An effective scatter correction method based on single scatter simulation for a 3D whole-body PET scanner

被引:13
|
作者
Gao Fei [1 ]
Ryoko, Yamada [2 ]
Mitsuo, Watanabe [2 ]
Liu Hua-Feng [1 ]
机构
[1] Zhejiang Univ, State Key Lab Modern Opt Instrumentat, Hangzhou 310027, Peoples R China
[2] Hamamatsu Photon KK, Cent Res Lab, Hamamatsu, Shizuoka 4348601, Japan
关键词
positron emission tomography; scatter correction; single scatter simulation; Monte Carlo; ALGORITHM;
D O I
10.1088/1674-1056/18/7/076
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Hamamatsu SHR74000 is a newly designed full three-dimensional (3D) whole body positron emission tomography (PET) scanner with small crystal size and large field of view (FOV). With the improvement of sensitivity, the scatter events increase significantly at the same time, especially for large objects. Monte Carlo simulations help us to understand the scatter phenomena and provide good references for scatter correction. In this paper, we introduce an effective scatter correction method based on single scatter simulation for the new PET scanner, which accounts for the full 3D scatter correction. With the results from Monte Carlo simulations, we implement a new scale method with special concentration on scatter events from out side the axial FOV and multiple scatter events. The effects of scatter correction are investigated and evaluated by phantom experiments; the results show good improvements in quantitative accuracy and contrast of the images, even for large objects.
引用
收藏
页码:3066 / 3072
页数:7
相关论文
共 50 条
  • [1] An effective scatter correction method based on single scatter simulation for a 3D whole-body PET scanner
    高飞
    Yamada Ryoko
    Watanabe Mitsuo
    刘华锋
    Chinese Physics B, 2009, 18 (07) : 3066 - 3072
  • [2] Parameterization of a model-based 3D whole-body PET scatter correction
    Wollenweber, SD
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1471 - 1475
  • [3] A single scatter simulation technique for scatter correction in 3D pet
    Watson, CC
    Newport, D
    Casey, ME
    THREE-DIMENSIONAL IMAGE RECONSTRUCTION IN RADIOLOGY AND NUCLEAR MEDICINE, 1996, 4 : 255 - 268
  • [4] Attenuation and Scatter Correction for Whole-body PET Using 3D Generative Adversarial Networks
    Yang, Xiaofeng
    Lei, Yang
    Dong, Xue
    Wang, Tonghe
    Higgins, Kristin
    Liu, Tian
    Shim, Hyunsuk
    Curran, Walter
    Mao, Hui
    Nye, Jonathon
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [5] Scatter Correction with Combined Single-Scatter Simulation and Monte Carlo Simulation for 3D PET
    Ye, Jinghan
    Song, Xiyun
    Hu, Zhiqiang
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [6] A method for calibrating 3D PET without scatter correction
    Bailey, DL
    Jones, T
    1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1046 - 1049
  • [7] Evaluation of simulation-based scatter correction for 3D PET cardiac imaging
    Watson, CC
    Newport, D
    Casey, ME
    deKemp, RA
    Beanlands, RS
    Schmand, M
    1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1373 - 1377
  • [8] Benefits of 3D Scatter Correction for the HRRT - a Large Axial FOV PET Scanner
    Sibomana, Merence
    Keller, Sune H.
    Stute, Simon
    Comtat, Claude
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2954 - 2957
  • [9] A CONVOLUTION-SUBTRACTION SCATTER CORRECTION METHOD FOR 3D PET
    BAILEY, DL
    MEIKLE, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03): : 411 - 424
  • [10] Out-of-Field Scatter Estimation in 3D Whole body PET
    Iatrou, Maria
    Manjeshwar, Ravindra M.
    Wollenweber, Scott D.
    Ross, Steven G.
    Stearns, Charles W.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3886 - +