Monte Carlo study of dosimetric impact of gadolinium contrast medium in transverse field MR-Linac system

被引:0
|
作者
Ohno, Takeshi [1 ]
Kubota, Takahiro [2 ]
Yano, Masayuki [3 ]
Fujiwara, Yasuhiro [1 ]
Araki, Fujio [1 ]
机构
[1] Kumamoto Univ, Fac Life Sci, Dept Hlth Sci, 4-24-1 Kuhonji, Kumamoto, Japan
[2] Kokura Mem Hosp, 3-2-1 Asano, Fukuoka, Japan
[3] Saga Heavy Ion Med Accelerator Tosu, Koga Machi, Tosu, Saga, Japan
关键词
Gadolinium contrast medium; MR-Linac; Magnetic field; Monte Carlo technique; Dose distribution; MV RADIOTHERAPY ACCELERATOR; MEGAVOLTAGE PHOTON BEAMS; HIGH SIGNAL INTENSITY; DENTATE NUCLEUS; MAGNETIC-FIELD; DOSE INCREASE; SCANNER; NANOPARTICLES; TRANSPORT; AGENT;
D O I
10.1016/j.ejmp.2021.05.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to evaluate the dosimetric impact of gadolinium contrast medium (Gadovist) in a transverse MR-Linac system using Monte Carlo methods. The dose distributions were calculated using two heterogeneous multi-layer phantoms consisting of Gadovist, water, bone, and lung. The photon beam was irradiated with a filed size of 5 x 5 cm2, and a transverse magnetic field of 0-3.0 T was applied perpendicular to the incident photon beam. Next, dose distributions for brain, head and neck (H&N), and lung cancer patients were calculated using a patient voxel-based phantom with and without replacing the patient's GTV with Gadovist. The dose at the water-Gadovist interface increased by 8% without a magnetic field. A similar dose increment was observed at 0.35 T. In contrast, the dose increment at the water-Gadovist interface was small at 1.5 T and a dose decrement of 5% was observed at 3.0 T. The dose variation at the lung-Gadovist interface was larger than that at the water-Gadovist interface. The mass collision stopping power ratio for Gadovist was 7% lower than that for water, whereas, the electron fluence spectra at the water-Gadovist interface increased by 17.5%. In a patient study, Gadovist increased the Dmean for brain, H&N, and lung cancer patients by 0.65-8.9%. The dose variation due to Gadovist grew large in the low-dose region in H&N and lung cancer. The GTV dose variation due to Gadovist in all treatment site was below 2% at 0-3 T if the Gadovist concentration was lower than 0.2 mmol/ml-1.
引用
收藏
页码:19 / 30
页数:12
相关论文
共 50 条
  • [41] Dosimetric verification of a Monte Carlo treatment planning system for MR-guided proton therapy
    Fuchs, H.
    Padilla-Cabal, F.
    Oborn, B.
    Traneus, E.
    Georg, D.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1544 - S1545
  • [42] Simulated dosimetric impact of online replanning for stereotactic body radiation therapy of lymph node oligometastases on the 1.5T MR-linac
    Winkel, Dennis
    Kroon, Petra S.
    Werensteijn-Honingh, Anita M.
    Bol, Gijsbert H.
    Raaymakers, Bas W.
    Jurgenliemk-Schulz, Ina M.
    ACTA ONCOLOGICA, 2018, 57 (12) : 1705 - 1712
  • [43] Monte Carlo dosimetric study of the medium dose rate CSM40 source
    Vijande, J.
    Granero, D.
    Perez-Calatayud, J.
    Ballester, F.
    APPLIED RADIATION AND ISOTOPES, 2013, 82 : 283 - 288
  • [44] Impact of field number and beam angle on ERE for lung stereotactic body radiotherapy with 1.5 T MR-Linac
    Ding, S.
    Liu, H.
    Wang, B.
    Li, Y.
    Li, R.
    Liu, B.
    Xia, Y.
    Huang, X.
    CANCER RADIOTHERAPIE, 2021, 25 (04): : 366 - 372
  • [45] Impact of Field Number and Beam Angle on ERE for Lung Cancers Radiotherapy Planning in 1.5 T MR-Linac
    Ding, S.
    Liu, H.
    Wang, B.
    Li, Y.
    Xia, Y. F.
    Huang, X.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E282 - E282
  • [46] Small Field Output Factor (FOF) Measurements During Commissioning of a 0.35T MR-Linac System
    Alghamwa, M.
    Sohn, J.
    Hawranko, R.
    Kim, S.
    Yadav, P.
    Das, I.
    Song, W.
    MEDICAL PHYSICS, 2022, 49 (06) : E228 - E228
  • [47] Dosimetric performance of the Elekta Unity MR-linac system: 2D and 3D dosimetry in anthropomorphic inhomogeneous geometry
    Pappas, E.
    Kalaitzakis, G.
    Boursianis, T.
    Zoros, E.
    Zourari, K.
    Pappas, E. P.
    Makris, D.
    Seimenis, I
    Efstathopoulos, E.
    Maris, T. G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (22):
  • [48] Dosimetric Advantage of MRI-Guided Online Adaptive Radiation Therapy in Abdomen: Initial Clinical Experience with a High-Field MR-Linac
    Ahunbay, E.
    Paulson, E.
    Chen, X.
    Hall, W.
    Straza, M.
    Li, X.
    MEDICAL PHYSICS, 2019, 46 (06) : E228 - E229
  • [49] Lung stereotactic body radiotherapy with an MR-linac - Quantifying the impact of the magnetic field and real-time tumor tracking
    Menten, Martin J.
    Fast, Martin F.
    Nill, Simeon
    Kamerling, Cornelis P.
    McDonald, Fiona
    Oelfke, Uwe
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 (03) : 461 - 466
  • [50] Monte Carlo Calculations of the Skin Dose for Longitudinal Linac-MR System Using Realistic Three-Dimensional Magnetic Field Modeling
    Keyvanloo, A.
    Burke, B.
    Tadic, T.
    Warkentin, B.
    Kirkby, C.
    Rathee, S.
    Fallone, B.
    MEDICAL PHYSICS, 2012, 39 (06) : 3957 - 3957