Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus

被引:195
|
作者
Tanner, Rebecca K. [1 ]
Fuller, Kate L. [1 ]
Ross, Megan L. R. [2 ]
机构
[1] Australian Inst Sport, Natl Sport Sci Qual Assurance Program, Canberra, ACT 2616, Australia
[2] Australian Inst Sport, Canberra, ACT, Australia
关键词
Portable lactate analysers; Lactate pro; Lactate scout; Lactate plus; Accuracy; Reliability; Linearity; ANALOX GM7; PERFORMANCE; PARAMETERS;
D O I
10.1007/s00421-010-1379-9
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Three portable blood lactate analysers, Lactate Pro (LP), Lactate Scout (LS) and Lactate Plus (L(+)), were evaluated. Analyser reliability and accuracy was assessed. For reliability, intra- and inter-analyser comparisons demonstrated that the LP (intra-TE = 0.5 mM, inter-TE = 0.4 mM) and L(+) (intra-TE = 0.4, inter-TE = 0.4 mM) displayed greater overall reliability than the LS (intra-TE = 1.0, inter-TE = 0.8 mM). At BLa < 4.0 mM, the LP (intra-TE = 0.1 mM) demonstrated greater reliability than the LS (intra-TE = 0.5 mM) and L(+) (intra-TE = 0.4 mM). At BLa > 8.0 mM, the LP (intra-TE = 0.5 mM, inter-TE = 0.4 mM) and L(+) (intra- and inter-TE = 0.4 mM) displayed greater reliability than the LS (intra-TE = 1.1 mM, inter-TE = 0.9 mM). For accuracy, the L(+) (SEE = 0.6 mM) compared more favourably to the LP than the LS (SEE = 1.1 mM). At BLa similar to 1.0-18.0 mM, the LS produced values that were up to 0.9 mM higher than the LP; the L(+) produced BLa that were within +/- 0.1 mM. All portable analysers tended to under-read the ABL 700 analyser. The suitability of the LP and L(+) as accurate analysers is supported by strong correlations (r = 0.91 and r = 0.94) and limits of agreement a parts per thousand currency sign2.1 mM. This study showed that the LP and L(+), compared well to each other, displayed good reliability and accuracy when compared to a laboratory-based analyser. Although the LS also displayed relatively good reliability, it was not as reliable or accurate as the LP or L(+).
引用
收藏
页码:551 / 559
页数:9
相关论文
共 50 条
  • [1] Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus
    Rebecca K. Tanner
    Kate L. Fuller
    Megan L. R. Ross
    European Journal of Applied Physiology, 2010, 109 : 551 - 559
  • [2] Evaluation of a portable lactate analyzer (Lactate Scout) in dogs
    Ferasin, Luca
    Dodkin, Stephen J.
    Amodio, Annarita
    Murray, Jane K.
    Papasouliotis, Kostas
    VETERINARY CLINICAL PATHOLOGY, 2007, 36 (01) : 36 - 39
  • [3] Evaluation of the Lactate Pro blood lactate analyser
    David B. Pyne
    Tanya Boston
    David T. Martin
    Andrew Logan
    European Journal of Applied Physiology, 2000, 82 : 112 - 116
  • [4] Evaluation of the Lactate Pro blood lactate analyser
    Pyne, DB
    Boston, T
    Martin, DT
    Logan, A
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2000, 82 (1-2) : 112 - 116
  • [5] Comparison of the Lactate Pro and Analox GM7 blood lactate analysers
    van Someren, KA
    Howatson, G
    Nunan, D
    Thatcher, R
    Shave, R
    INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 2005, 26 (08) : 657 - 661
  • [6] Identification of blood lactate parameters using the lactate pro portable analyzer
    van Someren, KA
    Howatson, G
    Nunan, D
    Thatcher, R
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2004, 36 (05): : S120 - S121
  • [7] Evaluation of the usefulness of the portable device Lactate Pro for measurement of lactate concentrations in equine whole blood
    van Oldruitenborgh-Oosterbaan, Marianne M. Sloet
    van den Broek, Erica T. W.
    Spierenburg, Astrid J.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2008, 20 (01) : 83 - 89A
  • [8] Effect of measuring blood lactate concentrations using different automated lactate analysers on blood lactate transition thresholds
    Buckley, JD
    Bourdon, PC
    Woolford, SM
    JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2003, 6 (04) : 408 - 421
  • [9] Determination of lactate in neonatal blood samples by a new portable device (Accutrend Lactate®).
    Gambke, BM
    Kalina, E
    Schlebusch, H
    CLINICAL CHEMISTRY, 2002, 48 (06) : A189 - A189
  • [10] Evaluation of the Lactate Plus monitor for plasma lactate concentration measurement in dogs
    Nye, Carolyn J.
    Musulin, Sarah E.
    Hanel, Rita M.
    Mariani, Christopher L.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2017, 27 (01) : 66 - 70