The effect of suture anchor design and orientation on suture abrasion: An in vitro study

被引:64
|
作者
Bardana, DD
Burks, RT
West, JR
Greis, PE
机构
[1] Univ Utah, Dept Orthoped, Salt Lake City, UT 84132 USA
[2] Kingston Gen Hosp, Dept Surg, Kingston, ON K7L 2V7, Canada
[3] Univ Cincinnati, Noyes Giannestras Biomech Lab, Dept Biomed Engn, Cincinnati, OH USA
关键词
sutures; equipment failure; suture anchor; abrasion; cyclic loading; arthroscopy;
D O I
10.1053/jars.2003.50032
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: To evaluate the effects of suture anchor design and orientation on suture abrasion in a cyclic model. Type of Study: In vitro. Methods: Biomechanical studies have shown suture breakage to be a predominant mode of failure in a suture anchor repair construct. It is possible that suture abrasion during knot tying or in vivo cyclic loading may contribute to early failure. This study specifically investigates suture abrasion caused by 17 commonly used suture anchors and demonstrates the effects of suture anchor angulation and rotation on suture abrasion. To eliminate target tissue as a source of failure, all anchors were implanted into a solid block of sawbones material and tested with No. 2 Ethibond Excel sutures (Ethicon, Somerville, NJ). The testing model focused on 3 variables: suture anchor type, suture pull angle (SA) and angle of anchor rotation (RA). Abrasion testing was then performed on a servohydraulic materials testing system by continually cycling the suture back and forth through each anchor with an excursion of 4 cm at a rate of 0.5 Hz under a load of 10 N until suture failure occurred. Results: Sutures performed significantly better when cycled in line with the anchor at 0degrees SA with 0degrees RA than they did at 45degrees SA with 0degrees RA or 45degrees SA with 90degrees RA. We found no significant difference between anchors tested at 45degrees SA with 0degrees RA and 45degrees SA with 90degrees RA. For tests performed using metallic suture anchors, all constructs failed by fraying of the suture. Constructs using biopolymer anchors and nonabsorbable polymeric anchors experienced a mixture of suture and anchor eyelet failures. Conclusions: In addition to the statistically significant detrimental effects of suture anchor angulation and rotation on suture abrasion, suture anchor eyelet design may also influence suture abrasion. Surgeons should be aware of the effects of anchor angulation, suture position in the eyelet, and design and composition of the eyelet to maximize the durability of the construct.
引用
收藏
页码:274 / 281
页数:8
相关论文
共 50 条
  • [21] Biomechanical Evaluation of Proximal Hamstring Repair: All-Suture Anchor Versus Titanium Suture Anchor
    Otto, Alexander
    DiCosmo, Alyssa M.
    Baldino, Joshua B.
    Mehl, Julian
    Obopilwe, Elifho
    Cote, Mark P.
    Imhoff, Andreas B.
    Beitzel, Knut
    Mazzocca, Augustus D.
    Coyner, Katherine
    ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2020, 8 (01)
  • [22] Biomechanical Comparison of a Knotless Suture Anchor With Standard Suture Anchor in the Repair of Type II SLAP Tears
    Sileo, Michael J.
    Lee, Steven J.
    Kremenic, Ian J.
    Orishimo, Karl
    Ben-Avi, Simon
    McHugh, Malachy
    Nicholas, Stephen J.
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2009, 25 (04): : 348 - 354
  • [23] Comparison of Suture Anchor Repair and Suture Tape Augmentation for Reconstruction of the Anterior Talofibular Ligament: A Biomechanical Study
    Shao, Qingfeng
    Hu, Dahai
    Chen, Xinru
    Wang, Huajun
    Xiao, Lei
    Teng, Qiang
    Hong, Jinsong
    Hou, Huige
    Zheng, Xiaofei
    ORTHOPAEDIC SURGERY, 2024,
  • [24] THE 2.5 MM PUSHLOCK SUTURE ANCHOR SYSTEM VERSUS A TRADITIONAL SUTURE ANCHOR FOR ULNAR COLLATERAL LIGAMENT INJURIES OF THE THUMB: A BIOMECHANICAL STUDY
    Jarrett, C. D.
    Mcgillivary, G. R.
    Hutton, W. C.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2010, 35E (02) : 139 - 143
  • [25] All-Suture Anchor vs. Knotless Suture Anchor for the Treatment of Anterior Shoulder Instability-A Prospective Cohort Study
    Minkus, Marvin
    Aigner, Annette
    Wolke, Julia
    Scheibel, Markus
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (05)
  • [26] Strength of damaged suture: An in vitro study
    Wright, Patrick B.
    Budoff, Jeffrey E.
    Yeh, Ming Long
    Kelm, Zachary S.
    Luo, Zong-Ping
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (12): : 1270 - 1275
  • [27] Glenoid suture anchor fixation strength: Effect of insertion angle
    Ilahi, OA
    Al-Fahl, T
    Bahrani, H
    Luo, ZP
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2004, 20 (06): : 609 - 613
  • [28] Arthroscopic Suture Anchor Scapholunate Capsulodesis
    Chung, Sze Ryn
    Carratala, Vicente
    Delgado, Pedro J.
    Kalapos, Robert
    Kakar, Sanjeev
    JOURNAL OF WRIST SURGERY, 2024,
  • [29] Intraoperative arthroscopic suture anchor reloading
    Antonogiannakis, E
    Yiannakopoulos, CK
    Karabalis, C
    Hiotis, I
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2005, 21 (07): : 898.e1 - 898.e4
  • [30] SIMPLE SUTURE AND ANCHOR IN RABBIT HIPS
    Garcia Filho, Fernando Cal
    Guarniero, Roberto
    de Godoy, Rui Maciel, Jr.
    Martins Pereira, Cesar Augusto
    Matos, Marcos Almeida
    Garcia, Lucas Cortizo
    ACTA ORTOPEDICA BRASILEIRA, 2012, 20 (05): : 280 - 284