Laser Doppler flowmetry: Characteristics of a modified single-fibre technique

被引:4
|
作者
Cai, H [1 ]
Rohman, H [1 ]
Larsson, SE [1 ]
Oberg, PA [1 ]
机构
[1] LINKOPING UNIV,DEPT ORTHOPAED,S-58185 LINKOPING,SWEDEN
关键词
iso-irradiance; laser beam; modified single fibre; ray tracing;
D O I
10.1007/BF02637015
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The single-fibre percutaneous laser Doppler technique has been used in previous studies of intramuscular blood flow. This method facilitates studies of blood flow in deep tissue volumes and minimises the tissue trauma. The technique has been further developed with the aim of improving the signal quality. This has been accomplished by modifying the geometry of the fibre tip. By melting the fibre core material, lenses of different shapes are formed. Flat, spherical and 'pear'-type tips have been manufactured and are evaluated theoretically and experimentally. The paraxial theory cannot accurately predict the position of zones of highest irradiance. Therefore, a ray-tracing program has been developed in the C language, by means of which some of the optical properties of the modified fibre tips can be simulated, Isoirradiance graphs and beam profiles are calculated for the three different fibre tips, Measured and calculated irradiance curves are used for evaluation of the properties of the ray-tracing model. The three types of fibre tips are also evaluated and compared in flow models. The sphere and pear-type probes show a higher flow sensitivity than the flat-end type, These improvements in flow sensitivity are interpreted as being related to the larger, strongly irradiated tissue volumes in front of the fibres. Intramuscular measurements with the pear-type probe show high sensitivities to induced blood flow changes.
引用
收藏
页码:2 / 8
页数:7
相关论文
共 50 条
  • [41] Laser doppler flowmetry in the hyperbaric environment
    G. Litscher
    K. O. Möller
    B. Ratzenhofer-komenda
    G. Schwarz
    T. Koop
    H. Kovac
    Lasers in Medical Science, 1997, 12 : 342 - 346
  • [42] Single-fibre response from an intact animal
    Rushton, WAH
    Barlow, HB
    NATURE, 1943, 152 : 597 - 598
  • [43] Laser Doppler flowmetry in the hyperbaric environment
    Litscher, G
    Moller, KO
    RatzenhoferKomenda, B
    Schwarz, G
    Koop, T
    Kovac, H
    LASERS IN MEDICAL SCIENCE, 1997, 12 (04) : 342 - 346
  • [44] Mathematical model of the single-fibre action potential
    Fleisher, S.M.
    Studer, M.
    Moschytz, G.S.
    Medical and Biological Engineering and Computing, 1984, 22 (05): : 433 - 439
  • [45] Velocity components in Laser Doppler Flowmetry
    Strömberg, T.
    JOURNAL OF VASCULAR RESEARCH, 2006, 43 : 5 - 5
  • [46] LASER DOPPLER FLOWMETRY IN EPISCLERAL VESSELS
    PETRIG, BL
    LANUTI, M
    RIVA, CE
    SCHWARTZ, B
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1991, 32 (04) : 944 - 944
  • [47] Evaluation of a new scanning laser Doppler flowmetry in comparison to a fundus camera-based laser Doppler flowmetry
    Okamoto, N
    Harino, S
    Suzuki, M
    Kitanishi, K
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 2803 - 2803
  • [48] Enhanced KLM Model for Single-Fibre Piezocomposite Transducers
    Angerer, Martin
    Zapf, Michael
    Gebhardt, Sylvia
    Neubert, Holger
    Ruiter, Nicole, V
    PROCEEDINGS OF THE 2020 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2020,
  • [49] HYPERTHERMAL LASER-DOPPLER-FLOWMETRY - A NEW DIAGNOSTIC-TECHNIQUE FOR MOTORIC NEUROPATHIES
    KOLTRINGER, P
    EBER, O
    LANGSTEGER, W
    LIND, P
    WAKONIG, P
    KLIMA, G
    MEDIZINISCHE WELT, 1989, 40 (09): : 267 - 269
  • [50] CUTANEOUS MICROCIRCULATORY FLOWMETRY EVALUATED BY LASER DOPPLER TECHNIQUE IN 100 HEALTHY-VOLUNTEERS
    MAUREL, A
    HAMON, P
    MACQUINMAVIER, I
    LAGRUE, G
    ADVANCES IN VASCULAR PATHOLOGY 1989, VOLS 1 & 2, 1989, 868 : 1177 - 1182