Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer

被引:14
|
作者
De Bernardo, Maddalena [1 ]
Borrelli, Maria [2 ]
Cembalo, Giovanni [1 ]
Rosa, Nicola [1 ]
机构
[1] Univ Salerno, Dept Med Surg & Dent, I-84081 Baronissi, Italy
[2] Heinrich Heine Univ Dusseldorf, Dept Ophthalmol, D-40204 Dusseldorf, Germany
来源
MEDICINA-LITHUANIA | 2019年 / 55卷 / 10期
关键词
intraocular pressure; Icare PRO; standing position; sitting position; supine position; EPISCLERAL VENOUS-PRESSURE; BODY POSITION; APPLANATION TONOMETRY; GLAUCOMA PROGRESSION; RISK-FACTOR; FLUCTUATION; PNEUMATONOMETER; DIFFERENCE; BLOOD; HEAD;
D O I
10.3390/medicina55100701
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives: It has been established that body position can play an important role in intraocular pressure (IOP) fluctuation. IOP has been previously shown to increase significantly when lying down, relative to sitting; this type of investigation has not been extensively reported for the standing (ST) position. Therefore, this study aims to look for eventual significant IOP changes while ST, sitting, and lying down. Materials and Methods: An Icare PRO was used to measure the IOP of 120 eyes of 60 healthy individuals, with age ranging from 21 to 55 years (mean 29.22 +/- 9.12 years), in sitting, supine and ST positions; IOP was measured again, 5 min after standing (ST-5m). Results: Mean IOP difference between sitting and ST position was 0.39 +/- 1.93 mmHg (95% CI: 0.04 to 0.74 mmHg) (p = 0.027); between sitting and ST-5m, it was -0.48 +/- 1.79 mmHg (95% CI: -0.8 to -0.16 mmHg) (p = 0.004); between the sitting and supine position, it was -1.16 +/- 1.9 mmHg (95% CI: -1.5 to -0.82 mmHg) (p < 0.001); between the supine and ST position, it was 1.55 +/- 2.04 mmHg (95% CI: 1.18 to 1.92 mmHg) (p < 0.001); between supine and ST-5m, it was 0.68 +/- 1.87 mmHg (95% CI: 0.34 to 1.02 mmHg) (p < 0.001); and between ST-5m and ST, it was 0.94 +/- 1.95 mmHg (95% CI: 0.58 to 1.29 mmHg) (p < 0.001). Mean axial eye length was 24.45 mm (95% CI: 24.22 to 24.69 mm), and mean central corneal thickness was 535.30 mu m (95% CI: 529.44 to 541.19 mu m). Conclusion: Increased IOP in the ST-5m position suggests that IOP measurements should be performed in this position too. The detection of higher IOP values in the ST-5m position than in the sitting one, may explain the presence of glaucoma damage or progression in apparently normal-tension or compensated patients.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Diagnostic accuracy of the iCare rebound tonometer compared to the Perkins applanation tonometer in assessing intraocular pressure in rural patients
    Li, Yifan
    Carpenter, Christopher R.
    Nicholson, Kathryn
    Milne, William Ken
    DIAGNOSIS, 2015, 2 (04) : 227 - 234
  • [32] Reproducibility of intraocular pressure self-measurement by ICare Home rebound tonometer and comparison with Goldman applanation tonometer
    Gaillard, R.
    Fenolland, J. R.
    Bonnel, S.
    Rosenberg, R.
    Marill, A. F.
    Bostanci, K.
    Giraud, J. M.
    Renard, J. P.
    ACTA OPHTHALMOLOGICA, 2015, 93
  • [33] Comparison of Intraocular Pressure Measurements between Icare Pro Rebound Tonometer and Tono-Pen XL Tonometer in Supine and Lateral Decubitus Body Positions
    Lee, Tae-Eun
    Yoo, Chungkwon
    Hwang, Jin-Young
    Lin, Shan
    Kim, Yong Yeon
    CURRENT EYE RESEARCH, 2015, 40 (09) : 923 - 929
  • [34] Comparison of Intraocular Pressure Measurements Between Icare Pro Rebound Tonometer and Tono-Pen XL Tonometer in Supine and Lateral Decubitus Body Positions
    Yoo, Chungkwon
    Lee, Tae-Eun
    Hwang, Jin-young
    Lin, Shan C.
    Kim, Yong Yeon
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [35] COMPARATIVE INTRAOCULAR-PRESSURE MEASUREMENTS WITH PNEUMATONOGRAPH AND GOLDMANN TONOMETER
    QUIGLEY, HA
    LANGHAM, ME
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1975, 80 (02) : 266 - 273
  • [36] Abstract title: An evaluation of the TonoVet rebound tonometer for measuring intraocular pressure in dogs and horses
    Knollinger, AM
    La Croix, NC
    Barrett, PM
    Miller, PE
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [37] Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber
    Eun Ji Lee
    Ki Ho Park
    Dong Myung Kim
    Young Cheol Yoo
    Shin Hee Kang
    Yu Jeong Kim
    Japanese Journal of Ophthalmology, 2008, 52 : 500 - 503
  • [38] Home assessment of diurnal intraocular pressure in healthy children using the Icare rebound tonometer
    Hsiao, Ya-Chuan
    Dzau, Jacqueline R.
    Flemmons, Meghan S.
    Asrani, Sanjay
    Jones, Sarah
    Freedman, Sharon F.
    JOURNAL OF AAPOS, 2012, 16 (01): : 58 - 60
  • [39] Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber
    Lee, Eun Ji
    Park, Ki Ho
    Kim, Dong Myung
    Yoo, Young Cheol
    Kang, Shin Hee
    Kim, Yu Jeong
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2008, 52 (06) : 500 - 503
  • [40] Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study
    Senay Asik Nacaroglu
    Emine Seker Un
    Mehmet Giray Ersoz
    Yelda Tasci
    International Journal of Ophthalmology, 2015, (03) : 540 - 543