Oxide Formation and Instrumental Mass Bias in MC-ICP-MS: An Isotopic Case Study of Neodymium

被引:12
|
作者
Freres, Evelyn [1 ]
Weis, Dominique [1 ]
Newman, Karla [2 ]
Amini, Marghaleray [1 ]
Gordon, Kathy [1 ]
机构
[1] Univ British Columbia, Pacific Ctr Isotop & Geochem Res, Dept Earth Ocean & Atmospher Sci, Vancouver, BC V6T 1Z4, Canada
[2] Trent Univ, Water Qual Ctr, 1600 West Bank Dr, Peterborough, ON K9J 8E5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
MC‐ ICP‐ MS; neodymium; isotope measurement; instrumental mass bias; oxide formation; INDUCTIVELY-COUPLED PLASMA; FRACTIONATION CORRECTION; OPERATING-CONDITIONS; SPECTROMETRY; ARGON; RATIOS; NITROGEN; TEMPERATURES; INTERFACE; SIGNALS;
D O I
10.1111/ggr.12381
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
High rates of oxide formation affect the magnitude and behaviour of instrumental mass bias for Nd isotopic measurements performed with multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS) instruments, causing the traditional correction methods (e.g., internal and external normalisation) to fail. Here, we investigate the instrumental conditions that govern oxide formation and provide an extensive data set describing how different oxide formation rates affect the measurement error of Nd isotopic ratios. Results are reported for several instrumental set-ups including wet and dry plasma, different introduction methods, the addition of N-2, and various sampler and skimmer cone geometries. The differences in the behaviour of Nd isotopic ratios observed for dry and wet plasma require several reaction mechanisms to explain why oxide formation is associated with a non-linear mass bias for some rare earth elements. We developed a simple mathematical model to describe the behaviour of Nd isotopic ratios for a range of oxide formation rates and different instrumental settings and present a qualitative model that predicts the isotopic offsets of Nd ratios based on the cumulative contributions of the major sources of mass bias. A series of analytical recommendations for the determination of accurate and precise Nd ratios by MC-ICP-MS is presented.
引用
收藏
页码:501 / 523
页数:23
相关论文
共 50 条
  • [31] Use of Zr for mass bias correction in strontium isotope ratio determinations using MC-ICP-MS
    Yang, Lu
    Peter, Charlotte
    Panne, Ulrich
    Sturgeon, Ralph E.
    Journal of Analytical Atomic Spectrometry, 2008, 23 (09): : 1269 - 1274
  • [32] Use of Zr for mass bias correction in strontium isotope ratio determinations using MC-ICP-MS
    Yang, Lu
    Peter, Charlotte
    Panne, Ulrich
    Sturgeon, Ralph E.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (09) : 1269 - 1274
  • [33] Rapid isotopic analysis of uranium particles by laser ablation MC-ICP-MS
    Wimpenny, Josh
    Samperton, Kyle M.
    Sotorrio, Pedro
    Wellons, Matthew S.
    Scott, Spencer M.
    Willingham, David
    Knight, Kim
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (04) : 827 - 840
  • [34] Copper Isotopic Analysis in Geological and Biological Reference Materials by MC-ICP-MS
    Sullivan, Kaj
    Layton-Matthews, Daniel
    Leybourne, Matthew
    Kidder, James
    Mester, Zoltan
    Yang, Lu
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2020, 44 (02) : 349 - 362
  • [35] Magnesium isotopic composition of igneous rock standards measured by MC-ICP-MS
    Huang, Fang
    Glessner, Justin
    Ianno, Adam
    Lundstrom, Craig
    Zhang, Zhaofeng
    CHEMICAL GEOLOGY, 2009, 268 (1-2) : 15 - 23
  • [36] Isobaric Spike Method for Absolute Isotopic Ratio Determination by MC-ICP-MS
    Shuai, Kang
    Li, Weiqiang
    Hui, Hejiu
    ANALYTICAL CHEMISTRY, 2020, 92 (07) : 4820 - 4828
  • [37] In situ Ti isotopic analysis by femtosecond laser ablation MC-ICP-MS
    Huang, Chao
    Wang, Hao
    Xie, Liewen
    Nie, Nicole X.
    Yang, Yueheng
    Zhao, Xinmiao
    Li, Jin
    Tian, Heng-Ci
    Wu, Shitou
    Xu, Lei
    Yang, Jinhui
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (10) : 2165 - 2175
  • [38] Instrumental mass-independent fractionation accounting for the sensitivity of the double spike proportion effect by MC-ICP-MS: a case study of Zn isotope measurements
    Wang, Chunhong
    Zhu, Xiang-kun
    Li, Jin
    Zhu, Zhiyong
    Ma, Jianxiong
    Chen, Yuelong
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (06) : 1529 - 1539
  • [39] Precise analysis of antimony isotopic composition in geochemical materials by MC-ICP-MS
    Sun, Guangyi
    Wu, Yunjie
    Feng, Xinbin
    Wu, Xian
    Li, Xinyu
    Deng, Qianwen
    Wang, Feiyue
    Fu, Xuewu
    CHEMICAL GEOLOGY, 2021, 582
  • [40] Precise measurement of stable neodymium isotopes of geological materials by using MC-ICP-MS
    Ma, Jinlong
    Wei, Gangjian
    Liu, Ying
    Ren, Zhongyuan
    Xu, Yigang
    Yang, Yonghong
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (12) : 1926 - 1931