Experimental study on the performance of an epithermal neutron flux monitor for BNCT

被引:14
|
作者
Guan, Xingcai [1 ]
Manabe, Masanobu [2 ]
Tamaki, Shingo [2 ]
Liu, Shuangtong [1 ]
Sato, Fuminobu [2 ]
Murata, Isao [2 ]
Wang, Tieshan [1 ]
机构
[1] Lanzhou Univ, Sch Nucl Sci & Technol, Tianshui South Rd 222, Lanzhou 730000, Gansu, Peoples R China
[2] Osaka Univ, Div Sustainable Energy & Environm Engn, Grad Sch Engn, Yamada Oka 2-1, Suita, Osaka 5650871, Japan
关键词
Epithermal neutron; Neutron flux monitor; BNCT neutron source; OKTAVIAN;
D O I
10.1016/j.apradiso.2016.04.003
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The performance of an epithermal neutron (0.5 eV < E-n < 10 keV) flux monitor designed for boron neutron capture therapy (BNCT) was experimentally studied by using a prototype monitor in an appropriate neutron field at the intense deuterium-tritium neutron source facility OKTAVIAN of Osaka University, Japan. It was convinced from the experimental results that the developed monitor worked well and the epithermal neutron fluxes in BNCT neutron sources can be measured within 5% by the monitor. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:28 / 32
页数:5
相关论文
共 50 条
  • [1] Monte Carlo optimization of an epithermal neutron flux monitor for BNCT
    Guan, Xingcai
    Murata, Isao
    Wang, Tieshan
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2017, 54 (10) : 1118 - 1122
  • [2] Performance verification of an epithermal neutron flux monitor using accelerator-based BNCT neutron sources
    Guan, X.
    Murata, I.
    Wang, T.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [3] Production of epithermal neutron beams for BNCT
    Bisceglie, E
    Colangelo, P
    Colonna, N
    Paticchio, V
    Santorelli, P
    Variale, V
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 476 (1-2): : 123 - 126
  • [4] Production of epithermal neutron beams for BNCT
    Colangelo, P
    Colonna, N
    Santorelli, P
    Variale, V
    Paticchio, V
    Maggipinto, G
    EXPERIMENTAL NUCLEAR PHYSICS IN EUROPE: ENPE 99: FACING THE NEXT MILLENNIUM, 1999, 495 : 454 - 457
  • [5] DETERMINATION OF THERMAL AND EPITHERMAL NEUTRON FLUXES WITH A SINGLE FLUX MONITOR
    MAENHAUT, W
    ADAMS, F
    HOSTE, J
    JOURNAL OF RADIOANALYTICAL CHEMISTRY, 1971, 9 (02): : 325 - &
  • [6] Experimental and theoretical evaluation of accelerator based epithermal neutron yields for BNCT
    Wielopolski, L
    Ludewig, H
    Powell, JR
    Raparia, D
    Alessi, JG
    Alburger, DE
    Zucker, MS
    Lowenstein, DI
    APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY, PTS 1 AND 2, 1999, 475 : 1033 - 1036
  • [7] Performance evaluation of the source description of the THOR BNCT epithermal neutron beam
    Liu, Yuan-Hao
    Tsai, Pi-En
    Yu, Hui-Ting
    Lin, Yi-Chun
    Huang, Yu-Shiang
    Huang, Chun-Kai
    Liu, Yen-Wan Hsueh
    Liu, Hong-Ming
    Jiang, Shiang-Huei
    APPLIED RADIATION AND ISOTOPES, 2011, 69 (12) : 1892 - 1896
  • [8] On the 252Cf primary and secondary gamma rays and epithermal neutron flux for BNCT
    Ghassoun, J.
    Merzouki, A.
    El Morabiti, A.
    Jehouani, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 263 (01): : 231 - 233
  • [9] Design study of neutron flux intensity monitor between ten and several hundred keV for BNCT
    Guan, Xingcai
    Murata, Isao
    Wang, Tieshan
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2016, 53 (08) : 1112 - 1119
  • [10] A biological comparison of epithermal neutron beams for BNCT
    Mason, A.
    Beynon, D.
    Green, S.
    Monk af Rosenschold, P.
    Giusti, V.
    Hopewell, J.
    Culbertson, C.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S244 - S245