Simulation-based Evaluation of OSEM Reconstruction Bias on low activity PET Data for the HRRT scanner

被引:3
|
作者
Dusch, Elodie [1 ]
Comtat, Claude [1 ]
Trebossen, Regine [1 ]
机构
[1] CEA, Serv Hosp Frederic Joliot, F-91406 Orsay, France
关键词
DOPAMINE TRANSPORTER; BINDING; BRAIN;
D O I
10.1109/NSSMIC.2009.5401956
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The ECAT High-Resolution Research Tomograph is a dedicated human brain positron emission tomography scanner. Due to gaps between its detector heads, 3D Ordinary Poisson OSEM algorithm is the reconstruction algorithm used in practice. For very low count statistics, it has been shown by several HRRT users groups that this algorithm converges to a biased solution. Unbiased quantification of low-activity regions is essential to measure accurately receptor-ligand kinetic parameters in brain receptor studies with reference tissue models. Using our simulation tool, we propose here to evaluate potential bias due to low-count statistics introduced by 3D OPOSEM on binding potential (BPND). We also compare OP-OS NEG-ML, a scaled gradient ascent algorithm with an accelerate rate of convergence for low activity values. For simulated [C-11]-(PEI)-I-2 dynamic data (1 hour acquisition, frame durations ranging from 1 to 5 minutes), we observe that 3D OP-OSEM and OP-OS NEC-ML converge to the same activity for the specific region, and also to the same BPND. We show using simulated cylinder phantom that the observed bias for this specific protocol is mainly due to the partial volume effect. This does not prevent the existence of a low count statistics bias with OP-OSEM for other protocols.
引用
收藏
页码:2770 / 2773
页数:4
相关论文
共 50 条
  • [1] Simulation-based evaluation of OSEM iterative reconstruction methods in dynamic brain PET studies
    Reilhac, Anthonin
    Tomeie, Sandrine
    Buvat, Irene
    Michell, Christian
    Keheren, Frank
    Costes, Nicolas
    [J]. NEUROIMAGE, 2008, 39 (01) : 359 - 368
  • [2] Spatial resolution of the HRRT PET scanner using 3D-OSEM PSF reconstruction
    Olesen, Oline Vinter
    Sibomana, Merence
    Keller, Sune Hogild
    Andersen, Flemming
    Jensen, Jorgen
    Holm, Soren
    Svarer, Claus
    Hojgaard, Liselotte
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3789 - 3790
  • [3] Simulation-based evaluation of NEG-ML iterative reconstruction of low count PET data
    Grezes-Besset, Louise
    Nuyts, Johan
    Boellard, Ronald
    Buvat, Irene
    Michel, Christian
    Pierre, Christian
    Costes, Nicolas
    Reilhac, Anthonin
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3009 - +
  • [4] An evaluation of the effect of filtering in 3-D OSEM reconstruction by using data from a high-resolution PET scanner
    Baghaei, H
    Uribe, JE
    Li, HD
    Wang, Y
    Aykac, M
    Liu, YQ
    Xing, T
    Wong, WH
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) : 2381 - 2386
  • [5] Evaluation of bias and variance in low-count OSEM list mode reconstruction
    Jian, Y.
    Planeta, B.
    Carson, R. E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (01): : 15 - 29
  • [6] Evaluation of Image Based Spatially Variant and Count Rate Dependant Point Spread Functions on the HRRT PET scanner
    Kotasidis, Fotis A.
    Angelis, Georgios I.
    Henderson, Jack
    Buckley, Anna
    Markiewicz, Pawel J.
    Green, Michael
    Anton-Rodriguez, Jose
    Lionheart, William R.
    Reader, Andrew J.
    Matthews, Julian C.
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3595 - +
  • [7] Evaluation of the 3D IMF-OSEM algorithm by using data from a high resolution PET scanner
    Baghaei, H
    Uribe, J
    Li, HD
    Wang, Y
    Aykac, M
    Liu, YQ
    Xing, T
    Wong, WH
    [J]. 2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1791 - 1795
  • [8] Examining publication bias-a simulation-based evaluation of statistical tests on publication bias
    Schneck, Andreas
    [J]. PEERJ, 2017, 5
  • [9] Reconstruction of attenuation maps for a PET/MR scanner based on the LSO background activity
    Kaltsas, T.
    Caldeira, L.
    Scheins, J.
    Tellmann, L.
    Kops, E. Rota
    Shah, N. J.
    Lerche, C.
    [J]. 2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [10] Data Mining in Intelligent SSD: Simulation-based Evaluation
    Jo, Yong-Yeon
    Chung, Moonjun
    Kim, Sang-Wook
    Oh, Hyunok
    [J]. 2016 INTERNATIONAL CONFERENCE ON BIG DATA AND SMART COMPUTING (BIGCOMP), 2016, : 123 - 128