Sensitive Voltammetric Determination of Cadmium(II) with 8-Hydroxyquinoline and Ionic Liquid Modified Carbon Paste Electrode

被引:2
|
作者
Jia, Shaohui [1 ]
Gong, Shixing [2 ]
Cheng, Yong [2 ]
Wang, Huiyun [1 ]
Liu, Hejian [2 ]
Zhang, Bo [1 ]
Sun, Wei [2 ]
机构
[1] Jining Med Univ, Sch Pharm, Jining 272100, Peoples R China
[2] Qingdao Univ Sci & Technol, Coll Chem & Mol Engn, Qingdao 266042, Peoples R China
关键词
8-Hydroxyquinoline; 1-Ethyl-3-methylimidazoliam ethylsulphate; Cadmium; Differential pulse anodic stripping voltammetry; Carbon paste electrode; STRIPPING VOLTAMMETRY; ELECTROCHEMICAL DETECTION; PERIPHERAL-BLOOD; FILM; LEAD(II); ANTIMONY; SPECTROMETRY; EXTRACTION; CHLORIDE; COPPER;
D O I
10.1002/jccs.201400177
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In this paper, 8-hydroxyquinoline (HQ) and ionic liquid (IL) modified carbon paste electrode was fabricated and used for the sensitive determination of cadmium(II) with differential pulse anodic stripping voltammetry (DPASV). The modified electrode was prepared by the addition of HQ and IL 1-ethyl-3-methylimidazoliam ethylsulphate as the modifiers into the traditional carbon paste mixture. Cd(II) was preconcentrated and reduced on the surface of the modified electrode at the potential of -1.0 V (vs. SCE) by the co-contributions from the formation of HQ-Cd(II) complex and the accumulation effect of IL. Then the reduced Cd on the electrode surface was reoxidized by DPASV with a sensitive oxidation peak appeared at -0.79 V (vs. SCE). Under the optimal conditions, the oxidation peak current was proportional to the Cd(II) concentration in the range from 0.03 to 2.0 mu mol/L with the detection limit as 5.0 nmol/L (3 sigma). The proposed method was successfully applied to the water samples detection with the recovery in the range from 95.6% to 96.6%.
引用
收藏
页码:1217 / 1222
页数:6
相关论文
共 50 条