Monte Carlo study of dosimetric impact of gadolinium contrast medium in transverse field MR-Linac system

被引:0
|
作者
Ohno, Takeshi [1 ]
Kubota, Takahiro [2 ]
Yano, Masayuki [3 ]
Fujiwara, Yasuhiro [1 ]
Araki, Fujio [1 ]
机构
[1] Kumamoto Univ, Fac Life Sci, Dept Hlth Sci, 4-24-1 Kuhonji, Kumamoto, Japan
[2] Kokura Mem Hosp, 3-2-1 Asano, Fukuoka, Japan
[3] Saga Heavy Ion Med Accelerator Tosu, Koga Machi, Tosu, Saga, Japan
关键词
Gadolinium contrast medium; MR-Linac; Magnetic field; Monte Carlo technique; Dose distribution; MV RADIOTHERAPY ACCELERATOR; MEGAVOLTAGE PHOTON BEAMS; HIGH SIGNAL INTENSITY; DENTATE NUCLEUS; MAGNETIC-FIELD; DOSE INCREASE; SCANNER; NANOPARTICLES; TRANSPORT; AGENT;
D O I
10.1016/j.ejmp.2021.05.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to evaluate the dosimetric impact of gadolinium contrast medium (Gadovist) in a transverse MR-Linac system using Monte Carlo methods. The dose distributions were calculated using two heterogeneous multi-layer phantoms consisting of Gadovist, water, bone, and lung. The photon beam was irradiated with a filed size of 5 x 5 cm2, and a transverse magnetic field of 0-3.0 T was applied perpendicular to the incident photon beam. Next, dose distributions for brain, head and neck (H&N), and lung cancer patients were calculated using a patient voxel-based phantom with and without replacing the patient's GTV with Gadovist. The dose at the water-Gadovist interface increased by 8% without a magnetic field. A similar dose increment was observed at 0.35 T. In contrast, the dose increment at the water-Gadovist interface was small at 1.5 T and a dose decrement of 5% was observed at 3.0 T. The dose variation at the lung-Gadovist interface was larger than that at the water-Gadovist interface. The mass collision stopping power ratio for Gadovist was 7% lower than that for water, whereas, the electron fluence spectra at the water-Gadovist interface increased by 17.5%. In a patient study, Gadovist increased the Dmean for brain, H&N, and lung cancer patients by 0.65-8.9%. The dose variation due to Gadovist grew large in the low-dose region in H&N and lung cancer. The GTV dose variation due to Gadovist in all treatment site was below 2% at 0-3 T if the Gadovist concentration was lower than 0.2 mmol/ml-1.
引用
收藏
页码:19 / 30
页数:12
相关论文
共 50 条
  • [31] The dosimetric impact of gadolinium-based contrast media in GBM brain patient plans for a MRI-Linac
    Ahmad, Syed Bilal
    Paudel, Moti Raj
    Sarfehnia, Arman
    Kim, Anthony
    Pang, Geordi
    Ruschin, Mark
    Sahgal, Arjun
    Keller, Brian M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (16): : N362 - N374
  • [32] Development and clinical application of a GPU-based Monte Carlo dose verification module and software for 1.5 T MR-LINAC
    Cheng, Bo
    Xu, Yuan
    Li, Shijun
    Ren, Qiang
    Pei, Xi
    Men, Kuo
    Dai, Jianrong
    Xu, Xie George
    MEDICAL PHYSICS, 2023, 50 (05) : 3172 - 3183
  • [33] Feasibility study: Polymer Gel Dosimetry On 0.35T MR-Linac System
    Marage, L.
    Boudet, J.
    Aubignac, L.
    Bessieres, I.
    Petitfils, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E769 - E769
  • [34] Impact of image deformation workflow in high-field MR-Linac treatment of head and neck cancer
    Christiansen, R. L.
    Johansen, J.
    Zukauskaite, R.
    Hansen, C. R.
    Mahmood, F.
    Bertelsen, A. S.
    Schytte, T.
    Hansen, O.
    Brink, C.
    Bernchou, U.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S904 - S904
  • [35] EVALUATING DOSIMETRIC IMPACT OF INTRA-FRACTION MOTION FOR ULTRA-HYPO-FRACTIONATED PROSTATE PATIENTS TREATED ON THE MR-LINAC
    Dang, Jennifer
    Kong, Vickie
    Li, Winnie
    Winter, Jeff
    Bayley, Andrew
    RADIOTHERAPY AND ONCOLOGY, 2020, 150 : S15 - S16
  • [36] Technical Note: Investigating the impact of field size on patient selection for the 1.5T MR-Linac
    Chuter, Robert W.
    Whitehurst, Philip
    Choudhury, Ananya
    van Herk, Marcel
    McWilliam, Alan
    MEDICAL PHYSICS, 2017, 44 (11) : 5667 - 5671
  • [37] Small Field Dosimetric Characterization of 7MV FFF Photons for Stereotactic MR-Guided Adaptive Radiotherapy Using the Unity MR-Linac
    Hwang, M.
    Chen, X.
    Oh, S.
    Lee, D.
    Pavord, D.
    Coopey, B.
    Sohn, J.
    MEDICAL PHYSICS, 2022, 49 (06) : E952 - E952
  • [38] A Monte Carlo Characterization of Positional Accuracy of a Double-Focused and Double-Stack Multileaf Collimator On a 0.35 T MR-Linac
    Khan, A.
    Simiele, E.
    DeWerd, L.
    Yadav, P.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [39] Extension and validation of a GPU-Monte Carlo dose engine gDPM for 1.5 T MR-LINAC online independent dose verification
    Li, Yongbao
    Ding, Shouliang
    Wang, Bin
    Liu, Hongdong
    Huang, Xiaoyan
    Song, Ting
    MEDICAL PHYSICS, 2021, 48 (10) : 6174 - 6183
  • [40] A Comparison of the Distortion in the Same Field MRI and MR-Linac System With a 3D Printed Phantom
    Liu, Xuechun
    Li, Zhenjiang
    Rong, Yi
    Cao, Minsong
    Li, Hongyu
    Jia, Chuntao
    Shi, Liting
    Lu, Weizhao
    Gong, Guanzhong
    Yin, Yong
    Qiu, Jianfeng
    FRONTIERS IN ONCOLOGY, 2021, 11