Noninvasive Blood Pressure Measurement Using the NICCI Monitor in Adolescents During Intraoperative Anesthetic Care

被引:3
|
作者
Owusu-Bediako, Kwaku [1 ]
Bekiroglu, Ismail [1 ]
Rice-Weimer, Julie [1 ]
Murillo-Deluquez, Marcelino [1 ]
Tobias, Joseph D. [1 ,2 ]
机构
[1] Nationwide Childrens Hosp, Dept Anesthesiol & Pain Med, Columbus, OH 43205 USA
[2] Ohio State Univ, Dept Anesthesiol & Pain Med, Columbus, OH 43210 USA
关键词
Continuous blood pressure monitoring; Invasive arterial pressure; Intraoperative monitoring; Pediatrics anesthesiology; DEVICE; ACCURACY; CNAP(TM); SURGERY;
D O I
10.14740/cr1378
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The intermittent measurement of blood pressure (BP) remains the standard of care during anesthesia or procedural sedation. To improve the early identification of hemodynamic compromise, various noninvasive BP devices have been developed which provide a continuous BP reading. The current study evaluates the accuracy of a novel continuous BP device, the NICCI system, in adolescents weighing 40 - 80 kg. Methods: During intraoperative anesthetic care, BP readings (systolic, diastolic, and mean) were captured from the arterial cannula and the NICC1 device every second. Results: The study cohort included 44 pediatric patients undergoing major orthopedic, cardiac, and neurosurgical procedures. A total of 383,126 pairs of systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP) values from the arterial cannula and the MCCI device were analyzed. The absolute difference for SBP, DBP, and MAP values from the MCCI monitor and the arterial cannula were 10 +/- 8, 9 +/- 7, and 9 +/- 7 mm Hg, respectively. The difference between the BP values from the MCCI and the arterial cannula was <= 10 mm Hg for 60% of the SBP readings, 67% of the DBP readings, and 56% of the MAP readings. Using Bland-Altman analysis, the bias was 2, 3, and 4 mm Hg for the SBP, DBP, and MAP. Conclusions: Although there were technical limitations related to patient size that affected its ability to meet the strict accuracy criteria set by the American National Standards Institute/Association for the Advancement of Medical Instrumentation standards for noninvasive BP measurement (ANSI/AAMI SP10), the NICCI system provided a continuous noninvasive beat-to-beat BP measurement which was clinically relevant during a significant portion of intraoperative care.
引用
收藏
页码:154 / 161
页数:8
相关论文
共 50 条
  • [1] An Evaluation of the NICCI Monitor in Providing Continuous, Noninvasive Blood Pressure Readings in Children During Intraoperative Anesthetic Care
    Bekiroglu, Ismail
    Owusu-Bediako, Kwaku
    Rice-Weimer, Julie
    Tobias, Joseph D.
    JOURNAL OF CLINICAL MEDICINE RESEARCH-CANADA, 2022, 14 (04): : 158 - 164
  • [2] Noninvasive Blood Pressure Measurement by the Nexfin Monitor During Reduced Arterial Pulsatility: A Feasibility Study
    Martina, Jerson R.
    Westerhof, Berend E.
    Van Goudoever, Jeroen
    De Jonge, Nicolaas
    Van Lieshout, Johannes J.
    Lahpor, Jaap R.
    De Mol, Bas A. J. M.
    ASAIO JOURNAL, 2010, 56 (03) : 221 - 227
  • [3] Measurement of Central Systolic Blood Pressure by Pulse Volume Plethysmography With a Noninvasive Blood Pressure Monitor
    Sung, Shih-Hsien
    Cheng, Hao-Min
    Chuang, Shao-Yuan
    Shih, Yuan-Ta
    Wang, Kang-Ling
    Chen, Ying-Hwa
    Lin, Shing-Jong
    Yu, Wen-Chung
    Chen, Chen-Huan
    AMERICAN JOURNAL OF HYPERTENSION, 2012, 25 (05) : 542 - 548
  • [4] Estimation of Central Aortic Blood Pressure using a Noninvasive Automatic Blood Pressure Monitor
    Shih, Yuan-Ta
    Sun, Yi-Jung
    Chen, Chen-Huan
    Cheng, Hao-min
    Wei-Chih, Hu
    13TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, VOLS 1-3, 2009, 23 (1-3): : 497 - +
  • [5] Validation of a tonometric noninvasive arterial blood pressure monitor in the intensive care setting
    Steiner, LA
    Johnston, AJ
    Salvador, R
    Czosnyka, M
    Menon, DK
    ANAESTHESIA, 2003, 58 (05) : 448 - 454
  • [6] Blood volume measurement - The next intraoperative monitor?
    Barker, SJ
    ANESTHESIOLOGY, 1998, 89 (06) : 1310 - 1312
  • [7] Noninvasive measurement of blood pressure variability: accuracy of the Finometer monitor and comparison with the Finapres device
    Maestri, R
    Pinna, GD
    Robbi, E
    Capomolla, S
    La Rovere, MT
    PHYSIOLOGICAL MEASUREMENT, 2005, 26 (06) : 1125 - 1136
  • [8] RELIABILITY OF THE DINAMAP NONINVASIVE MONITOR IN THE MEASUREMENT OF BLOOD-PRESSURE OF III ASIAN NEWBORNS
    CHIA, F
    ANG, AT
    WONG, TW
    TAN, KW
    FUNG, KP
    LEE, J
    KHIN, K
    CLINICAL PEDIATRICS, 1990, 29 (05) : 262 - 267
  • [9] ERRONEOUS BLOOD-PRESSURE MEASUREMENTS WITH A NONINVASIVE BLOOD-PRESSURE MONITOR
    EHLERS, KC
    GRABENBAUER, SA
    ANESTHESIOLOGY, 1991, 74 (05) : 967 - 967
  • [10] Device for the self-measurement of blood pressure that can monitor blood pressure during sleep
    Chonan, K
    Kikuya, M
    Araki, T
    Fujiwara, T
    Suzuki, M
    Michimata, M
    Hashimoto, J
    Ohkubo, T
    Hozawa, A
    Yamamoto, N
    Miyawaki, Y
    Matsubara, M
    Imai, Y
    BLOOD PRESSURE MONITORING, 2001, 6 (04) : 203 - 205