New, faster, image-based scatter correction for 3D PET

被引:0
|
作者
Watson, CC [1 ]
机构
[1] CTI PET Syst Inc, Knoxville, TN 37932 USA
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We report on a new numerical implementation of the single-scatter simulation scatter correction algorithm for 3D PET. Its primary advantage over the original implementation is that it is a much faster calculation, currently requiring less than 30 sec execution time per bed position for an adult thorax, thus making clinical whole-body scatter correction more practical. The new code runs on a single processor workstation CPU instead of a vector processor array, making it highly portable. It is modular and independent of any particular reconstruction code. The computed scatter contribution is now intrinsically scaled relative to the emission image and no longer requires normalization to the scatter tails in the sinogram when all activity is contained within the field of view, making it more robust against noise. The new algorithm has been verified against the original code on both phantom and human thorax studies. Initial results indicate that scatter correction may be accurately performed following, instead of prior to, either 3D reprojection or Fourier rebinning. Some evidence is presented that the single-scatter operator, when applied to an uncorrected emission image provides reasonable compensation for multiple scatter.
引用
收藏
页码:1637 / 1641
页数:3
相关论文
共 50 条
  • [21] Attenuation and scatter correction in fully 3D limited angle PET
    Pönisch, F
    Enghardt, W
    Lauckner, K
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 2115 - 2118
  • [22] A CONVOLUTION-SUBTRACTION SCATTER CORRECTION METHOD FOR 3D PET
    BAILEY, DL
    MEIKLE, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03): : 411 - 424
  • [23] Scatter correction techniques in 3D PET: a Monte Carlo evaluation
    Castiglioni, I.
    Cremonesi, O.
    Gilardi, M.C.
    Bettinardi, V.
    Rizzo, G.
    Savi, A.
    Bellotti, E.
    Fazio, F.
    IEEE Nuclear Science Symposium and Medical Imaging Conference, 1999, 1 : 738 - 742
  • [24] SCATTER CORRECTION IN 3-D PET
    LERCHER, MJ
    WIENHARD, K
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) : 649 - 657
  • [25] Parameterization of a model-based 3D whole-body PET scatter correction
    Wollenweber, SD
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1471 - 1475
  • [26] Image-based Fast 3D Reconstruction
    Xia, Jianjun
    Wang, Fei
    Zheng, Xiaocui
    PROCEEDINGS OF THE 2016 5TH INTERNATIONAL CONFERENCE ON MEASUREMENT, INSTRUMENTATION AND AUTOMATION (ICMIA 2016), 2016, 138 : 817 - 820
  • [27] An effective scatter correction method based on single scatter simulation for a 3D whole-body PET scanner
    高飞
    Yamada Ryoko
    Watanabe Mitsuo
    刘华锋
    Chinese Physics B, 2009, 18 (07) : 3066 - 3072
  • [28] An effective scatter correction method based on single scatter simulation for a 3D whole-body PET scanner
    Gao Fei
    Ryoko, Yamada
    Mitsuo, Watanabe
    Liu Hua-Feng
    CHINESE PHYSICS B, 2009, 18 (07) : 3066 - 3072
  • [29] Image-based 3D modelling: A review
    Remondino, Fabio
    El-Hakim, Sabry
    PHOTOGRAMMETRIC RECORD, 2006, 21 (115): : 269 - 291
  • [30] A Survey of Image-based 3D Reconstruction
    Fu, Xinfang