Evaluation of the exposure dose reduction in full-field digital mammography system

被引:0
|
作者
Tanigoshi, M [1 ]
Kobayashi, H [1 ]
Uchiyama, N [1 ]
Suzuki, T [1 ]
Kumazaki, T [1 ]
机构
[1] Nippon Med Coll, Dept Radiol, Bunkyo Ku, Tokyo, Japan
关键词
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We evaluated image quality characteristic of full-fielddigital mammography (Senographe 2000D; GEMS, CR-AL film;FUJIFILM) system in comparison with screen film mammography(Senographe DMR+;GEMS, Min-R 2000/Min-R 2000; Kodak) systemand investigated the possibility of exposure dose reduction in FFDMsystem. To evaluate each systems image quality characteristics, wemeasured characteristic curve, modulation transfer function MTF)value, and wiener spectrum ( W.S) value. image qualities and observer performance by RMI- 156 phantom and CDMAM phantom were alsoevaluated in accordance with the change of exposure dose. W.S value of FFDM become equal to that of SFM system in the amount of 60% exposure dose. The phantom studies reveled that the observer performance of FFDM system was also superior to that of SFM system. Those results indicated the possibility of exposure dose reduction in FFDM.
引用
收藏
页码:149 / 151
页数:3
相关论文
共 50 条
  • [21] Patient dose in full-field digital mammography: an Italian survey
    Gisella Gennaro
    Paola Baldelli
    Angelo Taibi
    Cosimo Di Maggio
    Mauro Gambaccini
    [J]. European Radiology, 2004, 14 : 645 - 652
  • [22] Patient dose in full-field digital mammography: an Italian survey
    Gennaro, G
    Baldelli, P
    Taibi, A
    Di Maggio, C
    Gambaccini, M
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (04) : 645 - 652
  • [23] Physical characteristics of a full-field digital mammography system
    Suryanarayanan, S
    Karellas, A
    Vedantham, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 533 (03): : 560 - 570
  • [24] Full-field digital mammography designed as a complete system
    Muller, S
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 1999, 31 (01) : 25 - 34
  • [25] EVALUATION OF MEAN GLANDULAR DOSE IN A FULL-FIELD DIGITAL MAMMOGRAPHY UNIT IN TABRIZ, IRAN
    Riabi, Hamed Alizadeh
    Mehnati, Parinaz
    Mesbahi, Asghar
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 142 (2-4) : 222 - 227
  • [26] Design and evaluation of a slot-scanning full-field digital mammography system
    Besson, GM
    Tesic, M
    [J]. RADIOLOGY, 2001, 221 : 464 - 464
  • [27] Design and evaluation of a slot-scanning Full-Field Digital Mammography system
    Besson, GM
    Koch, A
    Tesic, M
    Sottoriva, R
    Prieur-Drevron, P
    Munier, B
    Calais, E
    DeGroot, P
    [J]. MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 : 457 - 468
  • [28] Clinical evaluation of full-field digital mammography in a screening population
    Lewin, JM
    Hendrick, RE
    D'Orsi, CJ
    Moss, LJ
    Isaacs, P
    [J]. RADIOLOGY, 2000, 217 : 199 - 199
  • [29] A comparative evaluation of two full-field digital mammography units
    D Smith
    CP Lawinski
    A Mackenzie
    DS Evans
    NSA Wells
    D Emerton
    CA Lewis
    [J]. Breast Cancer Research, 2 (Suppl 2)
  • [30] Automatic evaluation of breast density for full-field digital mammography
    Lou, SL
    Fan, Y
    [J]. MEDICAL IMAGING 2000: IMAGE PROCESSING, PTS 1 AND 2, 2000, 3979 : 1362 - 1369