Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer

被引:14
|
作者
De Bernardo, Maddalena [1 ]
Borrelli, Maria [2 ]
Cembalo, Giovanni [1 ]
Rosa, Nicola [1 ]
机构
[1] Univ Salerno, Dept Med Surg & Dent, I-84081 Baronissi, Italy
[2] Heinrich Heine Univ Dusseldorf, Dept Ophthalmol, D-40204 Dusseldorf, Germany
来源
MEDICINA-LITHUANIA | 2019年 / 55卷 / 10期
关键词
intraocular pressure; Icare PRO; standing position; sitting position; supine position; EPISCLERAL VENOUS-PRESSURE; BODY POSITION; APPLANATION TONOMETRY; GLAUCOMA PROGRESSION; RISK-FACTOR; FLUCTUATION; PNEUMATONOMETER; DIFFERENCE; BLOOD; HEAD;
D O I
10.3390/medicina55100701
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives: It has been established that body position can play an important role in intraocular pressure (IOP) fluctuation. IOP has been previously shown to increase significantly when lying down, relative to sitting; this type of investigation has not been extensively reported for the standing (ST) position. Therefore, this study aims to look for eventual significant IOP changes while ST, sitting, and lying down. Materials and Methods: An Icare PRO was used to measure the IOP of 120 eyes of 60 healthy individuals, with age ranging from 21 to 55 years (mean 29.22 +/- 9.12 years), in sitting, supine and ST positions; IOP was measured again, 5 min after standing (ST-5m). Results: Mean IOP difference between sitting and ST position was 0.39 +/- 1.93 mmHg (95% CI: 0.04 to 0.74 mmHg) (p = 0.027); between sitting and ST-5m, it was -0.48 +/- 1.79 mmHg (95% CI: -0.8 to -0.16 mmHg) (p = 0.004); between the sitting and supine position, it was -1.16 +/- 1.9 mmHg (95% CI: -1.5 to -0.82 mmHg) (p < 0.001); between the supine and ST position, it was 1.55 +/- 2.04 mmHg (95% CI: 1.18 to 1.92 mmHg) (p < 0.001); between supine and ST-5m, it was 0.68 +/- 1.87 mmHg (95% CI: 0.34 to 1.02 mmHg) (p < 0.001); and between ST-5m and ST, it was 0.94 +/- 1.95 mmHg (95% CI: 0.58 to 1.29 mmHg) (p < 0.001). Mean axial eye length was 24.45 mm (95% CI: 24.22 to 24.69 mm), and mean central corneal thickness was 535.30 mu m (95% CI: 529.44 to 541.19 mu m). Conclusion: Increased IOP in the ST-5m position suggests that IOP measurements should be performed in this position too. The detection of higher IOP values in the ST-5m position than in the sitting one, may explain the presence of glaucoma damage or progression in apparently normal-tension or compensated patients.
引用
下载
收藏
页数:10
相关论文
共 50 条
  • [1] The influence of the tonometer position on canine intraocular pressure measurements using the Tonovet (R) rebound tonometer
    de Oliveira, Juliana Kravetz
    Montiani-Ferreira, Fabiano
    Williams, David L.
    OPEN VETERINARY JOURNAL, 2018, 8 (01) : 68 - 76
  • [2] Evaluation of Rebound Tonometer for Measuring Intraocular Pressure at Deviated Angle and Position
    Takenaka, Joji
    Mochizuki, Hideki
    Kunihara, Eriko
    Tanaka, Junko
    Kiuchi, Yoshiaki
    CURRENT EYE RESEARCH, 2011, 36 (05) : 422 - 428
  • [3] Evaluation of the Effect of Body Position on Intraocular Pressure Measured with Rebound Tonometer
    Mayali, Huseyin
    Tekin, Beyza
    Kayikcioglu, Ozcan Rasim
    Kurt, Emin
    Ilker, Suleyman Sami
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2019, 49 (01): : 6 - 9
  • [4] Rebound Tonometer: Ideal Tonometer for Measurement of Accurate Intraocular Pressure
    Suman, Suwarna
    Agrawal, Ajai
    Pal, Virendra K.
    Pratap, Vir B.
    JOURNAL OF GLAUCOMA, 2014, 23 (09) : 633 - 637
  • [5] Effect of Body Position on Intraocular Pressure Measured by Rebound Tonometer in Healthy Children
    Uzlu, Dilek
    Akyol, Nurettin
    Turk, Adem
    Gursoy, Nurcan
    Somuncu, Ahmet Mehmet
    Oruc, Yavuz
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2020, 50 (05): : 271 - 274
  • [6] Evaluation of monkey intraocular pressure by rebound tonometer
    Yu, Wenhan
    Cao, Guiqun
    Qiu, Jinghua
    Liu, Xuyang
    Ma, Jia
    Li, Ni
    Yu, Man
    Yan, Naihong
    Chen, Lei
    Pang, Iok-Hou
    MOLECULAR VISION, 2009, 15 (235-36): : 2196 - 2201
  • [7] Fluctuation of Intraocular Pressure with Noncontact Tonometer and Rebound Tonometer in the Normal Volunteers
    Tatara, Shunya
    Maeda, Fumiatsu
    Tsukahara, Yoshinosuke
    Yaoeda, Kiyoshi
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [8] Noninvasive measurement of rodent intraocular pressure with a rebound tonometer
    Wang, WH
    Millar, JC
    Pang, IH
    Wax, MB
    Clark, AF
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (12) : 4617 - 4621
  • [9] A new rebound tonometer for home monitoring of intraocular pressure
    André Rosentreter
    Kerstin S. Jablonski
    Anne C. Mellein
    Stergiani Gaki
    Arno Hueber
    Thomas S. Dietlein
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2011, 249 : 1713 - 1719
  • [10] A new rebound tonometer for home monitoring of intraocular pressure
    Rosentreter, Andre
    Jablonski, Kerstin S.
    Mellein, Anne C.
    Gaki, Stergiani
    Hueber, Arno
    Dietlein, Thomas S.
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 249 (11) : 1713 - 1719