Analytical model for dental implant insertion torque

被引:7
|
作者
Yang, Baixuan [1 ]
Irastorza-Landa, Ainara [2 ]
Heuberger, Peter [2 ]
Ploeg, Heidi-Lynn [1 ]
机构
[1] Queens Univ Kingston, Dept Mech & Mat Engn, McLaughlin Hall,130 Stuart St, Kingston, ON K7L 3N6, Canada
[2] Nobel Biocare Serv AG, Balz Zimmermann Str 7, CH-8302 Kloten, Switzerland
基金
加拿大自然科学与工程研究理事会;
关键词
Dental implant; Torque; Model; PRIMARY STABILITY; BONE-DENSITY; OSSEOINTEGRATION; MECHANICS;
D O I
10.1016/j.jmbbm.2022.105223
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Maximum insertion torque (IT) for threaded dental implants is a primary clinical measurement to assess implant anchorage, and strongly influences the clinical outcome. Insertion torque is influenced by surgical technique, implant designs, and patient factors such as bone density and quality. In this study, an analytical model was proposed for IT to estimate contributions from the thread and taper separately. The purpose of this study was to test if the analytical model could 1. differentiate the parallel-walled and tapered implant; and, 2. represent four factors: bone surrogate density, drill protocol, implant surface finish and cutting flute. The IT was modeled as the sum of the torques from the thread's inclined plane and interface shear stress from the tapered body integrated over the surface area, respectively, with two main parameters: effective force, F ', F ' and effective pressure, p '. The effective force, relates to the clamping force from the thread, while the effective pressure, p ', associates with the contact pressure at the bone-implant interface. The model performed well (R2 = 0.88-1.0) and differentiated between the parallel-walled (p ' = 0) and tapered implants (p ' = 0.12). The model's parameters could individually represent the effects of the four factors. High bone surrogate density, two-step drill protocol, and rough surface increased both F ' and p '. The cutting flute had opposing effects on F ' and p ' (beta 4 = 0.35 and -0.24, respectively); and therefore, had the lowest net effect on IT. The proposed analytical model therefore improves the understanding of the principal contributors to dental implant IT by considering thread and taper mechanics independently.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Effect of insertion torque on titanium implant osseointegration: an animal experimental study
    Duyck, Joke
    Roesems, Rutger
    Cardoso, Marcio V.
    Ogawa, Toru
    Camargos, Germana De Villa
    Vandamme, Katleen
    CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (02) : 191 - 196
  • [42] Development of a New Implant Primary Stability Parameter: Insertion Torque Revisited
    Degidi, Marco
    Daprile, Giuseppe
    Piattelli, Adriano
    Iezzi, Giovanna
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2013, 15 (05) : 637 - 644
  • [43] The Effect of Insertion Angles and Depths of Dental Implant on the Initial Stability
    Wu, Hsiang-Chun
    Tsai, Ming-Tzu
    Hsu, Jui-Ting
    APPLIED SCIENCES-BASEL, 2020, 10 (09):
  • [44] Evaluation of Insertion Energy as Novel Parameter for Dental Implant Stability
    Grobecker-Karl, Tanja
    Dickinson, Anthony
    Heckmann, Siegfried
    Karl, Matthias
    Steiner, Constanze
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (09) : 1 - 7
  • [45] Peak insertion torque values of five mini-implant systems under different insertion loads
    Quraishi, Erma
    Sherriff, Martyn
    Bister, Dirk
    JOURNAL OF ORTHODONTICS, 2014, 41 (02) : 102 - 109
  • [46] Correlation of Fractal Dimension Values with Implant Insertion Torque and Resonance Frequency Values at Implant Recipient Sites
    Suer, Berkay Tolga
    Yaman, Zekai
    Buyuksarac, Bora
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2016, 31 (01) : 55 - 62
  • [47] LIMITED EVIDENCE SUGGESTS NO CORRELATION BETWEEN IMPLANT INSERTION TORQUE AND IMPLANT SURVIVAL AND MARGINAL BONE LOSS
    Radi, Iman Abd-Elwahab
    Samy, Mohannad Alaa Eldin Yehya
    JOURNAL OF EVIDENCE-BASED DENTAL PRACTICE, 2023, 23 (02)
  • [48] Correlation Between Implant Geometry, Implant Surface, Insertion Torque, and Primary Stability: In Vitro Biomechanical Analysis
    Falco, Antonello
    Berardini, Marco
    Trisi, Paolo
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (04) : 824 - 830
  • [49] Mechanical device design for determining mini dental implants insertion torque
    Polo-Cano, Karolan
    Rodriguez-Heilbron, Carolina
    Roldan-Mckinley, Javier
    UIS INGENIERIAS, 2020, 19 (04): : 181 - 197
  • [50] Sensitivity and Specificity of Radiographic Methods for Predicting Insertion Torque of Dental Implants
    Gonzalez Cortes, Arthur Rodriguez
    Eimar, Hazem
    Barbosa, Jorge de Sa
    Costa, Claudio
    Arita, Emiko Saito
    Tamimi, Faleh
    JOURNAL OF PERIODONTOLOGY, 2015, 86 (05) : 646 - 655