Image quality and radiation exposure in 320-row temporal bone computed tomography

被引:19
|
作者
Bauknecht, H. C. [1 ]
Siebert, E. [1 ]
Dannenberg, A. [1 ]
Bohner, G. [1 ]
Jach, C. [1 ]
Diekmann, S. [1 ]
Scheurig, C. [2 ]
Klingebiel, R. [1 ]
机构
[1] Univ Med, Charite, Dept Neuroradiol, Berlin, Germany
[2] Univ Med, Charite, Dept Radiol, Berlin, Germany
关键词
cone beam computed tomography; tomography; X-ray computed; temporal bone; petrous bone; radiation dosage; HIGH-DENSITY STRUCTURES; CT SCANNER;
D O I
10.1259/dmfr/14139527
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim was to define image quality and radiation exposure in the recently introduced 320-row CT of the temporal bone (tb) in comparison to a 16-row tb CT. Methods: A cadaveric head phantom was used for repeated tb volume CT studies (80120 kV, 25-150 mAs), performed in a 320-row scanner (single rotation, 0.5 mm slice thickness, kernel FC 51) in comparison to 16-row helical CT using standard acquisition parameters (SAP) of 120 kV and 75 mAs (kernel FC 53). Qualitative image evaluation was performed by two radiologists using a 5-point visual analogue scale. Image noise (D(SD)) was determined by region of interest (ROI) based measurements in cadaveric as well as water phantom studies. Dosimetric measurements of the effective dose (ED) and organ dose (OD) of the lens were performed. Results: Image quality of 320-row tb CT was equivalent to 16-row CT for SAP scans, resulting in image noise levels (D(SD) 16-/320-row) of 109/237 and 206/446 for air and bone respectively. D(SD) differences were predominantly (>90%) attributable to the different kernels available for tb studies in 16- and 320-row CT. Radiation exposure for 16-/320-row SAP scans amounted to 0.36/0.30 mSv (ED) and 10.0/8.4 mGy (lens dose). Conclusion: 320-row volume acquisition in tb CT delivers equivalent image quality to 16-row CT while decreasing radiation exposure figures by one sixth. Image noise increase in 320-row CT is negligible with respect to image quality. Dentomaxillofacial Radiology (2010) 39, 199-206. doi: 10.1259/dmfr/14139527
引用
收藏
页码:199 / 206
页数:8
相关论文
共 50 条
  • [11] Effective temporal resolution and image quality of volume scanning in 320-row detector CT
    Urikura, Atsushi
    Hara, Takanori
    Yoshida, Tsukasa
    Nishimaru, Eiji
    Hoshino, Takashi
    Ichikawa, Katsuhiro
    Nakaya, Yoshihiro
    Endo, Masahiro
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (02) : 503 - 510
  • [12] Image quality and radiation doses of volumetric 320-row computed tomography angiography with prospective electrocardiogram-gating in the assessment of coronary arteries in neonates and infants
    Roik, Danuta
    Kucinska, Beata
    Roik, Marek
    Werner, Bozena
    KARDIOLOGIA POLSKA, 2022, 80 (05) : 567 - 574
  • [13] 320-row coronary computed tomography angiography (CCTA) with automatic exposure control (AEC): effect of 100 kV versus 120 kV on image quality and dose exposure
    Ernesto Di Cesare
    Antonio Gennarelli
    Alessandra Di Sibio
    Valentina Felli
    Marco Perri
    Alessandra Splendiani
    Giovanni Luca Gravina
    Antonio Barile
    Carlo Masciocchi
    La radiologia medica, 2016, 121 : 618 - 625
  • [14] 320-row coronary computed tomography angiography (CCTA) with automatic exposure control (AEC): effect of 100 kV versus 120 kV on image quality and dose exposure
    Di Cesare, Ernesto
    Gennarelli, Antonio
    Di Sibio, Alessandra
    Felli, Valentina
    Perri, Marco
    Splendiani, Alessandra
    Gravina, Giovanni Luca
    Barile, Antonio
    Masciocchi, Carlo
    RADIOLOGIA MEDICA, 2016, 121 (08): : 618 - 625
  • [15] The New 320-Row Computed Tomography With Lower Radiation Exposure to Rule-out Coronary Artery Disease in Patients with Atrial Fibrillation
    Pelliccia, Francesco
    Evangelista, Antonietta
    Pergolini, Amedeo
    Pasceri, Vincenzo
    Viceconte, Nicola
    Pannarale, Giuseppe
    Varrica, Salvatore
    Romitelli, Susanna
    Pellegrini, Massimo
    Gaudio, Carlo
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 58 (20) : B165 - B165
  • [16] Discordance of anatomical and functional coronary stenoses assesed by 320-row computed tomography
    Iwai, T.
    Lee, T.
    Murai, T.
    Takagi, T.
    Hishikari, K.
    Kakuta, T.
    EUROPEAN HEART JOURNAL, 2012, 33 : 93 - 94
  • [17] Evaluation of Pharyngoesophageal Segment Using 320-Row Area Detector Computed Tomography
    Wattanapan, Pattra
    Kagaya, Hitoshi
    Inamoto, Yoko
    Saitoh, Eiichi
    Shibata, Seiko
    Iida, Takatoshi
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2018, 127 (12): : 888 - 894
  • [18] REDUCTION OF RADIATION DOSE OF 320-ROW CORONARY COMPUTED TOMOGRAPHY ANGIOGRAPHY THROUGH PRIOR CORONARY CALCIUM SCANNING
    Dewey, Marc
    Zimmermann, Elke
    Wollenberg, Ulrike
    Rief, Matthias
    Greupner, Johannes
    Hamm, Bernd
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [19] Low Radiation Dose and High Image Quality of 320-Row Coronary Computed Tomography Angiography Using a Small Dose of Contrast Medium and Refined Scan Timing Prediction
    Shirasaka, Takashi
    Nagao, Michinobu
    Yamasaki, Yuzo
    Kojima, Tsukasa
    Kondo, Masatoshi
    Hamasaki, Hiroshi
    Kamitani, Takeshi
    Kato, Toyoyuki
    Asayama, Yoshiki
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2020, 44 (01) : 7 - 12
  • [20] Usefulness of 320-Row Area Detector Computed Tomography for the Diagnosis of Cystic Falx Meningioma
    Tabuchi, Sadaharu
    Nakajima, Sadao
    CASE REPORTS IN ONCOLOGY, 2013, 6 (02): : 362 - 366