Image quality and radiation exposure in 320-row temporal bone computed tomography

被引:19
|
作者
Bauknecht, H. C. [1 ]
Siebert, E. [1 ]
Dannenberg, A. [1 ]
Bohner, G. [1 ]
Jach, C. [1 ]
Diekmann, S. [1 ]
Scheurig, C. [2 ]
Klingebiel, R. [1 ]
机构
[1] Univ Med, Charite, Dept Neuroradiol, Berlin, Germany
[2] Univ Med, Charite, Dept Radiol, Berlin, Germany
关键词
cone beam computed tomography; tomography; X-ray computed; temporal bone; petrous bone; radiation dosage; HIGH-DENSITY STRUCTURES; CT SCANNER;
D O I
10.1259/dmfr/14139527
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim was to define image quality and radiation exposure in the recently introduced 320-row CT of the temporal bone (tb) in comparison to a 16-row tb CT. Methods: A cadaveric head phantom was used for repeated tb volume CT studies (80120 kV, 25-150 mAs), performed in a 320-row scanner (single rotation, 0.5 mm slice thickness, kernel FC 51) in comparison to 16-row helical CT using standard acquisition parameters (SAP) of 120 kV and 75 mAs (kernel FC 53). Qualitative image evaluation was performed by two radiologists using a 5-point visual analogue scale. Image noise (D(SD)) was determined by region of interest (ROI) based measurements in cadaveric as well as water phantom studies. Dosimetric measurements of the effective dose (ED) and organ dose (OD) of the lens were performed. Results: Image quality of 320-row tb CT was equivalent to 16-row CT for SAP scans, resulting in image noise levels (D(SD) 16-/320-row) of 109/237 and 206/446 for air and bone respectively. D(SD) differences were predominantly (>90%) attributable to the different kernels available for tb studies in 16- and 320-row CT. Radiation exposure for 16-/320-row SAP scans amounted to 0.36/0.30 mSv (ED) and 10.0/8.4 mGy (lens dose). Conclusion: 320-row volume acquisition in tb CT delivers equivalent image quality to 16-row CT while decreasing radiation exposure figures by one sixth. Image noise increase in 320-row CT is negligible with respect to image quality. Dentomaxillofacial Radiology (2010) 39, 199-206. doi: 10.1259/dmfr/14139527
引用
收藏
页码:199 / 206
页数:8
相关论文
共 50 条
  • [1] Image Quality of Coronary Computed Tomography Angiography with 320-Row Area Detector Computed Tomography in Children with Congenital Heart Disease
    Tada, Akihiro
    Sato, Shuhei
    Kanie, Yuichiro
    Tanaka, Takashi
    Inai, Ryota
    Akagi, Noriaki
    Morimitsu, Yusuke
    Kanazawa, Susumu
    PEDIATRIC CARDIOLOGY, 2016, 37 (03) : 497 - 503
  • [2] Image quality of coronary computed tomography angiography with 320-Row area detector computed tomography in children with congenital heart disease
    Qin, Jie
    Cai, Songwang
    Peng, Lingrong
    Tang, Wenjie
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 68 (16) : C142 - C142
  • [3] Image Quality of Coronary Computed Tomography Angiography with 320-Row Area Detector Computed Tomography in Children with Congenital Heart Disease
    Akihiro Tada
    Shuhei Sato
    Yuichiro Kanie
    Takashi Tanaka
    Ryota Inai
    Noriaki Akagi
    Yusuke Morimitsu
    Susumu Kanazawa
    Pediatric Cardiology, 2016, 37 : 497 - 503
  • [4] Imaging of the temporal bone in children using low-dose 320-row area detector computed tomography
    Tada, Akihiro
    Sato, Shuhei
    Masaoka, Yoshihisa
    Kanazawa, Susumu
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2017, 61 (04) : 489 - 493
  • [5] Coronary computed tomography angiography and endocardial leads - Image quality in 320-row CT using iterative reconstruction
    Garmer, Marietta
    Bonsels, Marc
    Metz, Frauke
    Klein-Wiele, Oliver
    Brandts, Bodo
    Groenemeyer, Dietrich
    CLINICAL IMAGING, 2018, 50 : 157 - 163
  • [6] Low-dose non-contrast examination of the temporal bone using volumetric 320-row computed tomography
    Pirimoglu, Berhan
    Sade, Recep
    Sakat, Muhammed Sedat
    Polat, Gokhan
    Kantarci, Mecit
    ACTA RADIOLOGICA, 2019, 60 (07) : 908 - 916
  • [7] Comparison of image quality between synthetic and patients' electrocardiogram-gated 320-row pediatric cardiac computed tomography
    Eriko Maeda
    Go Shirota
    Eisuke Shibata
    Shuhei Komatsu
    Kenji Ino
    Rumiko Torigoe
    Osamu Abe
    Pediatric Radiology, 2020, 50 : 180 - 187
  • [8] Comparison of image quality between synthetic and patients' electrocardiogram-gated 320-row pediatric cardiac computed tomography
    Maeda, Eriko
    Shirota, Go
    Shibata, Eisuke
    Komatsu, Shuhei
    Ino, Kenji
    Torigoe, Rumiko
    Abe, Osamu
    PEDIATRIC RADIOLOGY, 2020, 50 (02) : 180 - 187
  • [9] Prediction of Cerebrovascular Reserve Capacity by Computed Tomography Perfusion Using 320-Row Computed Tomography
    Takahashi, Satoshi
    Tanizaki, Yoshio
    Kimura, Hiroaki
    Akaji, Kazunori
    Kano, Tadashige
    Suzuki, Kentaro
    Takayama, Youhei
    Kanzawa, Takao
    Shidoh, Satoka
    Nakazawa, Masaki
    Yoshida, Kazunari
    Mihara, Ban
    JOURNAL OF STROKE & CEREBROVASCULAR DISEASES, 2015, 24 (05): : 939 - 945
  • [10] Effective temporal resolution and image quality of volume scanning in 320-row detector CT
    Atsushi Urikura
    Takanori Hara
    Tsukasa Yoshida
    Eiji Nishimaru
    Takashi Hoshino
    Katsuhiro Ichikawa
    Yoshihiro Nakaya
    Masahiro Endo
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 503 - 510