
    Parsons, D.
    Kalantari, F.
    Stojadinovic, S.
    MEDICAL PHYSICS, 2019, 46 (06) : E512 - E512
  • [13] A Monte Carlo model and its commissioning for the Leksell Gamma Knife Perfexion radiosurgery system
    Yuan, Jiankui
    Machtay, Mitchell
    MEDICAL PHYSICS, 2017, 44 (09) : 4910 - 4918
  • [14] Filmless Quality Assurance Tests for the Leksell Gamma Knife Icon™
    Maraghechi, B.
    Kim, T.
    Dise, J.
    Goddu, S.
    Mitchell, T.
    Kavanaugh, J.
    Zoberi, J.
    Mutic, S.
    Knutson, N.
    MEDICAL PHYSICS, 2020, 47 (06) : E785 - E785
  • [15] An independent verification of dosimetry for the Leksell Gamma Knife® treatment planning system
    Weldon, M.
    Gupta, N.
    Kanellitsas, C.
    MEDICAL PHYSICS, 2007, 34 (06) : 2416 - 2417
  • [16] Monte Carlo calculated output factors of a Leksell Gamma Knife unit
    Cheung, JYC
    Yu, KN
    Ho, RTK
    Yu, CP
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12): : N247 - N249
  • [17] 3-D treatment planning system-Leksell Gamma Knife treatment planning system
    Guo, Fanqing
    MEDICAL DOSIMETRY, 2018, 43 (02) : 177 - 183
  • [18] FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion radiosurgery system: Homogeneous media
    Battistoni, Giuseppe
    Cappucci, Fabrizio
    Bertolino, Nicola
    Brambilla, Maria Grazia
    Mainardi, Hae Song
    Torresin, Alberto
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (06): : 656 - 661
  • [19] Geant4-based Monte Carlo simulation of the Leksell Gamma Knife®
    Romano, F.
    Sabini, M. G.
    Cuttone, G.
    Russo, G.
    Mongelli, V.
    Foroni, R.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2581 - +