The effect of suture anchor design and orientation on suture abrasion: An in vitro study

被引:64
|
作者
Bardana, DD
Burks, RT
West, JR
Greis, PE
机构
[1] Univ Utah, Dept Orthoped, Salt Lake City, UT 84132 USA
[2] Kingston Gen Hosp, Dept Surg, Kingston, ON K7L 2V7, Canada
[3] Univ Cincinnati, Noyes Giannestras Biomech Lab, Dept Biomed Engn, Cincinnati, OH USA
关键词
sutures; equipment failure; suture anchor; abrasion; cyclic loading; arthroscopy;
D O I
10.1053/jars.2003.50032
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: To evaluate the effects of suture anchor design and orientation on suture abrasion in a cyclic model. Type of Study: In vitro. Methods: Biomechanical studies have shown suture breakage to be a predominant mode of failure in a suture anchor repair construct. It is possible that suture abrasion during knot tying or in vivo cyclic loading may contribute to early failure. This study specifically investigates suture abrasion caused by 17 commonly used suture anchors and demonstrates the effects of suture anchor angulation and rotation on suture abrasion. To eliminate target tissue as a source of failure, all anchors were implanted into a solid block of sawbones material and tested with No. 2 Ethibond Excel sutures (Ethicon, Somerville, NJ). The testing model focused on 3 variables: suture anchor type, suture pull angle (SA) and angle of anchor rotation (RA). Abrasion testing was then performed on a servohydraulic materials testing system by continually cycling the suture back and forth through each anchor with an excursion of 4 cm at a rate of 0.5 Hz under a load of 10 N until suture failure occurred. Results: Sutures performed significantly better when cycled in line with the anchor at 0degrees SA with 0degrees RA than they did at 45degrees SA with 0degrees RA or 45degrees SA with 90degrees RA. We found no significant difference between anchors tested at 45degrees SA with 0degrees RA and 45degrees SA with 90degrees RA. For tests performed using metallic suture anchors, all constructs failed by fraying of the suture. Constructs using biopolymer anchors and nonabsorbable polymeric anchors experienced a mixture of suture and anchor eyelet failures. Conclusions: In addition to the statistically significant detrimental effects of suture anchor angulation and rotation on suture abrasion, suture anchor eyelet design may also influence suture abrasion. Surgeons should be aware of the effects of anchor angulation, suture position in the eyelet, and design and composition of the eyelet to maximize the durability of the construct.
引用
收藏
页码:274 / 281
页数:8
相关论文
共 50 条
  • [1] Effect of suture anchor type, eyelet configuration, and loading condition on suture failure: An in vitro study
    Fidelis, Obinna P.
    Mulon, Pierre-Yves
    Anderson, David E.
    Crouch, Dustin L.
    VETERINARY SURGERY, 2025,
  • [2] Failure of suture material at suture anchor eyelets
    Meyer, DC
    Nyffeler, RW
    Fucentese, SF
    Gerber, C
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2002, 18 (09): : 1013 - 1019
  • [3] A knotless suture anchor - Design, function, and biomechanical testing
    Thal, R
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2001, 29 (05): : 646 - 649
  • [4] A Safe Technique to Ensure Not to Unload a Suture Anchor: Suture Dyeing
    Ocguder, Durmus Ali
    Sarikaya, Baran
    Eren, Semih
    Bozkurt, Ibrahim
    ARTHROSCOPY TECHNIQUES, 2023, 12 (04): : E545 - E547
  • [5] Arthroscopic Suture Anchor Tenodesis: Loop-Suture Technique
    Shon, Min Soo
    Koh, Kyoung Hwan
    Lim, Tae Kang
    Lee, Seung Won
    Park, Young Eun
    Yoo, Jae Chul
    ARTHROSCOPY TECHNIQUES, 2013, 2 (02): : E105 - E110
  • [6] Analysis of Suture Anchor Eyelet Position on Suture Failure Load
    Aktay, Sevima A.
    Kowaleski, Michael P.
    VETERINARY SURGERY, 2011, 40 (04) : 418 - 422
  • [7] Suture strength and angle of load application in a suture anchor eyelet
    Deakin, M
    Stubbs, D
    Bruce, W
    Goldberg, J
    Gillies, RM
    Walsh, WR
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2005, 21 (12): : 1447 - 1451
  • [8] The effect of two nonresorbable suture types on the mechanical performance over a metal suture anchor eyelet
    D. Acton
    A. Perry
    R. Evans
    A. Butler
    P. Stephens
    W. Bruce
    J. Goldberg
    D. Sonnabend
    W. R. Walsh
    Knee Surgery, Sports Traumatology, Arthroscopy, 2004, 12 : 165 - 168
  • [9] Material reaction to suture anchor
    Chow, JCY
    Gu, Y
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2004, 20 (03): : 314 - 316
  • [10] The effect of two nonresorbable suture types on the mechanical performance over a metal suture anchor eyelet
    Acton, D
    Perry, A
    Evans, R
    Butler, A
    Stephens, P
    Bruce, W
    Goldberg, J
    Sonnabend, D
    Walsh, WR
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2004, 12 (02) : 165 - 168