Pattern of isthmus conduction recovery using open cooled and solid large-tip catheters for radiofrequency ablation of typical atrial flutter

被引:30
|
作者
Ventura, R [1 ]
Klemm, H [1 ]
Lutomsky, B [1 ]
Demir, C [1 ]
Rostock, T [1 ]
Weiss, C [1 ]
Meinertz, T [1 ]
Willems, S [1 ]
机构
[1] Univ Hamburg, Hosp Eppendorf, Dept Cardiol, D-20246 Hamburg, Germany
关键词
isthmus recovery atrial flutter ablation;
D O I
10.1046/j.1540-8167.2004.04125.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cooled Versus Large-Tip Catheters for Flutter Ablation. Introduction: Open cooled-tip and solid 8-mm-tip catheters have demonstrated safety and effectiveness for radiofrequency current (RFC) ablation of typical atrial flutter (AFL). However, data from prospective and randomized studies in this setting are lacking. Methods and Results: One hundred thirty consecutive patients (104 men; 61 +/- 11 years) with AFL were randomized to undergo RFC catheter ablation either using a solid 8-mm-tip catheter (group A, 65degreesC, 70 W, 60 s) or an open irrigated-tip catheter (group B, 65degreesC, 50 W, 60 s, 17 mL/min flow). Endpoint was bidirectional conduction isthmus block. In cases of repeated (two times) transient isthmus block, the catheter was changed (crossed over) to the catheter used in the other randomization arm, but patients remained in the original group following intention-to-treat analysis. The selected endpoint could be achieved in all patients after 12 +/- 6 RFC pulses in group A and 10 +/- 7 RFC pulses in group B (P = 0.11). Procedure times were longer (159 +/- 38 min vs 138 +/- 37 min, P = 0.002) and x-ray exposures higher in group A (fluoroscopy time 25 +/- 17 min vs 21 +/- 10 min, P = 0.08; x-ray dosage 3,133 +/- 2,576 cGy . cm(2) vs 2,326 +/- 1,405 cGy . cm, P = 0.03). Transient isthmus block was observed in 23 group A patients and 12 group B patients (P = 0.03). Onset time of transient isthmus block ranged from 0.5 to 27 minutes. Repeated transient isthmus block occurred in 8 of the 23 patients in group A after 19 3 RFC applications. After crossover to the cooled-tip catheter, the endpoint was reached another 5 I RFC pulses. In group B, all patients could be treated without change of ablation catheter. After a follow-up of 14 +/- 2 months, 2 patients (3%) in group A and 1 patient (1.5%) in group B presented with AFL recurrence. Conclusion: Open cooled-tip catheters are more effective than solid large-tip catheters for AFL ablation. The greater effectiveness is evident in cases showing repeated conduction recovery within the cavotricuspid isthmus. Primary use of open irrigated-tip catheters should be considered for AFL ablation.
引用
收藏
页码:1126 / 1130
页数:5
相关论文
共 50 条
  • [31] Convective cooling effect on cooled-tip catheter compared to large-tip catheter radiofrequency ablation
    Pilcher, Thomas A.
    Sanford, Anna L.
    Saul, J. Philip
    Haemmerich, Dieter
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (12): : 1368 - 1374
  • [32] Single center experience of radiofrequency catheter ablation of typical atrial flutter:: comparison a cooled tip versus conventional ablation catheter
    Arentz, T
    Von Rosenthal, J
    Stockinger, J
    Blum, T
    Weber, R
    Bürkle, G
    Kalusche, D
    EUROPEAN HEART JOURNAL, 2003, 24 : 593 - 593
  • [33] Cavo-tricuspid isthmus ablation for typical atrial flutter using an open-irrigation versus 10 mm-tip radiofrequency catheter: a randomized study
    Kuhne, M.
    Ilg, K.
    Crawford, T.
    Good, E.
    Chugh, A.
    Jongnarangsin, K.
    Bogun, F.
    Pelosi, F.
    Morady, F.
    Oral, H.
    EUROPEAN HEART JOURNAL, 2010, 31 : 554 - 555
  • [34] Transcristal conduction during isthmus ablation of typical atrial flutter:: Influence on success criteria
    Anselme, F
    Savouré, A
    Ouali, S
    Cribier, A
    Saoudi, N
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (02) : 184 - 189
  • [35] Usefulness of unipolar electrograms to detect isthmus block after radiofrequency ablation of typical atrial flutter
    Villacastin, J
    Almendral, J
    Arenal, A
    Castellano, NP
    Gonzalez, S
    Ortiz, M
    García, J
    Vallbona, B
    Moreno, J
    Portales, JF
    Torrecilla, EG
    CIRCULATION, 2000, 102 (25) : 3080 - 3085
  • [36] Clinical implication of the double potentials at the cavotricuspid isthmus for radiofrequency catheter ablation of typical atrial flutter
    Sasano, T
    Okishige, K
    JOURNAL OF ELECTROCARDIOLOGY, 2002, 35 (04) : 321 - 326
  • [37] Prospective randomized comparison of cooled radiofrequency versus standard radiofrequency energy for ablation of typical atrial flutter
    Atiga, WL
    Worley, SJ
    Hummel, J
    Berger, RD
    Gohn, DC
    Mandalakas, NJ
    Kalbfleisch, S
    Halperin, H
    Donahue, K
    Tomaselli, G
    Calkins, H
    Daoud, E
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08): : 1172 - 1178
  • [38] Clinical significance of residual slow cavotricuspid isthmus conduction after ablation of typical atrial flutter
    Takahashi, R
    Iesaka, Y
    Takahashi, A
    Hiroe, M
    Marumo, F
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (11): : 1902 - 1907
  • [39] Apparent Bidirectional Conduction Block Following Radiofrequency Catheter Ablation of Typical Atrial Flutter
    Robert F. Quintos
    Tawfik Barakat
    Andrew Mecca
    Brian Olshansky
    Journal of Interventional Cardiac Electrophysiology, 2001, 5 : 109 - 118
  • [40] Apparent bidirectional conduction block following radiofrequency catheter ablation of typical atrial flutter
    Quintos, RF
    Barakat, T
    Mecca, A
    Olshansky, B
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2001, 5 (01) : 109 - 118