Clinical implementation of varian enhanced dynamic wedges into the pinnacle treatment planning system: Monte Carlo dosimetry and patient-specific quality assurance techniques

被引:0
|
作者
Ahmad, M.
Deng, J.
Chen, Z.
Lund, M.
Kimmett, J.
Moran, M.
Nath, R.
机构
[1] Yale Univ, Sch Med, New Haven, CT USA
[2] Yale New Haven Med Ctr, New Haven, CT 06504 USA
[3] William W Backus Hosp, Norwich, NY USA
关键词
D O I
10.1118/1.2760762
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present a commissioning study of enhanced dynamic wedges (EDWs) into the Pinnacle3 treatment planning system (TPS) and to evaluate various QA techniques for validating the treatment plans using EDWs. Method and Materials: Dose distribution modeling for EDW fields in the Pinnacle3 TPS is based on a combination of open‐field beam data and the Varian “Golden Segmented Treatment Table” (GSTT) unique to each photon beam. To validate the EDW models, dose profiles of 6− and 10−MV photon beams from a Clinac 2100C/D were measured in water for depths from near‐surface to 31.5 cm for a wide range of field sizes and wedge angles using a linear detector array of 25 energy‐compensated diodes (LDA‐25) to integrate the dose at multiple points in the EDW field for each exposure. The EDW output factors for square fields from 4 to 20 cm wide were measured in solid water using a cylindrical ionization chamber at a depth of 10 cm on central axis. The 6− and 10−MV photon beams emerging from the treatment head of the Clinac 2100C/D were fully simulated using the BEAMnrc and DOSXYZnrc was used to calculate the central‐axis percentage depth doses and dose profiles for the open and dynamically‐wedged fields in a water phantom. Film and MapCHECK data were compared with Pinnacle3 and MC predictions. Results: (1) Pinnacle collapsed‐cone convolution algorithm and MC simulations agreed with the measured EDW dose distributions to an accuracy of better than 3%. (2) Measured EDW output factors used for monitor‐unit calculation in Pinnacle3 TPS agreed with the MC predictions within 1–2%. (3) EDW‐based treatment plans were satisfactorily validated using MapCHECK and chamber measurements. Conclusions: We have successfully modeled EDWs into the Pinnacle3 TPS and demonstrated that MapCHECK and films dosimetry constitute valuable QA tools for validating planned dose distributions resulting from dynamically‐wedged beams. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2425 / 2425
页数:1
相关论文
共 50 条
  • [1] Clinical implementation of enhanced dynamic wedges into the Pinnacle treatment planning system: Monte Carlo validation and patient-specific QA
    Ahmad, Munir
    Deng, Jun
    Lund, Molly W.
    Chen, Zhe
    Kimmett, James
    Moran, Meena S.
    Nath, Ravinder
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (02): : 447 - 465
  • [2] Implementation of enhanced dynamic wedges in Pinnacle treatment planning system
    Alaei, P
    Higgins, PD
    Gerbi, BJ
    MEDICAL DOSIMETRY, 2005, 30 (04) : 228 - 232
  • [3] Clinical Implementation of Varian TrueBeam Portal Dosimetry: Comparison to Sun Nuclear MapCHECK2 for Patient-Specific Quality Assurance
    Roles, S.
    Broekhoven, B.
    Godwin, W.
    Rapchak, A.
    Mart, C.
    McDonald, D.
    Owen, G.
    Maynard, M.
    Peng, J.
    Fallon, K.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Monte Carlo-Based Patient-Specific Quality Assurance System for Carbon-Ion Radiotherapy: System Development and Clinical Implementation
    Yun, Y.
    Lee, S. H.
    Han, M. C.
    Kim, J. S.
    MEDICAL PHYSICS, 2024, 51 (10) : 7684 - 7684
  • [5] Implementation of a Monte Carlo dosimetry method for patient-specific internal emitter therapy
    Furhang, EE
    Chui, CS
    Kolbert, KS
    Larson, SM
    Sgouros, G
    MEDICAL PHYSICS, 1997, 24 (07) : 1163 - 1172
  • [6] Clinical implementation of PerFRACTION™ for pre-treatment patient-specific quality assurance
    Sang-Won Kang
    Boram Lee
    Changhoon Song
    Keun-Yong Eeom
    Bum-Sup Jang
    In Ah Kim
    Jae-Sung Kim
    Jin-Beom Chung
    Seonghee Kang
    Woong Cho
    Dong-Suk Shin
    Jin-Young Kim
    Minsoo Chun
    Journal of the Korean Physical Society, 2022, 80 : 516 - 525
  • [7] Clinical implementation of PerFRACTION™ for pre-treatment patient-specific quality assurance
    Kang, Sang-Won
    Lee, Boram
    Song, Changhoon
    Eeom, Keun-Yong
    Jang, Bum-Sup
    Kim, In Ah
    Kim, Jae-Sung
    Chung, Jin-Beom
    Kang, Seonghee
    Cho, Woong
    Shin, Dong-Suk
    Kim, Jin-Young
    Chun, Minsoo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 80 (06) : 516 - 525
  • [8] Monte Carlo modeling of the ModuLeaf miniature MLC for small field dosimetry and quality assurance of the clinical treatment planning system
    Crop, F.
    Reynaert, N.
    Pittomvils, G.
    Paelinck, L.
    De Gersem, W.
    De Wagter, C.
    Vakaet, L.
    De Neve, W.
    Thierens, H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (11): : 3275 - 3290
  • [9] Implementation of enhanced dynamic wedges into XIO treatment planning system
    Rodrigues, C.
    Batel, V.
    Germano, S.
    Monteiro Grillo, I.
    Pinto, J.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S512 - S513
  • [10] Clinical implementation of a Monte Carlo treatment planning system
    Ma, CM
    Mok, E
    Kapur, A
    Pawlicki, T
    Findley, D
    Brain, S
    Forster, K
    Boyer, AL
    MEDICAL PHYSICS, 1999, 26 (10) : 2133 - 2143